Comparability of childhood blood pressure measurements with two different devices

IF 1.3 4区 医学 Q4 PHYSIOLOGY Clinical Physiology and Functional Imaging Pub Date : 2022-11-28 DOI:10.1111/cpf.12802
Lukas Streese, Christoph Hauser, Henner Hanssen
{"title":"Comparability of childhood blood pressure measurements with two different devices","authors":"Lukas Streese,&nbsp;Christoph Hauser,&nbsp;Henner Hanssen","doi":"10.1111/cpf.12802","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Higher blood pressure levels are associated with vascular dysfunction as early as childhood. Here, we aim to compare two blood pressure devices for use in childhood populations within a school setting to screen for elevated blood pressure in children.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Systolic and diastolic blood pressure, mean arteriolar pressure (MAP) as well as heart rate (HR) were measured with the validated Oscillomate 9002 and Mindray VS-900 in 82 randomly selected children between 10 and 12 years taking part in the EXAMIN YOUTH study in 2020. Measurements were performed after 5 min at rest in a sitting position in the school environment.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Both devices showed strong correlations for systolic (<i>r</i> = 0.62, <i>t</i>[80] = 7, <i>p</i> &lt; 0.001) and diastolic blood pressure (<i>r</i> = 0.73, <i>t</i>[80] = 9.52, <i>p</i> &lt; 0.001), MAP (<i>r</i> = 0.75, <i>t</i>[80] = 10.22, <i>p</i> &lt; 0.001) and HR (<i>r</i> = 0.89, <i>t</i>[79] = 17.77, <i>p</i> &lt; 0.001). There was no evidence for a statistically significant difference of both devices for systolic (110 ± 9 mmHg vs. 111 ± 9 mmHg, <i>p</i> = 0.574), diastolic blood pressure (67 ± 9 mmHg vs. 66 ± 9 mmHg, <i>p</i> = 0.301) as well as MAP (81 ± 8 mmHg vs. 80 ± 8 mmHg, <i>p</i> = 0.400) and HR (83 ± 12 BPM vs. 83 ± 11 BPM, <i>p</i> = 0.994).</p>\n </section>\n \n <section>\n \n <h3> Discussion</h3>\n \n <p>Both devices, the older Oscillomate 9002 and the current Mindray VS-900 showed good agreement for the measurement of blood pressure in school children. It therefore appears to be feasible to measure childhood blood pressure with either device or replace one device with another in prospective long-term studies or screening programmes as long as both are validated for use in children.</p>\n </section>\n </div>","PeriodicalId":10504,"journal":{"name":"Clinical Physiology and Functional Imaging","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2022-11-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Physiology and Functional Imaging","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/cpf.12802","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHYSIOLOGY","Score":null,"Total":0}
引用次数: 2

Abstract

Introduction

Higher blood pressure levels are associated with vascular dysfunction as early as childhood. Here, we aim to compare two blood pressure devices for use in childhood populations within a school setting to screen for elevated blood pressure in children.

Methods

Systolic and diastolic blood pressure, mean arteriolar pressure (MAP) as well as heart rate (HR) were measured with the validated Oscillomate 9002 and Mindray VS-900 in 82 randomly selected children between 10 and 12 years taking part in the EXAMIN YOUTH study in 2020. Measurements were performed after 5 min at rest in a sitting position in the school environment.

Results

Both devices showed strong correlations for systolic (r = 0.62, t[80] = 7, p < 0.001) and diastolic blood pressure (r = 0.73, t[80] = 9.52, p < 0.001), MAP (r = 0.75, t[80] = 10.22, p < 0.001) and HR (r = 0.89, t[79] = 17.77, p < 0.001). There was no evidence for a statistically significant difference of both devices for systolic (110 ± 9 mmHg vs. 111 ± 9 mmHg, p = 0.574), diastolic blood pressure (67 ± 9 mmHg vs. 66 ± 9 mmHg, p = 0.301) as well as MAP (81 ± 8 mmHg vs. 80 ± 8 mmHg, p = 0.400) and HR (83 ± 12 BPM vs. 83 ± 11 BPM, p = 0.994).

Discussion

Both devices, the older Oscillomate 9002 and the current Mindray VS-900 showed good agreement for the measurement of blood pressure in school children. It therefore appears to be feasible to measure childhood blood pressure with either device or replace one device with another in prospective long-term studies or screening programmes as long as both are validated for use in children.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
两种不同装置测量儿童血压的可比性
高血压早在儿童时期就与血管功能障碍有关。在这里,我们的目的是比较两种血压设备用于儿童人群在学校设置筛选儿童血压升高。方法:在2020年的EXAMIN YOUTH研究中,随机选择82名10 - 12岁的儿童,使用经过验证的Oscillomate 9002和Mindray VS-900测量收缩压、舒张压、平均小动脉压(MAP)和心率(HR)。在学校环境中静坐休息5分钟后进行测量。结果两种装置在收缩压(r = 0.62, t[80] = 7, p < 0.001)、舒张压(r = 0.73, t[80] = 9.52, p < 0.001)、MAP (r = 0.75, t[80] = 10.22, p < 0.001)和HR (r = 0.89, t[79] = 17.77, p < 0.001)方面均表现出较强的相关性。两种装置在收缩压(110±9 mmHg vs 111±9 mmHg, p = 0.574)、舒张压(67±9 mmHg vs 66±9 mmHg, p = 0.301)、MAP(81±8 mmHg vs 80±8 mmHg, p = 0.400)和心率(83±12 BPM vs 83±11 BPM, p = 0.994)方面无统计学差异。这两种设备,旧的振荡9002和当前的迈瑞VS-900在测量学龄儿童血压方面表现出良好的一致性。因此,在前瞻性长期研究或筛查计划中,只要两种设备都被证实适用于儿童,用其中一种设备测量儿童血压或用另一种设备替换一种设备似乎是可行的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.40
自引率
5.60%
发文量
62
审稿时长
6-12 weeks
期刊介绍: Clinical Physiology and Functional Imaging publishes reports on clinical and experimental research pertinent to human physiology in health and disease. The scope of the Journal is very broad, covering all aspects of the regulatory system in the cardiovascular, renal and pulmonary systems with special emphasis on methodological aspects. The focus for the journal is, however, work that has potential clinical relevance. The Journal also features review articles on recent front-line research within these fields of interest. Covered by the major abstracting services including Current Contents and Science Citation Index, Clinical Physiology and Functional Imaging plays an important role in providing effective and productive communication among clinical physiologists world-wide.
期刊最新文献
Microvascular effects of a mixed meal tolerance test: a model validation study. PET/MRI in paediatric inflammatory bowel disease - a prospective accuracy study. ‘Effects of dehydration on central blood pressure in young healthy adults’ The concurrent validity of a portable ultrasound probe for muscle thickness measurements. Tortuosity of the left anterior descending artery is associated with hypertension and is not independently related to physical performance: A cardiac computed tomography study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1