Investigation of repulsion effects during lithotripsy

A. Rühm, M. Baumgartl, D. Steigenhöfer, T. Pongratz, M. Bader, W. Khoder, R. Sroka
{"title":"Investigation of repulsion effects during lithotripsy","authors":"A. Rühm, M. Baumgartl, D. Steigenhöfer, T. Pongratz, M. Bader, W. Khoder, R. Sroka","doi":"10.1515/plm-2013-0054","DOIUrl":null,"url":null,"abstract":"Abstract Background: Stone therapy has been developing towards minimally invasive endourological techniques. This includes endoscopic lithotripsy techniques to reduce the urolith size prior to removal. In this context, stone repulsion constitutes an undesired effect of the mechanical impact on the stones and should therefore be minimized to facilitate the intervention and to improve its outcome. Objectives: To compare the repulsion characteristics of two stand-alone, handheld lithotripsy devices (LithoBreaker® and StoneBreaker™) and one Ho:YAG laser using dedicated test models. Materials and methods: Tests with single-use mechanical probes (1 and 2 mm in diameter) as well as Ho:YAG laser pulses delivered by optical fibers with core diameters of 230, 365 and 600 μm were conducted in underwater set-ups by measuring displacements of non-breakable spherical test masses (steel balls having diameters of 3–5 mm and masses of 0.12, 0.27 and 0.52 g respectively; and a lead ball with a diameter of 5 mm and a mass of 0.72 g). The tests were carried out in an acrylic glass tube with an inner diameter of 12 mm used as an ureter model and on a pendulum set-up. Additionally, hard and soft dampers were compared in case of the LithoBreaker®. Mean and standard deviations of the displacements and related quantities were evaluated for all sets of experiments and statistical analyses were performed (Student’s t-test, Mann-Whitney U-test, Shapiro-Wilk test) to test the influence of device type and device parameters. Results: The repulsion distance in the ureter model using the 1-mm probes was 19.4±1.9 cm with hard and 21.5±2.7 cm with soft damper for the LithoBreaker® device and 20.3±3.0 cm for the StoneBreaker™ device (with standard damper). The repulsion distance using the 2-mm probes was 14.8±1.3 cm with hard and 20.5±2.5 cm with soft damper for the LithoBreaker® and 17.1±2.0 cm for the StoneBreaker™. In pendulum tests, the 2-mm probes again resulted in smaller displacements than the 1-mm probes, except for the LithoBreaker® with soft damper. For a fixed probe diameter, no statistically significant influence of the damper was observed in case of the LithoBreaker® (p>0.05). However, the StoneBreaker™ with 2-mm probes produced smaller repulsions than the LithoBreaker® with 2-mm probes and hard damper. The pendulum tests with Ho:YAG laser pulses showed that the repulsion is proportional to the fiber core diameter and the laser pulse energy. The laser-induced momentum is about a factor 100 lower than the one generated with the mechanical lithotripsy devices. Conclusions: The electromechanical LithoBreaker® and the pneumatic StoneBreaker™ produced significantly different repulsions (p<0.05). The laser-induced momentum transfer is significantly lower compared to the mechanical lithotripsy devices (p<0.05). Zusammenfassung Hintergrund: Die Steintherapie hat sich zunehmend hin zu minimal-invasiven endourologischen Techniken entwickelt. Dies schließt endoskopische Steinzertrümmerungstechniken ein, die eingesetzt werden, um die Größe des Uroliths vor der Entfernung zu reduzieren. In diesem Zusammenhang stellt der Rückstoß des Steins einen unerwünschten Effekt des mechanischen Stoßes dar, der daher minimiert werden sollte, um den Eingriff zu erleichtern und seinen Erfolg zu verbessern. Zielsetzung: Vergleich der Rückstoß-Charakteristiken von zwei eigenständigen tragbaren Lithotripsie-Geräten (LithoBreaker® und StoneBreaker™) und einem Ho:YAG-Laser mit Hilfe spezieller Testmodelle. Materialien und Methoden: In Wasserumgebung wurden Tests mit mechanischen Einwegsonden (1 und 2 mm im Durchmesser) sowie mit über optische Fasern (Kerndurchmesser von 230, 365 und 600 μm) übertragenen Ho:YAG-Laserpulsen durchgeführt, indem die Auslenkungen von unzerbrechlichen kugelförmigen Testmassen gemessen wurden (Stahlkugeln