Application value of magnetic resonance 3D-SPACE full volume scanning sequence in direct imaging of lower extremity deep venous thrombosis

Jilin Nie
{"title":"Application value of magnetic resonance 3D-SPACE full volume scanning sequence in direct imaging of lower extremity deep venous thrombosis","authors":"Jilin Nie","doi":"10.3760/CMA.J.ISSN.1008-1372.2019.12.014","DOIUrl":null,"url":null,"abstract":"Objective \nTo evaluate the capability of magnetic resonance imaging 3D-sampling perfection with application optimize contrasts using different flip angle evolution (MRI 3D-SPACE) full-volume scanning sequence to detect lower extremity deep vein thrombosis (LEDVT), and to explore the characteristics of the images of LEDVT and surrounding soft tissue at different stages. \n \n \nMethods \nMRI 3D-SPACE full volume T1WI and T2WI scanning of lower limb were performed on 45 patients diagnosed of LEDVT via digital subtraction angiography (DSA) by Siemens Prisma 3T MRI scanner. 43 cases completed scanning. The location and quantity of vascular segment lesions were recorded, of which the diagnostic sensitivity, specificity and accuracy were calculated comparing to the gold standard of DSA results. The characteristics of magnetic resonance imaging signals of thrombus and thrombus surrounding soft tissue at different times were summarized. \n \n \nResults \nAccording to the location of thrombosis, there were 15 cases of central type, 17 cases of peripheral type and 11 cases of mixed type. In all imaging levels, 272 segments of embolic vessels were found in 3D-space T1WI and 261 segments in 3D-space T2WI. Compared with the DSA results, the sensitivity, specificity, and accuracy of the two groups in the diagnosis of deep venous thrombosis were: T1 WI: Common iliac vein 97.10%, 97.90%, 98.20%; External iliac vein 96.80%, 97.60%, 97.10%; Internal iliac vein 96.40%, 97.20%, 97.20%; Femoral vein 98.10%, 98.20%, 98.50%; Popliteal vein 94.40%, 93.50%, 93.70%; The anterior tibial vein is 94.30%, 88.50% and 90.40%. The posterior tibial vein is 93.40%, 85.20%, 85.30%. T2WI: Common iliac vein 97.10%, 97.90%, 97.10%; External iliac vein 96.50%, 96.20%, 96.00%; Internal iliac vein 96.20%, 96.60%, 96.20%; Femoral vein 98.10%, 98.20%, 98.50%; Popliteal vein 93.40%, 92.50%, 92.70%; The anterior tibial vein is 90.30%, 86.50% and 85.40%; Posterior tibial vein 89.40%, 84.20%, 83.30%. According to the onset time, 17 cases in the acute phase, the thrombus was equal or slightly higher and uneven signal in the two groups of sequences, the perivascular and muscle tissue gap signal increased; 21 cases in the subacute phase, the thrombus was high in both groups signal; 5 cases in the chronic phase, the thrombus showed equal or slightly lower signal in the two groups of sequences, the blood vessel wall was irregular, and the surrounding soft tissue was mild or no obvious edema. \n \n \nConclusions \nMRI 3D-space sequence full volume scanning technology has high sensitivity, high specificity and high accuracy for the diagnosis of lower extremity deep vein thrombosis, and can reflect the characteristics of thrombus and soft tissue signal around thrombus in different periods. It has high application value for the early detection of thrombosis and the stage diagnosis of the course of disease. \n \n \nKey words: \nMagnetic resonance imaging; Imaging, three-dimensional; Lower extremity deep vein thrombosis","PeriodicalId":15276,"journal":{"name":"中国医师杂志","volume":"718 1","pages":"1813-1816"},"PeriodicalIF":0.0000,"publicationDate":"2019-12-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医师杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.1008-1372.2019.12.014","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective To evaluate the capability of magnetic resonance imaging 3D-sampling perfection with application optimize contrasts using different flip angle evolution (MRI 3D-SPACE) full-volume scanning sequence to detect lower extremity deep vein thrombosis (LEDVT), and to explore the characteristics of the images of LEDVT and surrounding soft tissue at different stages. Methods MRI 3D-SPACE full volume T1WI and T2WI scanning of lower limb were performed on 45 patients diagnosed of LEDVT via digital subtraction angiography (DSA) by Siemens Prisma 3T MRI scanner. 