{"title":"[Transcutaneous bone conduction implant with self-drilling screws : A new method for fixation of an active transcutaneous bone conduction implant. German version].","authors":"Assen Koitschev, Marcus Neudert, Thomas Lenarz","doi":"10.1007/s00106-023-01294-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The active transcutaneous bone conduction implant (tBCI; BONEBRIDGE™ BCI 601; MED-EL, Innsbruck, Austria) is fixed to the skull with two self-tapping screws in predrilled screw channels. The aim of this prospective study was to evaluate the safety and effectiveness of fixation with self-drilling screws instead of the self-tapping screws, in order to simplify the surgical procedure.</p><p><strong>Materials and methods: </strong>Nine patients (mean age 37 ± 16 years, range 14-57 years) were examined pre- and 12 months postoperatively for word recognition scores (WRS) at 65 dB SPL, sound-field (SF) thresholds, bone conduction thresholds (BC), health-related quality of life (Assessment of Quality of Life, AQOL-8D questionnaire), and adverse events (AE).</p><p><strong>Results: </strong>Due to avoidance of one surgical step, the surgical technique was simplified. Mean WRS in SF was 11.1 ± 22.2% (range 0-55%) pre- and 77.2 ± 19.9% (range 30-95%) postoperatively; mean SF threshold (pure tone audiometry, PTA<sub>4</sub>) improved from 61.2 ± 14.3 dB HL (range 37.0-75.3 dB HL) to 31.9 ± 7.2 dB HL (range 22.8-45.0 dB HL); mean BC thresholds were constant at 16.7 ± 6.8 dB HL (range 6.3-27.5 dB HL) pre- and 14.2 ± 6.2 dB HL (range 5.8-23.8 dB HL) postoperatively. AQOL-8D mean utility score increased from 0.65 ± 0.18 preoperatively to 0.82 ± 0.17 postoperatively. No device-related adverse events occurred.</p><p><strong>Conclusion: </strong>Implant fixation by means of self-drilling screws was safe and effective in all nine patients. There was significant audiological benefit 12 months after implantation.</p>","PeriodicalId":55052,"journal":{"name":"Hno","volume":null,"pages":null},"PeriodicalIF":0.8000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hno","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00106-023-01294-x","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/4/27 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"OTORHINOLARYNGOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The active transcutaneous bone conduction implant (tBCI; BONEBRIDGE™ BCI 601; MED-EL, Innsbruck, Austria) is fixed to the skull with two self-tapping screws in predrilled screw channels. The aim of this prospective study was to evaluate the safety and effectiveness of fixation with self-drilling screws instead of the self-tapping screws, in order to simplify the surgical procedure.
Materials and methods: Nine patients (mean age 37 ± 16 years, range 14-57 years) were examined pre- and 12 months postoperatively for word recognition scores (WRS) at 65 dB SPL, sound-field (SF) thresholds, bone conduction thresholds (BC), health-related quality of life (Assessment of Quality of Life, AQOL-8D questionnaire), and adverse events (AE).
Results: Due to avoidance of one surgical step, the surgical technique was simplified. Mean WRS in SF was 11.1 ± 22.2% (range 0-55%) pre- and 77.2 ± 19.9% (range 30-95%) postoperatively; mean SF threshold (pure tone audiometry, PTA4) improved from 61.2 ± 14.3 dB HL (range 37.0-75.3 dB HL) to 31.9 ± 7.2 dB HL (range 22.8-45.0 dB HL); mean BC thresholds were constant at 16.7 ± 6.8 dB HL (range 6.3-27.5 dB HL) pre- and 14.2 ± 6.2 dB HL (range 5.8-23.8 dB HL) postoperatively. AQOL-8D mean utility score increased from 0.65 ± 0.18 preoperatively to 0.82 ± 0.17 postoperatively. No device-related adverse events occurred.
Conclusion: Implant fixation by means of self-drilling screws was safe and effective in all nine patients. There was significant audiological benefit 12 months after implantation.
期刊介绍:
HNO is an internationally recognized journal and addresses all ENT specialists in practices and clinics dealing with all aspects of ENT medicine, e.g. prevention, diagnostic methods, complication management, modern therapy strategies and surgical procedures.
Review articles provide an overview on selected topics and offer the reader a summary of current findings from all fields of ENT medicine.
Freely submitted original papers allow the presentation of important clinical studies and serve the scientific exchange.
Case reports feature interesting cases and aim at optimizing diagnostic and therapeutic strategies.
Review articles under the rubric ''Continuing Medical Education'' present verified results of scientific research and their integration into daily practice.