{"title":"A realization of super-fast scanning power supply for proton therapy","authors":"Jiazhe Li, Rui Li, Huanting Lu","doi":"10.1587/elex.18.20210501","DOIUrl":null,"url":null,"abstract":"This paper introduces a method to achieve fast scanning power supply for Proton Therapy. To meet the specific needs of this power supply, a new structure consists of dynamic module (high voltage part) and static module (low voltage part) has been designed and tested. The simulation results and actual waveform were listed in this paper. The results showed that the current rising slope can reach 90kA/s.","PeriodicalId":13437,"journal":{"name":"IEICE Electron. Express","volume":"1 1","pages":"20210501"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEICE Electron. Express","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1587/elex.18.20210501","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
This paper introduces a method to achieve fast scanning power supply for Proton Therapy. To meet the specific needs of this power supply, a new structure consists of dynamic module (high voltage part) and static module (low voltage part) has been designed and tested. The simulation results and actual waveform were listed in this paper. The results showed that the current rising slope can reach 90kA/s.