Quality Control of Conventional X-Ray Tube in Three Tertiary Hospitals in South-East, Nigeria

Nworah Nf, C. Nzotta, H. Chiegwu, Oyekunle Eo, Ikegwuonu Nc, Ugwuanyi Dc
{"title":"Quality Control of Conventional X-Ray Tube in Three Tertiary Hospitals in South-East, Nigeria","authors":"Nworah Nf, C. Nzotta, H. Chiegwu, Oyekunle Eo, Ikegwuonu Nc, Ugwuanyi Dc","doi":"10.48153/jrrs.v32i1.223078","DOIUrl":null,"url":null,"abstract":"\n \n \n \nBackground:Quality control of conventional x-ray tube ensures that the patient integral radiation dose is minimized and image quality is improved, by controlling the x-ray beam to reduce scatter radiation. \n Objective:To assess x-ray tubes for half value layer (HVL), x-ray field and light field congruence using known standard. \nMethodology:The HVLs were measured using calibrated, non-invasive, digital multifunctional detector meter that incorporate computer output. The detector was positioned at the center of the collimated beam axis with focus-to-image distance (FID) of 100 cm. Tube potentials of 80 and 100 kVp were selected, and used to make exposures. The corresponding HVLs were then recorded. Also, x-ray field and light field congruence were measured by placing 18 cm x 24 cm cassette loaded with film at FID of 100 cm. The collimator light was in ON position and metal markers were used to delineate the periphery of the light field. Misalignment was calculated from the developed radiographs using standard formula. \nResults:The HVL ranged from 3.40 to 4.4mmAl. Also, the sum of the misalignment in both orthogonal directions ranged from 7.21 to 9.70 % of FID in all three centers. \nConclusion:The HVL were within standard limit at 80 and 100 kVp in all the centers studied. However, x-ray field and light field were grossly misaligned. \n  \n \n \n \n \n ","PeriodicalId":16919,"journal":{"name":"Journal of Radiography and Radiation Sciences","volume":"4 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-03-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiography and Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.48153/jrrs.v32i1.223078","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background:Quality control of conventional x-ray tube ensures that the patient integral radiation dose is minimized and image quality is improved, by controlling the x-ray beam to reduce scatter radiation.  Objective:To assess x-ray tubes for half value layer (HVL), x-ray field and light field congruence using known standard. Methodology:The HVLs were measured using calibrated, non-invasive, digital multifunctional detector meter that incorporate computer output. The detector was positioned at the center of the collimated beam axis with focus-to-image distance (FID) of 100 cm. Tube potentials of 80 and 100 kVp were selected, and used to make exposures. The corresponding HVLs were then recorded. Also, x-ray field and light field congruence were measured by placing 18 cm x 24 cm cassette loaded with film at FID of 100 cm. The collimator light was in ON position and metal markers were used to delineate the periphery of the light field. Misalignment was calculated from the developed radiographs using standard formula. Results:The HVL ranged from 3.40 to 4.4mmAl. Also, the sum of the misalignment in both orthogonal directions ranged from 7.21 to 9.70 % of FID in all three centers. Conclusion:The HVL were within standard limit at 80 and 100 kVp in all the centers studied. However, x-ray field and light field were grossly misaligned.    
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
尼日利亚东南部三所三级医院常规x射线管质量控制
背景:传统x射线管的质量控制是通过控制x射线束来减少散射辐射,以保证患者整体辐射剂量的最小化和图像质量的提高。目的:用已知标准评价x射线管半值层(HVL)、x光场和光场一致性。方法:hvl测量使用校准,非侵入性,数字式多功能检测器,包括计算机输出。探测器位于准直光束轴线的中心,聚焦像距离(FID)为100 cm。选择80和100 kVp的管电位进行曝光。然后记录相应的hvl。此外,通过将装有薄膜的18 cm x 24 cm盒式磁带放置在100 cm的FID处,测量x射线场和光场一致性。准直器光处于ON位置,用金属标记圈定光场外围。用标准公式从显影的x线片中计算不对准。结果:HVL范围为3.40 ~ 4.4mmAl。此外,在两个正交方向上,三个中心的误差总和为FID的7.21 ~ 9.70%。结论:各中心HVL均在80、100 kVp的标准范围内。然而,x射线场和光场严重错位。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
CT features, Pathological findings and Clinical data of Follicular Dendritic Cell Sarcoma in the Mediastinum: Case Report and Literature Review KNOWLEDGE AND PRACTICE OF IMAGING OF CHILDREN WITH SUSPECTED CASES OF NON-ACCIDENTAL INJURIES AMONG RADIOGRAPHERS WITHIN MAIDUGURI METROPOLIS, NORTHEASTERN NIGERIA Effective doses in chest and abdominal radiography following the ICRP recommendations of 1991 and 2007 in a regional hospital. Investigation of the Effect of Solar Panel Temperature on Power Output Efficiency in Brass, Nigeria. An Assessment of the Lead rubber aprons in Radiodiagnostic centres in South-South Nigeria
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1