{"title":"Determination of trace amounts of copper using a flow-through spectrophotometric sensor","authors":"J. M. Colsa Herrera, F. Sanchez Rojas, C. Bosch Ojeda, A. García de Torres, J. M. Cano Pavón","doi":"10.1002/1098-2728(2000)12:5<241::AID-LRA4>3.0.CO;2-W","DOIUrl":null,"url":null,"abstract":"<p>A flow-through spectrophotometric sensor for determination of copper based on retention of Cu(I)-1,2 ciclohexanedione thiosemicarbazone complex on C<sub>18</sub> bonded phase beads packed in a flow-cell was developed. The method is highly selective for copper; it features a detection limit of 50 ng mL<sup>−1</sup> and a linear range from at least 0.1 to 10 μg mL<sup>−1</sup>. The method is subject to very few interferences and is more sensitive than the batch procedure. The proposed method was applied to the determination of copper in catalysts and alloys. © 2000 John Wiley & Sons, Inc. Lab Robotics and Automation 12:241–245, 2000</p>","PeriodicalId":100863,"journal":{"name":"Laboratory Robotics and Automation","volume":"12 5","pages":"241-245"},"PeriodicalIF":0.0000,"publicationDate":"2000-12-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/1098-2728(2000)12:5<241::AID-LRA4>3.0.CO;2-W","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory Robotics and Automation","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/1098-2728%282000%2912%3A5%3C241%3A%3AAID-LRA4%3E3.0.CO%3B2-W","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
A flow-through spectrophotometric sensor for determination of copper based on retention of Cu(I)-1,2 ciclohexanedione thiosemicarbazone complex on C18 bonded phase beads packed in a flow-cell was developed. The method is highly selective for copper; it features a detection limit of 50 ng mL−1 and a linear range from at least 0.1 to 10 μg mL−1. The method is subject to very few interferences and is more sensitive than the batch procedure. The proposed method was applied to the determination of copper in catalysts and alloys. © 2000 John Wiley & Sons, Inc. Lab Robotics and Automation 12:241–245, 2000
用流动分光光度传感器测定痕量铜
建立了一种基于Cu(I)-1,2环己二酮硫代氨基脲配合物在流动池中C18键合相珠上的保留量测定铜的流动分光光度传感器。该方法对铜有很高的选择性;其检测限为50 ng mL - 1,线性范围至少为0.1至10 μg mL - 1。该方法受干扰很小,比批处理法更灵敏。该方法可用于催化剂和合金中铜的测定。©2000 John Wiley &儿子,Inc。实验室机器人与自动化,2000:241 - 245
本文章由计算机程序翻译,如有差异,请以英文原文为准。