{"title":"Feasibility of Tc-99 m sestamibi uptake quantification with few-projection emission tomography.","authors":"A M García-Esparza, H M Garnica-Garza","doi":"10.3233/XST-221303","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Molecular breast imaging uses Tc-99 m sestamibi to obtain functional images of the breast. Determining the Tc-99 m sestamibi uptake in volumes of interest in the breast may be useful in assessing the response to neoadjuvant chemotherapy or for the purposes of breast cancer risk assessment.</p><p><strong>Purpose: </strong>To determine, using Monte Carlo simulation, if emission tomography can be used to quantify the uptake of Tc-99 m sestamibi in molecular breast imaging and if so, to determine the accuracy as a function of the number of projections used in the reconstruction process.</p><p><strong>Methods: </strong>In this study, two voxelized breast models are implemented with different ratios of fibroglandular to fatty tissue and tumoral masses of varying dimensions. Monte Carlo simulation is used to calculate sets of projections, which assumes that each tumoral mass contains a given Tc-99 m activity. Projections are also calculated for a calibration phantom in order to correlate the known activity with the image pixel value. For each case, the total number of calculated projections is 36 and the reconstruction is carried out for 36, 18, 9, 7 and 5 projections, respectively, using an open source image reconstruction toolbox.</p><p><strong>Results: </strong>Study data show that determination of Tc-99 m sestamibi uptake with and average error of 7% can be carried out with as little as 7 projections.</p><p><strong>Conclusions: </strong>Molecular breast emission tomography enables to accurately determine the Tc-99 m sestamibi tumoral mass uptake with the number of projections very close to the number of images currently acquired in clinical practice.</p>","PeriodicalId":49948,"journal":{"name":"Journal of X-Ray Science and Technology","volume":"31 2","pages":"285-299"},"PeriodicalIF":1.7000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of X-Ray Science and Technology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3233/XST-221303","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Molecular breast imaging uses Tc-99 m sestamibi to obtain functional images of the breast. Determining the Tc-99 m sestamibi uptake in volumes of interest in the breast may be useful in assessing the response to neoadjuvant chemotherapy or for the purposes of breast cancer risk assessment.
Purpose: To determine, using Monte Carlo simulation, if emission tomography can be used to quantify the uptake of Tc-99 m sestamibi in molecular breast imaging and if so, to determine the accuracy as a function of the number of projections used in the reconstruction process.
Methods: In this study, two voxelized breast models are implemented with different ratios of fibroglandular to fatty tissue and tumoral masses of varying dimensions. Monte Carlo simulation is used to calculate sets of projections, which assumes that each tumoral mass contains a given Tc-99 m activity. Projections are also calculated for a calibration phantom in order to correlate the known activity with the image pixel value. For each case, the total number of calculated projections is 36 and the reconstruction is carried out for 36, 18, 9, 7 and 5 projections, respectively, using an open source image reconstruction toolbox.
Results: Study data show that determination of Tc-99 m sestamibi uptake with and average error of 7% can be carried out with as little as 7 projections.
Conclusions: Molecular breast emission tomography enables to accurately determine the Tc-99 m sestamibi tumoral mass uptake with the number of projections very close to the number of images currently acquired in clinical practice.
期刊介绍:
Research areas within the scope of the journal include:
Interaction of x-rays with matter: x-ray phenomena, biological effects of radiation, radiation safety and optical constants
X-ray sources: x-rays from synchrotrons, x-ray lasers, plasmas, and other sources, conventional or unconventional
Optical elements: grazing incidence optics, multilayer mirrors, zone plates, gratings, other diffraction optics
Optical instruments: interferometers, spectrometers, microscopes, telescopes, microprobes