The Influence of Different Implant Abutment Sizes on Fracture Resistance of Two Monolithic Ceramic Crown Systems

Enaam Y. Abdel-Rahman, Sahar Y. Abdel-Aziz, Zainab El Sharkawy
{"title":"The Influence of Different Implant Abutment Sizes on Fracture Resistance of Two Monolithic Ceramic Crown Systems","authors":"Enaam Y. Abdel-Rahman, Sahar Y. Abdel-Aziz, Zainab El Sharkawy","doi":"10.21608/adjg.2022.74968.1360","DOIUrl":null,"url":null,"abstract":"Purpose: to study the effect of different implant abutment sizes on fracture resistance of two ceramic monolithic crown systems. Material and Methods: Research Ethics committee approval of faculty of Dental Medicine for Girls was obtained (REC18-065). Forty (40) Titanium implant supported abutments were fabricated and divided into2 groups (n=20 each) according to abutment diameter; Group (1): implant abutment diameter 4mm and Group (2): implant supported abutment diameter 4.5mm. Samples of each group were divided into 2 sub-groups (n=10 each) according to type of monolithic ceramic crowns. Subgroup(A): zirconia ceramic crowns, Subgroup(B): vita suprinity crowns. All crowns were cemented to implant abutments by Rely X resin cement, then subjected to thermal cycling and seated on a universal testing machine and subjected to fracture resistance test. Data were statistically analyzed . Results: It was found that the highest mean ± SD values of fracture resistance were recorded for large diameter Zirconia subgroup (1032.2±174.78 N) followed by large diameter Vita suprinity subgroup mean ± SD values (935.7±83.71 N) then small diameter Zirconia subgroup (930.3±69.53 N), meanwhile the lowest mean ± SD values were recorded with small diameter Vita suprinity (815.9±101.52 N). Conclusion: The mean fracture values of zirconia were higher than that of vita suprinity crowns, while the crowns were supported on large implant abutment diameter had higher fracture resistance than that supported on smaller implant abutment diameter.","PeriodicalId":7493,"journal":{"name":"Al-Azhar Dental Journal for Girls","volume":"67 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Al-Azhar Dental Journal for Girls","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.21608/adjg.2022.74968.1360","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: to study the effect of different implant abutment sizes on fracture resistance of two ceramic monolithic crown systems. Material and Methods: Research Ethics committee approval of faculty of Dental Medicine for Girls was obtained (REC18-065). Forty (40) Titanium implant supported abutments were fabricated and divided into2 groups (n=20 each) according to abutment diameter; Group (1): implant abutment diameter 4mm and Group (2): implant supported abutment diameter 4.5mm. Samples of each group were divided into 2 sub-groups (n=10 each) according to type of monolithic ceramic crowns. Subgroup(A): zirconia ceramic crowns, Subgroup(B): vita suprinity crowns. All crowns were cemented to implant abutments by Rely X resin cement, then subjected to thermal cycling and seated on a universal testing machine and subjected to fracture resistance test. Data were statistically analyzed . Results: It was found that the highest mean ± SD values of fracture resistance were recorded for large diameter Zirconia subgroup (1032.2±174.78 N) followed by large diameter Vita suprinity subgroup mean ± SD values (935.7±83.71 N) then small diameter Zirconia subgroup (930.3±69.53 N), meanwhile the lowest mean ± SD values were recorded with small diameter Vita suprinity (815.9±101.52 N). Conclusion: The mean fracture values of zirconia were higher than that of vita suprinity crowns, while the crowns were supported on large implant abutment diameter had higher fracture resistance than that supported on smaller implant abutment diameter.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
不同种植基台尺寸对两种整体陶瓷冠系统抗断裂性能的影响
目的:研究不同种植基台尺寸对两种陶瓷全冠系统抗断裂性能的影响。材料和方法:获得女口腔医学院伦理委员会批准(REC18-065)。制作钛种植体支撑基牙40个,根据基牙直径分为2组,每组20个;组(1):种植体基台直径4mm;组(2):种植体支撑基台直径4.5mm。每组样本按整体瓷冠类型分为2个亚组,每组n=10。亚组(A):氧化锆陶瓷冠;亚组(B): vita superity冠。所有冠用Rely X树脂水泥与种植基台粘合,热循环后置于万能试验机上进行抗断裂测试。对数据进行统计学分析。结果:大直径Zirconia亚组的骨折阻力平均值±SD值最高(1032.2±174.78 N),其次是大直径Vita超优亚组的平均值±SD值(935.7±83.71 N),其次是小直径Zirconia亚组的平均值±SD值(930.3±69.53 N),而小直径Vita超优亚组的平均值±SD值最低(815.9±101.52 N)。氧化锆牙冠的平均断裂值高于优质牙冠,而种植基牙直径大的牙冠比种植基牙直径小的牙冠具有更高的抗断裂能力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Efficacy of Aloe vera in Treatment of Experimentally Induced Oral Lichenoid Reaction Evaluation of Local Versus Intravenous Administration of Adipose Tissue-Derived Stem Cells on Regeneration of Oral Epithelium of Chemotherapy Treated Rats The Influence of Different Implant Abutment Sizes on Fracture Resistance of Two Monolithic Ceramic Crown Systems Effect of Irradiation with 980-nm Diode Laser on the Microhardness of Young and Old Root Canal Dentin after Treatment with Chemical Solutions Evaluation of Occlusal Pressure Pattern and Dimensional Changes of the Lined Dentures Subjected to Microwave Disinfection Protocol
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1