Jérémy Alexandre, Kevin Tan, Tiago P Almeida, Josep Sola, Bruce S Alpert, Jay Shah
{"title":"Validation of the Aktiia blood pressure cuff for clinical use according to the ANSI/AAMI/ISO 81060-2:2013 protocol.","authors":"Jérémy Alexandre, Kevin Tan, Tiago P Almeida, Josep Sola, Bruce S Alpert, Jay Shah","doi":"10.1097/MBP.0000000000000639","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Assess the accuracy and precision of the Aktiia initialization oscillometric upper-arm cuff device (Aktiia SA, Neuchâtel, Switzerland) for home blood pressure (BP) monitoring in the general population according to the American National Standards Institute / Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard.</p><p><strong>Methods: </strong>Three trained observers validated BP measurements performed using the Aktiia cuff versus BP measurements performed using a standard mercury sphygmomanometer. Two ISO 81060-2 criteria were used to validate the Aktiia cuff. Criterion 1 evaluated, for both SBP and DBP, whether the mean error between BP readings performed by the Aktiia cuff and auscultation was ≤±5 mmHg, and whether the SD of the error was ≤8 mmHg. Criterion 2 assessed whether, for the SBP and DBP of each individual subject, the SD of the averaged paired determinations per subject of the Aktiia cuff and of the auscultation met the criteria listed in the table of Averaged Subject Data Acceptance.</p><p><strong>Results: </strong>Mean differences between the Aktiia cuff and the standard mercury sphygmomanometer (criterion 1) were 1.3 ± 7.11 mmHg for SBP and -0.2 ± 5.46 mmHg for DBP. The SD of the averaged paired differences per subject (criterion 2) was 6.55 mmHg for SBP and 5.15 mmHg for DBP.</p><p><strong>Conclusion: </strong>Aktiia initialization cuff complies with the requirements of the ANSI/AAMI/ISO guidelines and can be safely recommended for BP measurements in the adult population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/5c/40/bpmj-28-109.PMC9981319.pdf","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000639","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/2/17 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 3
Abstract
Objective: Assess the accuracy and precision of the Aktiia initialization oscillometric upper-arm cuff device (Aktiia SA, Neuchâtel, Switzerland) for home blood pressure (BP) monitoring in the general population according to the American National Standards Institute / Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard.
Methods: Three trained observers validated BP measurements performed using the Aktiia cuff versus BP measurements performed using a standard mercury sphygmomanometer. Two ISO 81060-2 criteria were used to validate the Aktiia cuff. Criterion 1 evaluated, for both SBP and DBP, whether the mean error between BP readings performed by the Aktiia cuff and auscultation was ≤±5 mmHg, and whether the SD of the error was ≤8 mmHg. Criterion 2 assessed whether, for the SBP and DBP of each individual subject, the SD of the averaged paired determinations per subject of the Aktiia cuff and of the auscultation met the criteria listed in the table of Averaged Subject Data Acceptance.
Results: Mean differences between the Aktiia cuff and the standard mercury sphygmomanometer (criterion 1) were 1.3 ± 7.11 mmHg for SBP and -0.2 ± 5.46 mmHg for DBP. The SD of the averaged paired differences per subject (criterion 2) was 6.55 mmHg for SBP and 5.15 mmHg for DBP.
Conclusion: Aktiia initialization cuff complies with the requirements of the ANSI/AAMI/ISO guidelines and can be safely recommended for BP measurements in the adult population.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.