Micro-Red Blood Cell, Fragmented Red Blood Cell, Platelet Distribution Width, Mean Platelet Volume, and Platelet-Large Cell Ratio on Sysmex XN Series Hematology Analyzers Can Be Used for the Reflex Test of Impedance Platelet Count in Clinical Practice.

IF 3.7 3区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY Archives of pathology & laboratory medicine Pub Date : 2024-01-01 DOI:10.5858/arpa.2022-0030-OA
Si Chen, Zhigang Mao, Shuang Wang, Jiamin Deng, Hongyan Liao, Qin Zheng
{"title":"Micro-Red Blood Cell, Fragmented Red Blood Cell, Platelet Distribution Width, Mean Platelet Volume, and Platelet-Large Cell Ratio on Sysmex XN Series Hematology Analyzers Can Be Used for the Reflex Test of Impedance Platelet Count in Clinical Practice.","authors":"Si Chen, Zhigang Mao, Shuang Wang, Jiamin Deng, Hongyan Liao, Qin Zheng","doi":"10.5858/arpa.2022-0030-OA","DOIUrl":null,"url":null,"abstract":"<p><strong>Context.—: </strong>Platelet (PLT) counting with impedance (PLT-I) is widely used but has low specificity. PLT counting with fluorescence (PLT-F), tested by the Sysmex XN series with high specificity, can be a complementary method to PLT-I.</p><p><strong>Objective.—: </strong>To identify red blood cell (RBC)- and PLT-related parameters as potential influencing factors for PLT-I and establish PLT reflex test rules with PLT-F.</p><p><strong>Design.—: </strong>We prospectively tested both PLT-I and PLT-F in all 3480 samples. In a development data set of 3000 samples, differences between the reflex and nonreflex groups were compared and influencing factors for PLT-I were identified by logistic regression. The area under the receiver operating characteristic (ROC) curve and cutoff values were obtained by ROC curve analysis. Validation was conducted in the remaining 480 samples (validation data set).</p><p><strong>Results.—: </strong>PLT-F showed comparable results with immunoplatelet counting. In logistic regression, increased micro-RBC absolute count (micro-RBC#), fragmented RBC absolute count (FRC#), PLT distribution width (PDW), mean PLT volume (MPV), PLT-large cell ratio (P-LCR), and immature PLT fraction absolute count (IPF#) were influencing factors for PLT-I. In ROC curve analysis, the cutoff values of micro-RBC#, FRC#, PDW, MPV, and P-LCR were 0.64 × 106/μL, 0.082 × 106/μL, 15.40 fL, 11.15 fL, and 33.95%, respectively. The areas under the ROC curve of micro-RBC# and FRC# were 0.77 and 0.79, respectively.</p><p><strong>Conclusions.—: </strong>Micro-RBC#, FRC#, PDW, MPV, P-LCR, and IPF# were factors affecting PLT-I. Among them, micro-RBC# and FRC# were the most impactful factors. From our study results, micro-RBC#, FRC#, MPV, PDW, and P-LCR can be used to establish reflex test rules for PLT counting in clinical work.</p>","PeriodicalId":8305,"journal":{"name":"Archives of pathology & laboratory medicine","volume":" ","pages":"61-67"},"PeriodicalIF":3.7000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archives of pathology & laboratory medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5858/arpa.2022-0030-OA","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Context.—: Platelet (PLT) counting with impedance (PLT-I) is widely used but has low specificity. PLT counting with fluorescence (PLT-F), tested by the Sysmex XN series with high specificity, can be a complementary method to PLT-I.

Objective.—: To identify red blood cell (RBC)- and PLT-related parameters as potential influencing factors for PLT-I and establish PLT reflex test rules with PLT-F.

Design.—: We prospectively tested both PLT-I and PLT-F in all 3480 samples. In a development data set of 3000 samples, differences between the reflex and nonreflex groups were compared and influencing factors for PLT-I were identified by logistic regression. The area under the receiver operating characteristic (ROC) curve and cutoff values were obtained by ROC curve analysis. Validation was conducted in the remaining 480 samples (validation data set).

