Use of FIA systems for on-line dilution in multielement determination by ICP–MS

José Anchieta Gomes Neto, José Bento Borba Silva, Ivan Gonçalves Souza, Adilson José Curtius
{"title":"Use of FIA systems for on-line dilution in multielement determination by ICP–MS","authors":"José Anchieta Gomes Neto,&nbsp;José Bento Borba Silva,&nbsp;Ivan Gonçalves Souza,&nbsp;Adilson José Curtius","doi":"10.1002/(SICI)1098-2728(1999)11:4<240::AID-LRA10>3.0.CO;2-6","DOIUrl":null,"url":null,"abstract":"<p>A flow-injection (FI) system to match concentrations was used as an auto-diluter in multielement determination by inductively coupled plasma–mass spectrometry (ICP–MS). The flow system comprised loop-based injection or a timed valve that introduced a variable sample volume into a spray chamber through a standard Meinhard nebulizer of an ICP–MS. Routinely analyzed samples such as water, plant, and steel were selected. The accuracy of multielement determination was checked against water standard reference material from the National Institute of Standards and Technology (1643d), plant standard reference material from the National Bureau of Standards (1572 citrus leaves), and steel standard reference material from the National Bureau of Standards (AISI 4340). The measuring system was calibrated with a multielement solution, yielding a linear plot with good precision [relative standard deviation (RSD) &lt; 3%, n = 12]. The results were in agreement at a 95% confidence level with the certified values for the reference materials and also with those obtained by continuous aspiration and by (FI) with a discrete volume. © 1999 John Wiley &amp; Sons, Inc. Lab Robotics and Automation 11: 240–247, 1999</p>","PeriodicalId":100863,"journal":{"name":"Laboratory Robotics and Automation","volume":"11 4","pages":"240-247"},"PeriodicalIF":0.0000,"publicationDate":"1999-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/(SICI)1098-2728(1999)11:4<240::AID-LRA10>3.0.CO;2-6","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory Robotics and Automation","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/%28SICI%291098-2728%281999%2911%3A4%3C240%3A%3AAID-LRA10%3E3.0.CO%3B2-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5

Abstract

A flow-injection (FI) system to match concentrations was used as an auto-diluter in multielement determination by inductively coupled plasma–mass spectrometry (ICP–MS). The flow system comprised loop-based injection or a timed valve that introduced a variable sample volume into a spray chamber through a standard Meinhard nebulizer of an ICP–MS. Routinely analyzed samples such as water, plant, and steel were selected. The accuracy of multielement determination was checked against water standard reference material from the National Institute of Standards and Technology (1643d), plant standard reference material from the National Bureau of Standards (1572 citrus leaves), and steel standard reference material from the National Bureau of Standards (AISI 4340). The measuring system was calibrated with a multielement solution, yielding a linear plot with good precision [relative standard deviation (RSD) < 3%, n = 12]. The results were in agreement at a 95% confidence level with the certified values for the reference materials and also with those obtained by continuous aspiration and by (FI) with a discrete volume. © 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 240–247, 1999

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
ICP-MS中在线稀释FIA系统的应用
在电感耦合等离子体质谱(ICP-MS)测定多元素时,采用流动注射(FI)系统作为自动稀释剂进行浓度匹配。流动系统包括基于回路的注射或定时阀,通过ICP-MS的标准Meinhard雾化器将可变样本量引入喷雾室。常规分析的样本,如水,植物和钢铁被选择。根据美国国家标准技术研究院的水标准标准物质(1643d)、美国国家标准局的植物标准标准物质(1572柑橘叶)和美国国家标准局的钢铁标准标准物质(AISI 4340)对多元素测定的准确性进行了检验。测量系统用多元素溶液进行校准,得到精度良好的线性图[相对标准偏差(RSD) <3%, n = 12]。结果在95%的置信水平上与标准物质的认证值一致,也与连续抽吸和离散体积(FI)获得的值一致。©1999 John Wiley &儿子,Inc。实验室机器人与自动化,1999
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Flow system for calibration and interference examination in two-component analysis Development of hardware and software for on-line sample preparation using sequential-injection analysis Automation of procedures to measure ash content for polyolefin and ryton poly(phenylene sulfide) resins Spectrophotometric determination of phosphite in fertilizers in a flow injection system with online sample preparation Determination of trace amounts of copper using a flow-through spectrophotometric sensor
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1