{"title":"Analytical performances of a glycated albumin assay that is traceable to standard reference materials and reference range determination.","authors":"Xinran Tao, Ryosuke Koguma, Yoko Nagai, Takuji Kohzuma","doi":"10.1002/jcla.24509","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Glycated albumin (GA) is an intermediate-term marker for monitoring glycemic control (preceding 2-3 weeks) in patients with diabetes mellitus. We evaluated the performance of Lucica Glycated Albumin-L, a new GA assay that is traceable to standard reference materials and determined the reference range in healthy subjects without diabetes.</p><p><strong>Methods: </strong>The performance and reference range studies were conducted in accordance with Clinical and Laboratory Standards Institute (CLSI) Guidelines. The traceability was established using reference material recommended by the Japan Society of Clinical Chemistry (JSCC).</p><p><strong>Results: </strong>The coefficient of variation (CV) of overall repeatability, within-laboratory precision, and overall reproducibility values of GA values were not more than 2.6%, 3.3%, and 1.6%, respectively, among laboratories. The GA values showed good linearity from 173 to 979 mmol/mol (9.4%-54.9%) across the assay range. The GA reference range in 262 healthy subjects was between 183 and 259 mmol/mol (9.9%-14.2%) while that of subjects with diabetes was 217-585 mmol/mol (11.8-32.6%). The reagent was stable for 2 months on the bench at room temperature. The limits of blank, detection, and qualification were 6.9, 7.9, and 9.7 μmol/L for GA concentration, and 3.8, 7.0, and 21.8 μmol/L for albumin concentration, respectively. Hemoglobin slightly affected the assay, while other classical interfering substances had no significant impact.</p><p><strong>Conclusions: </strong>The present GA assay shows comparable performance to current clinical assays and could be used for intermediate-term monitoring of glycemic control in diabetes patients.</p>","PeriodicalId":49817,"journal":{"name":"Microelectronics International","volume":"4 1","pages":"e24509"},"PeriodicalIF":0.7000,"publicationDate":"2022-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9280011/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Microelectronics International","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jcla.24509","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/5/20 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"ENGINEERING, ELECTRICAL & ELECTRONIC","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Glycated albumin (GA) is an intermediate-term marker for monitoring glycemic control (preceding 2-3 weeks) in patients with diabetes mellitus. We evaluated the performance of Lucica Glycated Albumin-L, a new GA assay that is traceable to standard reference materials and determined the reference range in healthy subjects without diabetes.
Methods: The performance and reference range studies were conducted in accordance with Clinical and Laboratory Standards Institute (CLSI) Guidelines. The traceability was established using reference material recommended by the Japan Society of Clinical Chemistry (JSCC).
Results: The coefficient of variation (CV) of overall repeatability, within-laboratory precision, and overall reproducibility values of GA values were not more than 2.6%, 3.3%, and 1.6%, respectively, among laboratories. The GA values showed good linearity from 173 to 979 mmol/mol (9.4%-54.9%) across the assay range. The GA reference range in 262 healthy subjects was between 183 and 259 mmol/mol (9.9%-14.2%) while that of subjects with diabetes was 217-585 mmol/mol (11.8-32.6%). The reagent was stable for 2 months on the bench at room temperature. The limits of blank, detection, and qualification were 6.9, 7.9, and 9.7 μmol/L for GA concentration, and 3.8, 7.0, and 21.8 μmol/L for albumin concentration, respectively. Hemoglobin slightly affected the assay, while other classical interfering substances had no significant impact.
Conclusions: The present GA assay shows comparable performance to current clinical assays and could be used for intermediate-term monitoring of glycemic control in diabetes patients.
期刊介绍:
Microelectronics International provides an authoritative, international and independent forum for the critical evaluation and dissemination of research and development, applications, processes and current practices relating to advanced packaging, micro-circuit engineering, interconnection, semiconductor technology and systems engineering. It represents a current, comprehensive and practical information tool. The Editor, Dr John Atkinson, welcomes contributions to the journal including technical papers, research papers, case studies and review papers for publication. Please view the Author Guidelines for further details.
Microelectronics International comprises a multi-disciplinary study of the key technologies and related issues associated with the design, manufacture, assembly and various applications of miniaturized electronic devices and advanced packages. Among the broad range of topics covered are:
• Advanced packaging
• Ceramics
• Chip attachment
• Chip on board (COB)
• Chip scale packaging
• Flexible substrates
• MEMS
• Micro-circuit technology
• Microelectronic materials
• Multichip modules (MCMs)
• Organic/polymer electronics
• Printed electronics
• Semiconductor technology
• Solid state sensors
• Thermal management
• Thick/thin film technology
• Wafer scale processing.