{"title":"A NEW SPECTROSCOPY METHOD FOR THE QUANTITATIVE DETERMINATION OF IRON(III) BASED ON CURCUMIN REAGENT","authors":"Q. Tran, Dinh-Vu Le","doi":"10.47743/ACHI-2021-1-0004","DOIUrl":null,"url":null,"abstract":"The paper describes a simple, sensitive, and selective spectroscopy method for the determination of Fe by curcumin reagent and iron(III). In acid media, a 1:2 complex was formed; its properties were identified by UV-VIS, FTIR, Raman, and MS spectra. The maximum absorbance of UV-Vis spectra appeared at 518 nm. The parameters of the method for the determination of Fe was validated. Beer’s law was obeyed in the range 0.2×10 M to 2.0×10mol/L of Fe. The limited of detection (LOD) and limited of quantitation (LOQ) has been to be 0.5.0×10 6 and 1.5×10 mol/L. This method was applied to determine the concentration of ferrous total in various samples.","PeriodicalId":6958,"journal":{"name":"Acta Chemica Iasi","volume":"1 1","pages":"45-64"},"PeriodicalIF":0.4000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Chemica Iasi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.47743/ACHI-2021-1-0004","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 1
Abstract
The paper describes a simple, sensitive, and selective spectroscopy method for the determination of Fe by curcumin reagent and iron(III). In acid media, a 1:2 complex was formed; its properties were identified by UV-VIS, FTIR, Raman, and MS spectra. The maximum absorbance of UV-Vis spectra appeared at 518 nm. The parameters of the method for the determination of Fe was validated. Beer’s law was obeyed in the range 0.2×10 M to 2.0×10mol/L of Fe. The limited of detection (LOD) and limited of quantitation (LOQ) has been to be 0.5.0×10 6 and 1.5×10 mol/L. This method was applied to determine the concentration of ferrous total in various samples.