Electrode contact noise in surface biopotential measurements

M. Fernandez, R. Pallàs-Areny
{"title":"Electrode contact noise in surface biopotential measurements","authors":"M. Fernandez, R. Pallàs-Areny","doi":"10.1109/IEMBS.1992.5760887","DOIUrl":null,"url":null,"abstract":"Skin-Electrode interface noise is important in high resolution surface biopotential recordings. In order to determine its contribution, we have carried out several measurements. First, we measured the noise power spectrum density of several pairs of Ag/AgQ electrodes placed face to face and then placed on the body surface (leg and arm). We simultaneously measured the impedance of these pairs of electrodes and calculated the thermal noise associated to their real part. We found that noise at low frequencies is larger than noise at high frequencies, the power spectrum density increasing as 1/f. We have also found that calculated thermal noise is lower than measured electrode noise. Next, we measured noise and impedance for different electrode areas. For electrodes placed face to face we found that noise and resistance increase when electrode area decreases. For electrodes placed on the body surface, \"noise\" power spectrum density (probably EMG) increases when electrode area increases. We also calculated the Probability Density Function (PDF) of electrode noise. We found that it looks gaussian for electrodes placed face to face, but not for electrodes placed on the body surface.","PeriodicalId":6457,"journal":{"name":"1992 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society","volume":"45 1","pages":"123-124"},"PeriodicalIF":0.0000,"publicationDate":"1992-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"9","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1992 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IEMBS.1992.5760887","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 9

Abstract

Skin-Electrode interface noise is important in high resolution surface biopotential recordings. In order to determine its contribution, we have carried out several measurements. First, we measured the noise power spectrum density of several pairs of Ag/AgQ electrodes placed face to face and then placed on the body surface (leg and arm). We simultaneously measured the impedance of these pairs of electrodes and calculated the thermal noise associated to their real part. We found that noise at low frequencies is larger than noise at high frequencies, the power spectrum density increasing as 1/f. We have also found that calculated thermal noise is lower than measured electrode noise. Next, we measured noise and impedance for different electrode areas. For electrodes placed face to face we found that noise and resistance increase when electrode area decreases. For electrodes placed on the body surface, "noise" power spectrum density (probably EMG) increases when electrode area increases. We also calculated the Probability Density Function (PDF) of electrode noise. We found that it looks gaussian for electrodes placed face to face, but not for electrodes placed on the body surface.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
表面生物电位测量中的电极接触噪声
皮肤电极界面噪声是高分辨率表面生物电位记录的重要内容。为了确定它的作用,我们进行了几次测量。首先,我们测量了几对面对面放置的Ag/AgQ电极的噪声功率谱密度,然后放置在身体表面(腿和手臂)。我们同时测量了这些电极对的阻抗,并计算了与它们的实部相关的热噪声。我们发现低频噪声比高频噪声大,功率谱密度以1/f的速度增加。我们还发现计算的热噪声比测量的电极噪声低。接下来,我们测量了不同电极区域的噪声和阻抗。对于面对面放置的电极,我们发现随着电极面积的减小,噪声和电阻增大。对于放置在体表的电极,“噪声”功率谱密度(可能是肌电图)随着电极面积的增加而增加。计算了电极噪声的概率密度函数(PDF)。我们发现面对面放置的电极看起来是高斯分布,但放置在身体表面的电极则不是高斯分布。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Welcome Message from the Organizing Committee A model of injury potential for myelinated nerve fiber Plenary keynote lecture President's welcome message Images of Electrocardiographic Imaging
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1