mit Durchmessern von 3, 4 bzw. 5 mm und Massen von 0,12, 0,27 bzw. 0,52 g, sowie eine Bleikugel mit einem Durchmesser von 5 mm und einer Masse von 0,72 g). Die Tests wurden in einer Acrylglas-Röhre mit 12 mm Innendurchmesser, welche als Ureter-Modell diente, sowie an einem Pendelaufbau durchgeführt. Zusätzlich wurden im Falle des LithoBreaker® harte und weiche Dämpfer miteinander verglichen. Mittelwerte und Standardabweichungen der Auslenkungen und der damit zusammenhängenden Messgrößen wurden für alle Experimentserien berechnet, und statistische Analysen wurden durchgeführt (Student’s t-Test, Mann-Whitney U-Test, Shapiro-Wilk-Test), um den Einfluss des Gerätetyps und der Geräteparameter zu testen. Ergebnisse: Die Rückstoßdistanz im Ureter-Modell bei Verwendung der 1 mm-Sonden betrug 19,4±1,9 cm mit hartem und 21,5±2,7 cm mit weichem Dämpfer für das LithoBreaker®-Gerät und 20,3±3,0 cm für das StoneBreaker™-Gerät (mit Standard-Dämpfer). Die Rückstoßdistanz bei Verwendung der 2 mm-Sonden lag bei 14,8±1,3 cm mit hartem und 20,5±2,5 cm mit weichem Dämpfer für den LithoBreaker® und 17,1±2,0 cm für den StoneBreaker™. In den Pendelversuchen mit den 2 mm-Sonden wurden erneut kleinere Auslenkungen als mit den 1 mm-Sonden gemessen, mit Ausnahme des LithoBreaker® mit weichem Dämpfer. Bei festem Sondendurchmesser wurde im Falle des LithoBreaker® kein statistisch signifikanter Einfluss des Dämpfers beobachtet (p>0,05). Der StoneBreaker™ mit 2 mm-Sonden erzeugte jedoch kleinere Rückstöße als der LithoBreaker® mit 2 mm-Sonden und hartem Dämpfer. Die Pendeltests mit Ho:YAG-Laserpulsen zeigten, dass der Rückstoß sich proportional zum Faserkerndurchmesser und der Laserpulsenergie verhält. Der laser-induzierte Impuls ist etwa um den Faktor 100 kleiner als der, der mit den mechanischen Lithotripsie-Geräten erzeugt wurde. Schlussfolgerungen: Mit dem elektromechanischen LithoBreaker® und dem pneumatischen StoneBreaker™ werden signifikant unterschiedliche Rückstöße erzeugt (p<0,05). Der laser-induzierte Impulsübertrag ist signifikant kleiner als bei den mechanischen Lithotripsie-Geräten (p<0,05).","PeriodicalId":20126,"journal":{"name":"Photonics & Lasers in Medicine","volume":"75 1","pages":"47 - 55"},"PeriodicalIF":0.0000,"publicationDate":"2014-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Photonics & Lasers in Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/plm-2013-0054","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

Abstract Background: Stone therapy has been developing towards minimally invasive endourological techniques. This includes endoscopic lithotripsy techniques to reduce the urolith size prior to removal. In this context, stone repulsion constitutes an undesired effect of the mechanical impact on the stones and should therefore be minimized to facilitate the intervention and to improve its outcome. Objectives: To compare the repulsion characteristics of two stand-alone, handheld lithotripsy devices (LithoBreaker® and StoneBreaker™) and one Ho:YAG laser using dedicated test models. Materials and methods: Tests with single-use mechanical probes (1 and 2 mm in diameter) as well as Ho:YAG laser pulses delivered by optical fibers with core diameters of 230, 365 and 600 μm were conducted in underwater set-ups by measuring displacements of non-breakable spherical test masses (steel balls having diameters of 3–5 mm and masses of 0.12, 0.27 and 0.52 g respectively; and a lead ball with a diameter of 5 mm and a mass of 0.72 g). The tests were carried out in an acrylic glass tube with an inner diameter of 12 mm used as an ureter model and on a pendulum set-up. Additionally, hard and soft dampers were compared in case of the LithoBreaker®. Mean and standard deviations of the displacements and related quantities were evaluated for all sets of experiments and statistical analyses were performed (Student’s t-test, Mann-Whitney U-test, Shapiro-Wilk test) to test the influence of device type and device parameters. Results: The repulsion distance in the ureter model using the 1-mm probes was 19.4±1.9 cm with hard and 21.5±2.7 cm with soft damper for the LithoBreaker® device and 20.3±3.0 cm for the StoneBreaker™ device (with standard