Anne-Sofie Madsen Staples, Julie Schwartz, Kezia Ann Friis Præstmark, Marie Sand Traberg
{"title":"Novel Robust Needle Tip Design Enables Needle Reuse and Reduced Skin Trauma in Combination With Autoinjector Needle Shields.","authors":"Anne-Sofie Madsen Staples, Julie Schwartz, Kezia Ann Friis Præstmark, Marie Sand Traberg","doi":"10.1177/19322968231190408","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Pen needles and autoinjectors are necessary for millions of patients needing injectable drug treatment but pose economic and environmental burdens. A durable device with a multiuse needle could reduce cost and improve user experience. This study explores a novel robust needle tip (EXP) designed for multiple uses and durability against hooking.</p><p><strong>Method: </strong>Needle robustness was investigated through a structural analysis. Furthermore, EXP and control needles (NF30, NF28) were evaluated in an in-vivo porcine model as pen needles or embedded in autoinjectors to study the resulting increase in skin blood perfusion (SBP). The SBP was assessed by laser speckle contrast analysis (LASCA) of 192 randomized and blinded needle insertions.</p><p><strong>Results: </strong>Forming a 33 µm hook against a hard surface requires 0.92 N for the NF30 control needle and 5.38 N for EXP. The EXP did not induce more tissue trauma than the NF30. There was a positive relation between needle diameter and SBP (<i>P</i> < .05). Furthermore, the presence of an autoinjector shield and applied force of 10 N was found to significantly reduce SBP for worn EXP needles (<i>P</i> < .05) compared to insertions without autoinjector shield.</p><p><strong>Conclusions: </strong>The investigated robust needle EXP is on par with the single-use needle NF30 in terms of tissue trauma, which is further reduced by combining the needle with a needle shield. These results should encourage the innovation and development of durable, reusable injection systems with pharmacoeconomic and environmental value and a simplified and enhanced user experience for patients.</p>","PeriodicalId":15475,"journal":{"name":"Journal of Diabetes Science and Technology","volume":" ","pages":"352-360"},"PeriodicalIF":4.1000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11874290/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Diabetes Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/19322968231190408","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/8/9 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Pen needles and autoinjectors are necessary for millions of patients needing injectable drug treatment but pose economic and environmental burdens. A durable device with a multiuse needle could reduce cost and improve user experience. This study explores a novel robust needle tip (EXP) designed for multiple uses and durability against hooking.
Method: Needle robustness was investigated through a structural analysis. Furthermore, EXP and control needles (NF30, NF28) were evaluated in an in-vivo porcine model as pen needles or embedded in autoinjectors to study the resulting increase in skin blood perfusion (SBP). The SBP was assessed by laser speckle contrast analysis (LASCA) of 192 randomized and blinded needle insertions.
Results: Forming a 33 µm hook against a hard surface requires 0.92 N for the NF30 control needle and 5.38 N for EXP. The EXP did not induce more tissue trauma than the NF30. There was a positive relation between needle diameter and SBP (P < .05). Furthermore, the presence of an autoinjector shield and applied force of 10 N was found to significantly reduce SBP for worn EXP needles (P < .05) compared to insertions without autoinjector shield.
Conclusions: The investigated robust needle EXP is on par with the single-use needle NF30 in terms of tissue trauma, which is further reduced by combining the needle with a needle shield. These results should encourage the innovation and development of durable, reusable injection systems with pharmacoeconomic and environmental value and a simplified and enhanced user experience for patients.
期刊介绍:
The Journal of Diabetes Science and Technology (JDST) is a bi-monthly, peer-reviewed scientific journal published by the Diabetes Technology Society. JDST covers scientific and clinical aspects of diabetes technology including glucose monitoring, insulin and metabolic peptide delivery, the artificial pancreas, digital health, precision medicine, social media, cybersecurity, software for modeling, physiologic monitoring, technology for managing obesity, and diagnostic tests of glycation. The journal also covers the development and use of mobile applications and wireless communication, as well as bioengineered tools such as MEMS, new biomaterials, and nanotechnology to develop new sensors. Articles in JDST cover both basic research and clinical applications of technologies being developed to help people with diabetes.