The SOCA type of eye lens dosimeter testing to measure Hp(3) dose

A. T. Desviana, F. Adhisti, B. Bunawas
{"title":"The SOCA type of eye lens dosimeter testing to measure Hp(3) dose","authors":"A. T. Desviana, F. Adhisti, B. Bunawas","doi":"10.1063/5.0061158","DOIUrl":null,"url":null,"abstract":"Interventional cardiology and radiology workers are exposed with occupational radiation. Radiation workers must use not only personal dosimeter; but also eye lens dosimeter in order to maintain the dose that does not exceed the dose limit value set by Nuclear Energy Regulatory Agency (Bapeten) which is 20 mSv a year with no single year exceeding 50 mSv. In order to provide a reliable with affordable price dosimeter, a new eye lens dosimeter was developed at NuklindoLab using TLD-900 (CaSO4:Dy) disc for measurement of dose in term of Hp(3). The type testing including homogeneity test, linearity of dose response, angular dependence test, and determining detection threshold were done after the irradiation carried out in Secondary Standard Dosimetry Laboratory (SSDL) Agensi Nuklear Malaysia. The method of the irradiation was based on ISO 4037-3 and the reference of the type testing was based on ISO 12794. The coefficient of variance (CV) value from homogeneity test results 4.8%, the linearity test result R2 values vary from 0.998 to 0.999, the angular dependence result ranges from 0% to 19% with average value 12.9%, and the detection threshold of SOCA dosimeter value was 0.35 mSv.","PeriodicalId":20561,"journal":{"name":"PROCEEDINGS OF THE 6TH INTERNATIONAL SYMPOSIUM ON CURRENT PROGRESS IN MATHEMATICS AND SCIENCES 2020 (ISCPMS 2020)","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-07-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"PROCEEDINGS OF THE 6TH INTERNATIONAL SYMPOSIUM ON CURRENT PROGRESS IN MATHEMATICS AND SCIENCES 2020 (ISCPMS 2020)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1063/5.0061158","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Interventional cardiology and radiology workers are exposed with occupational radiation. Radiation workers must use not only personal dosimeter; but also eye lens dosimeter in order to maintain the dose that does not exceed the dose limit value set by Nuclear Energy Regulatory Agency (Bapeten) which is 20 mSv a year with no single year exceeding 50 mSv. In order to provide a reliable with affordable price dosimeter, a new eye lens dosimeter was developed at NuklindoLab using TLD-900 (CaSO4:Dy) disc for measurement of dose in term of Hp(3). The type testing including homogeneity test, linearity of dose response, angular dependence test, and determining detection threshold were done after the irradiation carried out in Secondary Standard Dosimetry Laboratory (SSDL) Agensi Nuklear Malaysia. The method of the irradiation was based on ISO 4037-3 and the reference of the type testing was based on ISO 12794. The coefficient of variance (CV) value from homogeneity test results 4.8%, the linearity test result R2 values vary from 0.998 to 0.999, the angular dependence result ranges from 0% to 19% with average value 12.9%, and the detection threshold of SOCA dosimeter value was 0.35 mSv.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
采用SOCA型眼晶状体剂量计检测Hp(3)剂量
介入心脏病学和放射学工作者暴露在职业性辐射中。辐射工作人员不仅要使用个人剂量计;还有眼透镜剂量计,以保持剂量不超过核能监管机构(Bapeten)规定的剂量限值,即每年20毫西弗,没有一年超过50毫西弗。为了提供一种可靠且价格合理的剂量计,NuklindoLab开发了一种新的眼透镜剂量计,使用TLD-900 (CaSO4:Dy)光盘测量Hp(3)的剂量。辐照后在马来西亚原子能机构二级标准剂量测定实验室(SSDL)进行类型试验,包括均匀性试验、剂量响应线性、角依赖性试验和检测阈值的确定。辐照方法参照ISO 4037-3,型式试验参照ISO 12794。同质性检验结果的方差系数(CV)为4.8%,线性度检验结果R2为0.998 ~ 0.999,角依赖性结果为0% ~ 19%,平均值为12.9%,SOCA剂量计值的检测阈值为0.35 mSv。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Singular Integral Operators, Quantitative Flatness, and Boundary Problems Mirzakhani’s Curve Counting and Geodesic Currents The conception of national nuclear emergency preparedness system from the state defense perspective Computer security regulation for nuclear installation: A suggestion Optimal reinsurance combination of quota-share and stop-loss reinsurance based on conditional-tail-expectation (CTE) optimization
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1