damper). The repulsion distance using the 2-mm probes was 14.8±1.3 cm with hard and 20.5±2.5 cm with soft damper for the LithoBreaker® and 17.1±2.0 cm for the StoneBreaker™. In pendulum tests, the 2-mm probes again resulted in smaller displacements than the 1-mm probes, except for the LithoBreaker® with soft damper. For a fixed probe diameter, no statistically significant influence of the damper was observed in case of the LithoBreaker® (p>0.05). However, the StoneBreaker™ with 2-mm probes produced smaller repulsions than the LithoBreaker® with 2-mm probes and hard damper. The pendulum tests with Ho:YAG laser pulses showed that the repulsion is proportional to the fiber core diameter and the laser pulse energy. The laser-induced momentum is about a factor 100 lower than the one generated with the mechanical lithotripsy devices. Conclusions: The electromechanical LithoBreaker® and the pneumatic StoneBreaker™ produced significantly different repulsions (p<0.05). The laser-induced momentum transfer is significantly lower compared to the mechanical lithotripsy devices (p<0.05). Zusammenfassung Hintergrund: Die Steintherapie hat sich zunehmend hin zu minimal-invasiven endourologischen Techniken entwickelt. Dies schließt endoskopische Steinzertrümmerungstechniken ein, die eingesetzt werden, um die Größe des Uroliths vor der Entfernung zu reduzieren. In diesem Zusammenhang stellt der Rückstoß des Steins einen unerwünschten Effekt des mechanischen Stoßes dar, der daher minimiert werden sollte, um den Eingriff zu erleichtern und seinen Erfolg zu verbessern. Zielsetzung: Vergleich der Rückstoß-Charakteristiken von zwei eigenständigen tragbaren Lithotripsie-Geräten (LithoBreaker® und StoneBreaker™) und einem Ho:YAG-Laser mit Hilfe spezieller Testmodelle. Materialien und Methoden: In Wasserumgebung wurden Tests mit mechanischen Einwegsonden (1 und 2 mm im Durchmesser) sowie mit über optische Fasern (Kerndurchmesser von 230, 365 und 600 μm) übertragenen Ho:YAG-Laserpulsen durchgeführt, indem die Auslenkungen von unzerbrechlichen kugelförmigen Testmassen gemessen wurden (Stahlkugeln mit Durchmessern von 3, 4 bzw. 5 mm und Massen von 0,12, 0,27 bzw. 0,52 g, sowie eine Bleikugel mit einem Durchmesser von 5 mm und einer Masse von 0,72 g). Die Tests wurden in einer Acrylglas-Röhre mit 12 mm Innendurchmesser, welche als Ureter-Modell diente, sowie an einem Pendelaufbau durchgeführt. Zusätzlich wurden im Falle des LithoBreaker® harte und weiche Dämpfer miteinander verglichen. Mittelwerte und Standardabweichungen der Auslenkungen und der damit zusammenhängenden Messgrößen wurden für alle Experimentserien berechnet, und statistische Analysen wurden durchgeführt (Student’s t-Test, Mann-Whitney U-Test, Shapiro-Wilk-Test), um den Einfluss des Gerätetyps und der Geräteparameter zu testen. Ergebnisse: Die Rückstoßdistanz im Ureter-Modell bei Verwendung der 1 mm-Sonden betrug 19,4±1,9 cm mit hartem und 21,5±2,7 cm mit weichem Dämpfer für das LithoBreaker®-Gerät und 20,3±3,0 cm für das StoneBreaker™-Gerät (mit Standard-Dämpfer). Die Rückstoßdistanz bei Verwendung der 2 mm-Sonden lag bei 14,8±1,3 cm mit hartem und 20,5±2,5 cm mit weichem Dämpfer für den LithoBreaker® und 17,1±2,0 cm für den StoneBreaker™. In den Pendelversuchen mit den 2 mm-Sonden wurden erneut kleinere Auslenkungen als mit den 1 mm-Sonden gemessen, mit Ausnahme des LithoBreaker® mit weichem Dämpfer. Bei festem Sondendurchmesser wurde im Falle des LithoBreaker® kein statistisch signifikanter Einfluss des Dämpfers beobachtet (p>0,05). Der StoneBreaker™ mit 2 mm-Sonden erzeugte jedoch kleinere Rückstöße als der LithoBreaker® mit 2 mm-Sonden und hartem Dämpfer. Die Pendeltests mit Ho:YAG-Laserpulsen zeigten, dass der Rückstoß sich proportional zum Faserkerndurchmesser und der Laserpulsenergie verhält. Der laser-induzierte Impuls ist etwa um den Faktor 100 kleiner als der, der mit den mechanischen Lithotripsie-Geräten erzeugt wurde. Schlussfolgerungen: Mit dem elektromechanischen LithoBreaker® und dem pneumatischen StoneBreaker™ werden signifikant unterschiedliche Rückstöße erzeugt (p<0,05). Der laser-induzierte Impulsübertrag ist signifikant kleiner als bei den mechanischen Lithotripsie-Geräten (p<0,05).