Results.—: PLT-F showed comparable results with immunoplatelet counting. In logistic regression, increased micro-RBC absolute count (micro-RBC#), fragmented RBC absolute count (FRC#), PLT distribution width (PDW), mean PLT volume (MPV), PLT-large cell ratio (P-LCR), and immature PLT fraction absolute count (IPF#) were influencing factors for PLT-I. In ROC curve analysis, the cutoff values of micro-RBC#, FRC#, PDW, MPV, and P-LCR were 0.64 × 106/μL, 0.082 × 106/μL, 15.40 fL, 11.15 fL, and 33.95%, respectively. The areas under the ROC curve of micro-RBC# and FRC# were 0.77 and 0.79, respectively.

Conclusions.—: Micro-RBC#, FRC#, PDW, MPV, P-LCR, and IPF# were factors affecting PLT-I. Among them, micro-RBC# and FRC# were the most impactful factors. From our study results, micro-RBC#, FRC#, MPV, PDW, and P-LCR can be used to establish reflex test rules for PLT counting in clinical work.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Sysmex XN 系列血液分析仪上的微红细胞、碎红细胞、血小板分布宽度、平均血小板体积和血小板-大细胞比可用于临床实践中的阻抗血小板计数反射测试。
背景:血小板(PLT)阻抗计数法(PLT-I)被广泛使用,但特异性较低。Sysmex XN 系列测试的荧光血小板计数法(PLT-F)具有很高的特异性,可以作为 PLT-I 的补充方法:确定作为 PLT-I 潜在影响因素的红细胞(RBC)和 PLT 相关参数,并建立 PLT-F 的 PLT 反射检验规则:我们对所有 3480 份样本中的 PLT-I 和 PLT-F 进行了前瞻性测试。在 3000 份样本的开发数据集中,比较了反射组和非反射组之间的差异,并通过逻辑回归确定了 PLT-I 的影响因素。通过 ROC 曲线分析得出接收者操作特征曲线下面积和临界值。对剩余的 480 份样本(验证数据集)进行了验证:PLT-F与免疫血小板计数结果相当。在逻辑回归中,微小红细胞绝对计数(micro-RBC#)、碎小红细胞绝对计数(FRC#)、PLT 分布宽度(PDW)、PLT 平均体积(MPV)、PLT-大细胞比值(P-LCR)和未成熟 PLT 部分绝对计数(IPF#)的增加是 PLT-I 的影响因素。在 ROC 曲线分析中,micro-RBC#、FRC#、PDW、MPV 和 P-LCR 的临界值分别为 0.64 × 106/μL、0.082 × 106/μL、15.40 fL、11.15 fL 和 33.95%。微量 RBC# 和 FRC# 的 ROC 曲线下面积分别为 0.77 和 0.79:结论:微RBC#、FRC#、PDW、MPV、P-LCR和IPF#是影响PLT-I的因素。其中,微RBC#和FRC#是影响最大的因素。从我们的研究结果来看,在临床工作中,微RBC#、FRC#、MPV、PDW和P-LCR可用于建立PLT计数的反射测试规则。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
9.20
自引率
2.20%
发文量
369
审稿时长
3-8 weeks
期刊介绍: Welcome to the website of the Archives of Pathology & Laboratory Medicine (APLM). This monthly, peer-reviewed journal of the College of American Pathologists offers global reach and highest measured readership among pathology journals. Published since 1926, ARCHIVES was voted in 2009 the only pathology journal among the top 100 most influential journals of the past 100 years by the BioMedical and Life Sciences Division of the Special Libraries Association. Online access to the full-text and PDF files of APLM articles is free.
期刊最新文献
New Entities and Concepts in Salivary Gland Tumor Pathology: The Role of Molecular Alterations. Update on Sinonasal Tract Malignancies: Advances in Diagnostic Modalities. Update on Salivary Gland Fine-Needle Aspiration and the Milan System for Reporting Salivary Gland Cytopathology. BRAF Exon 15 Mutations in the Evaluation of Well-Differentiated Epithelial Nephroblastic Neoplasms in Children: A Report From the Children's Oncology Group Study AREN03B2. Neoplastic Progression in Intraductal Papillary Neoplasm of the Bile Duct.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1