43 cases completed scanning. The location and quantity of vascular segment lesions were recorded, of which the diagnostic sensitivity, specificity and accuracy were calculated comparing to the gold standard of DSA results. The characteristics of magnetic resonance imaging signals of thrombus and thrombus surrounding soft tissue at different times were summarized. Results According to the location of thrombosis, there were 15 cases of central type, 17 cases of peripheral type and 11 cases of mixed type. In all imaging levels, 272 segments of embolic vessels were found in 3D-space T1WI and 261 segments in 3D-space T2WI. Compared with the DSA results, the sensitivity, specificity, and accuracy of the two groups in the diagnosis of deep venous thrombosis were: T1 WI: Common iliac vein 97.10%, 97.90%, 98.20%; External iliac vein 96.80%, 97.60%, 97.10%; Internal iliac vein 96.40%, 97.20%, 97.20%; Femoral vein 98.10%, 98.20%, 98.50%; Popliteal vein 94.40%, 93.50%, 93.70%; The anterior tibial vein is 94.30%, 88.50% and 90.40%. The posterior tibial vein is 93.40%, 85.20%, 85.30%. T2WI: Common iliac vein 97.10%, 97.90%, 97.10%; External iliac vein 96.50%, 96.20%, 96.00%; Internal iliac vein 96.20%, 96.60%, 96.20%; Femoral vein 98.10%, 98.20%, 98.50%; Popliteal vein 93.40%, 92.50%, 92.70%; The anterior tibial vein is 90.30%, 86.50% and 85.40%; Posterior tibial vein 89.40%, 84.20%, 83.30%. According to the onset time, 17 cases in the acute phase, the thrombus was equal or slightly higher and uneven signal in the two groups of sequences, the perivascular and muscle tissue gap signal increased; 21 cases in the subacute phase, the thrombus was high in both groups signal; 5 cases in the chronic phase, the thrombus showed equal or slightly lower signal in the two groups of sequences, the blood vessel wall was irregular, and the surrounding soft tissue was mild or no obvious edema. Conclusions MRI 3D-space sequence full volume scanning technology has high sensitivity, high specificity and high accuracy for the diagnosis of lower extremity deep vein thrombosis, and can reflect the characteristics of thrombus and soft tissue signal around thrombus in different periods. It has high application value for the early detection of thrombosis and the stage diagnosis of the course of disease. Key words: Magnetic resonance imaging; Imaging, three-dimensional; Lower extremity deep vein thrombosis
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
磁共振3D-SPACE全容积扫描序列在下肢深静脉血栓直接成像中的应用价值
目的评价不同翻角演化(MRI 3D-SPACE)全容积扫描序列对下肢深静脉血栓形成(LEDVT)检测的磁共振成像3d采样完美性和应用优化对比能力,探讨LEDVT及周围软组织不同阶段的图像特征。方法对45例经Siemens Prisma 3T MRI数字减影血管造影(DSA)诊断为LEDVT的患者行MRI 3D-SPACE下肢全容积T1WI、T2WI扫描。完成扫描43例。记录血管段病变的位置和数量,并与DSA结果的金标准进行比较,计算诊断的敏感性、特异性和准确性。总结了血栓及血栓周围软组织不同时间的磁共振成像信号特征。结果根据血栓形成部位不同,中心型15例,外周型17例,混合型11例。在所有成像水平上,3d空间T1WI发现栓塞血管272段,3d空间T2WI发现栓塞血管261段。与DSA结果比较,两组诊断深静脉血栓的敏感性、特异性和准确性分别为:T1 WI:髂总静脉97.10%、97.90%、98.20%;髂外静脉96.80%、97.60%、97.10%;髂内静脉96.40%、97.20%、97.20%;股静脉98.10%,98.20%,98.50%;腘静脉94.40%,93.50%,93.70%;胫前静脉占94.30%、88.50%、90.40%。胫后静脉分别为93.40%、85.20%、85.30%。T2WI:髂总静脉97.10%、97.90%、97.10%;髂外静脉96.50%、96.20%、96.00%;髂内静脉96.20%、96.60%、96.20%;股静脉98.10%,98.20%,98.50%;腘静脉93.40%、92.50%、92.70%;胫前静脉分别为90.30%、86.50%和85.40%;胫后静脉89.40%,84.20%,83.30%。根据发病时间,17例在急性期,两组序列中血栓大小相等或略高且信号不均匀,血管周围和肌肉组织间隙信号增高;亚急性期21例,两组血栓信号均较高;慢性期5例,两组序列血栓信号相等或稍低,血管壁不规则,周围软组织轻度或无明显水肿。结论MRI三维空间序列全容积扫描技术对下肢深静脉血栓的诊断具有高灵敏度、高特异性和高准确性,能反映血栓及不同时期血栓周围软组织信号的特征。对血栓的早期发现和病程的分期诊断具有很高的应用价值。关键词:磁共振成像;三维成像;下肢深静脉血栓
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
中国医师杂志
中国医师杂志 Medicine-Medicine (all)
CiteScore
0.10
自引率
0.00%
发文量
20937
期刊介绍:
期刊最新文献
Effect of sivelestat sodium on early inflammatory reaction and pulmonary edema in rats with smoke inhalation induced lung injury Effect of CEA on neurological function and coagulation function in patients with ischemic stroke Intraoperative real-time imaging-guided laparoscopic liver resection Application progress of laparoscopy in the treatment of liver tumors Analysis of risk of endobronchial biopsy-induced bleeding in different locations of lung cancer
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1