{"title":"The Application of Simple and Easy to Implement Decoupling Pulse Scheme Combinations to Effect Decoupling of Large Values with Reduced Artifacts","authors":"K. Klika","doi":"10.1155/2014/289638","DOIUrl":null,"url":null,"abstract":"Of the various problems in decoupling one nucleus type from another using standard decoupling pulse schemes for broadband decoupling, a particular challenge is to effect full, artifact-free decoupling when the size of the coupling constant is very large. Herein it is demonstrated that 1H decoupling of the 31P NMR spectrum of diethyl phosphonate HP(=O)(OCH2CH3)2 can be accomplished with reduced artifacts despite the large value of 693 Hz by using a combination of decoupling pulse schemes involving continuous-wave (CW) irradiation and either adiabatic-pulse decoupling (APD), MPF decoupling, or traditional composite-pulse decoupling (CPD) schemes such as WALTZ or GARP. The considered strategy is simple, efficient, and easy to implement on most instruments. The best result was attained for a combination of CW and CPD using GARP with a standard pulse width of 60 μs. Altogether, the advantages of the methodology include low power requirements, complete decoupling, tolerance of a range of large values, large bandwidth for normal-sized values, and the suppression of sidebands.","PeriodicalId":14329,"journal":{"name":"International Journal of Spectroscopy","volume":"49 1","pages":"1-9"},"PeriodicalIF":0.0000,"publicationDate":"2014-04-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"13","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Spectroscopy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2014/289638","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 13
Abstract
Of the various problems in decoupling one nucleus type from another using standard decoupling pulse schemes for broadband decoupling, a particular challenge is to effect full, artifact-free decoupling when the size of the coupling constant is very large. Herein it is demonstrated that 1H decoupling of the 31P NMR spectrum of diethyl phosphonate HP(=O)(OCH2CH3)2 can be accomplished with reduced artifacts despite the large value of 693 Hz by using a combination of decoupling pulse schemes involving continuous-wave (CW) irradiation and either adiabatic-pulse decoupling (APD), MPF decoupling, or traditional composite-pulse decoupling (CPD) schemes such as WALTZ or GARP. The considered strategy is simple, efficient, and easy to implement on most instruments. The best result was attained for a combination of CW and CPD using GARP with a standard pulse width of 60 μs. Altogether, the advantages of the methodology include low power requirements, complete decoupling, tolerance of a range of large values, large bandwidth for normal-sized values, and the suppression of sidebands.