Implementation of carbon fibre treatment couches in the XiO® and Monaco® Treatment Planning Systems

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Polish Journal of Medical Physics and Engineering Pub Date : 2020-12-01 DOI:10.2478/pjmpe-2020-0025
Christoffel Jacobus van Reenen, C. Trauernicht
{"title":"Implementation of carbon fibre treatment couches in the XiO® and Monaco® Treatment Planning Systems","authors":"Christoffel Jacobus van Reenen, C. Trauernicht","doi":"10.2478/pjmpe-2020-0025","DOIUrl":null,"url":null,"abstract":"Abstract Purpose: Carbon fibre treatment couches on linear accelerators provide a strong, rigid framework for patient support. Patient safety is a priority, therefore the dosimetric properties of treatment couches need to be accurately incorporated in treatment plans, to minimize differences between planned and delivered dose. This study aims to determine the attenuation effect of treatment couches for 3-D Conformal Radiotherapy (3-D CRT) and to validate the implementation thereof in the XiO and Monaco treatment planning systems (TPS). Material and methods: Attenuation measurements were performed on the ELEKTA Connexion couches of the ELEKTA Precise and Synergy-Agility linear accelerators. Measurements were made at 10° intervals in RMI-457 Solid water (30 cm x 30 cm x 30 cm) using a PTW Farmer-type ionization chamber (TW30013) positioned at the accelerator’s isocentre. The percentage attenuation was calculated as the ratio of the electrometer readings for parallel-opposed fields. The Computed Tomography (CT) data sets of the set-ups were obtained on a Philips Big Bore 16-slice CT scanner and exported to the TPS. The individual couch structures were delineated and electron density (ED) values were assigned using the commissioned CT-to-ED curve. Test treatment plans were generated with 100MU per field at 10° gantry intervals. Results: The percentage attenuation was determined to be within 2% and 3% for beams perpendicular to the couch surface for XiO and Monaco, respectively. The maximum attenuation was observed for oblique fields which was significantly higher than the manufacturer specified values. TPS validation showed an agreement to 1% for XiO and Monaco. At extreme oblique angles, both planning systems overestimated this effect up to a maximum of 4%. Conclusions: Couch attenuation differs significantly with gantry angle and beam energy. As a result, the treatment couch models should be included in all treatment planning calculations.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":null,"pages":null},"PeriodicalIF":0.7000,"publicationDate":"2020-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2020-0025","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Abstract Purpose: Carbon fibre treatment couches on linear accelerators provide a strong, rigid framework for patient support. Patient safety is a priority, therefore the dosimetric properties of treatment couches need to be accurately incorporated in treatment plans, to minimize differences between planned and delivered dose. This study aims to determine the attenuation effect of treatment couches for 3-D Conformal Radiotherapy (3-D CRT) and to validate the implementation thereof in the XiO and Monaco treatment planning systems (TPS). Material and methods: Attenuation measurements were performed on the ELEKTA Connexion couches of the ELEKTA Precise and Synergy-Agility linear accelerators. Measurements were made at 10° intervals in RMI-457 Solid water (30 cm x 30 cm x 30 cm) using a PTW Farmer-type ionization chamber (TW30013) positioned at the accelerator’s isocentre. The percentage attenuation was calculated as the ratio of the electrometer readings for parallel-opposed fields. The Computed Tomography (CT) data sets of the set-ups were obtained on a Philips Big Bore 16-slice CT scanner and exported to the TPS. The individual couch structures were delineated and electron density (ED) values were assigned using the commissioned CT-to-ED curve. Test treatment plans were generated with 100MU per field at 10° gantry intervals. Results: The percentage attenuation was determined to be within 2% and 3% for beams perpendicular to the couch surface for XiO and Monaco, respectively. The maximum attenuation was observed for oblique fields which was significantly higher than the manufacturer specified values. TPS validation showed an agreement to 1% for XiO and Monaco. At extreme oblique angles, both planning systems overestimated this effect up to a maximum of 4%. Conclusions: Couch attenuation differs significantly with gantry angle and beam energy. As a result, the treatment couch models should be included in all treatment planning calculations.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在XiO®和Monaco®治疗计划系统中实施碳纤维治疗沙发
目的:碳纤维治疗沙发的线性加速器提供了一个强大的,刚性的框架,为患者的支持。患者安全是一个优先事项,因此治疗躺椅的剂量学特性需要准确地纳入治疗计划,以尽量减少计划剂量和交付剂量之间的差异。本研究旨在确定3-D适形放疗(3-D CRT)治疗沙发的衰减效果,并验证其在XiO和Monaco治疗计划系统(TPS)中的实施。材料和方法:衰减测量在ELEKTA精密和协同-敏捷线性加速器的ELEKTA Connexion沙发上进行。在RMI-457固体水(30 cm x 30 cm x 30 cm)中以10°间隔进行测量,使用位于加速器等心的PTW farmer型电离室(TW30013)。衰减百分比计算为静电计读数的比例为平行对立场。设置的计算机断层扫描(CT)数据集在飞利浦大孔16层CT扫描仪上获得,并导出到TPS。利用委托的ct -ED曲线描绘了单个沙发结构,并分配了电子密度(ED)值。每个地块100亩,10°龙门间隔的试验处理方案。结果:在XiO和Monaco,垂直于沙发表面的梁的衰减百分比分别在2%和3%以内。在斜场中观察到的最大衰减明显高于制造商规定的值。TPS验证显示XiO和Monaco的协议为1%。在极端倾斜角度下,两个规划系统都高估了这种影响,最大可达4%。结论:基底衰减随龙门架角度和光束能量有显著差异。因此,治疗沙发模型应包括在所有的治疗计划计算。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
期刊最新文献
Parametrization of subsegmental infarcts using high spatial resolution 2DSTE and synthetic ultrasonic data Automated differential diagnostics of respiratory diseases using an electronic stethoscope Evaluating the impact of anatomical changes on dose distributions in head and neck cancer Comparing eDQE and eNEQ metrics – is there an alternative approach to assessing image quality in digital mammography? Vascular stiffness in cold pressor test hyper-reactors
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1