Kyungmo Sung, Seonghoon Jo, Jaewook Lee, Jeong Hoan Park, Young Hoon Park, Jeongjoo Moon, Sung June Kim, Joonsoo Jeong, Jonghwan Lee, Kyungsik Eom
{"title":"Computational analysis of multichannel magnetothermal neural stimulation using magnetic resonator array.","authors":"Kyungmo Sung, Seonghoon Jo, Jaewook Lee, Jeong Hoan Park, Young Hoon Park, Jeongjoo Moon, Sung June Kim, Joonsoo Jeong, Jonghwan Lee, Kyungsik Eom","doi":"10.1007/s13534-023-00267-x","DOIUrl":null,"url":null,"abstract":"<p><p>Heating nanoparticles with a magnetic field could facilitate selective remote control of neural activity in deep tissue. However, current magnetothermal stimulation approaches are limited to single-channel stimulation. Here, we investigated various designs for multichannel magnetothermal stimulation based on an array of resonant coils that are driven by a single loop coil. Using a tuning capacitor that allows resonant coils to resonate at the operating frequency, each coil's ON and OFF resonance can be controlled, enabling us to select stimulation channels. We found that smaller inner diameters of resonant coils produce more localized magnetic fields while larger coils produce magnetic fields over a longer distance. The constructed multichannel resonant coil arrays can provide a high enough magnetic field intensity to raise the temperature of nanoparticles by 8 °C when we apply 35.2 W into the loop coil that is spaced 1 mm from the target neurons. This multichannel stimulation using a simple resonant circuit approach would be useful for clinical applications of magnetothermal neural stimulation.</p>","PeriodicalId":46898,"journal":{"name":"Biomedical Engineering Letters","volume":"13 2","pages":"209-219"},"PeriodicalIF":3.2000,"publicationDate":"2023-02-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10130299/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Engineering Letters","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s13534-023-00267-x","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/5/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Heating nanoparticles with a magnetic field could facilitate selective remote control of neural activity in deep tissue. However, current magnetothermal stimulation approaches are limited to single-channel stimulation. Here, we investigated various designs for multichannel magnetothermal stimulation based on an array of resonant coils that are driven by a single loop coil. Using a tuning capacitor that allows resonant coils to resonate at the operating frequency, each coil's ON and OFF resonance can be controlled, enabling us to select stimulation channels. We found that smaller inner diameters of resonant coils produce more localized magnetic fields while larger coils produce magnetic fields over a longer distance. The constructed multichannel resonant coil arrays can provide a high enough magnetic field intensity to raise the temperature of nanoparticles by 8 °C when we apply 35.2 W into the loop coil that is spaced 1 mm from the target neurons. This multichannel stimulation using a simple resonant circuit approach would be useful for clinical applications of magnetothermal neural stimulation.
期刊介绍:
Biomedical Engineering Letters (BMEL) aims to present the innovative experimental science and technological development in the biomedical field as well as clinical application of new development. The article must contain original biomedical engineering content, defined as development, theoretical analysis, and evaluation/validation of a new technique. BMEL publishes the following types of papers: original articles, review articles, editorials, and letters to the editor. All the papers are reviewed in single-blind fashion.