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
碎石过程中斥力效应的研究
背景:结石治疗已向微创泌尿系统技术发展。这包括内镜下碎石术,在手术前减少尿石的大小。在这种情况下,石头的排斥构成了对石头的机械冲击的不希望的影响,因此应尽量减少,以促进干预和改善其结果。目的:使用专用测试模型比较两种独立的手持式碎石设备(LithoBreaker®和StoneBreaker™)和一种Ho:YAG激光器的排斥力特性。材料和方法:采用一次性机械探头(直径为1和2 mm)和芯径为230、365和600 μm的光纤输送的Ho:YAG激光脉冲,在水下装置中通过测量不易破碎的球形试验质量(直径为3-5 mm,质量分别为0.12、0.27和0.52 g的钢球)的位移进行试验;和直径为5毫米,质量为0.72 g的铅球)。试验在内径为12毫米的丙烯酸玻璃管中进行,作为输尿管模型,并在摆摆装置上进行。此外,在碎石机®的情况下,比较了硬阻尼器和软阻尼器。对所有实验集的位移和相关量的均值和标准差进行评估,并进行统计分析(Student 's t检验、Mann-Whitney u检验、Shapiro-Wilk检验),以检验设备类型和设备参数的影响。结果:使用1 mm探头的输尿管模型中,硬探头的斥力距离为19.4±1.9 cm,软探头的斥力距离为21.5±2.7 cm,标准探头的斥力距离为20.3±3.0 cm。使用2毫米探头的斥力距离为14.8±1.3 cm(硬阻尼),20.5±2.5 cm(软阻尼),LithoBreaker®和StoneBreaker™的斥力距离为17.1±2.0 cm(软阻尼)。在摆锤测试中,除了带有软阻尼器的LithoBreaker®外,2mm探头的位移也比1mm探头小。对于固定探头直径,碎石机®没有观察到阻尼器的统计显著影响(p>0.05)。然而,带有2mm探头的StoneBreaker产生的排斥比带有2mm探头和硬阻尼器的LithoBreaker产生的排斥更小。用Ho:YAG激光脉冲进行的摆试验表明,斥力与光纤芯径和激光脉冲能量成正比。激光诱导的动量比机械碎石装置产生的动量低约100倍。结论:机电碎石机®和气动碎石机™产生了显著不同的排斥力(p0,05)。Der StoneBreaker™mit 2 mm-Sonden erzeugte jedoch kleinere Rückstöße Der StoneBreaker®mit 2 mm-Sonden und hartem Dämpfer。Die Pendeltests mit Ho:YAG-Laserpulsen zeigten, dasass der rckstoß sich proportional zum Faserkerndurchmesser and der laserpulsenerggie verhält。激光工业脉冲系统是由激光工业驱动的,激光工业驱动的,激光工业驱动的,激光工业驱动的,激光机械驱动的Lithotripsie-Geräten激光工业驱动的。Schlussfolgerungen: Mit dem elektromechanischen LithoBreaker®和dem pneumatischen StoneBreaker™是显著的unterschiedliche Rückstöße erzeugt (p< 0.05)。激光诱导脉冲效应显著(p< 0.05) Lithotripsie-Geräten。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
FLIM and PLIM in biomedical research – An innovative way to combine autofluorescence and oxygen measurements Use of a 1318 nm Nd:YAG laser for the resection of limited forms of pulmonary tuberculosis Can laser therapy be the answer for radiodermatitis in anal cancer patients? Two case reports Hydrogen peroxide detection in viable and apoptotic tumor cells under action of cisplatin and bleomycin LMTB winner of the Innovation Award Berlin Brandenburg 2015
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1