{"title":"Accuracy of X-Ray Exposure Parameters in some Radio-diagnostic Centres in Kaduna State, Nigeria","authors":"Abeabe, Christopher Ishiekwen, A. Aliyu","doi":"10.48153/jrrs.v32i1.223228","DOIUrl":null,"url":null,"abstract":"\n \n \n \nBackground: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. \nObjective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. \nMethodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. \nResults: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. \nConclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres. \n \n \n \n \n \n ","PeriodicalId":16919,"journal":{"name":"Journal of Radiography and Radiation Sciences","volume":"29 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-03-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiography and Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.48153/jrrs.v32i1.223228","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury.
Objective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection.
Methodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape.
Results: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance.
Conclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres.