Engineering Analysis of Non-Braided Polycaprolactone Bioresorbable Flow Diverters for Aneurysms.

IF 1.7 4区 医学 Q4 BIOPHYSICS Journal of Biomechanical Engineering-Transactions of the Asme Pub Date : 2023-11-01 DOI:10.1115/1.4063001
Mohammad Robiul Hossan, Vishal Barot, Seth Harriet, Lauren Peters, Alex Christopher Matsayko, Andrew Bauer, Khalid Hossain
{"title":"Engineering Analysis of Non-Braided Polycaprolactone Bioresorbable Flow Diverters for Aneurysms.","authors":"Mohammad Robiul Hossan, Vishal Barot, Seth Harriet, Lauren Peters, Alex Christopher Matsayko, Andrew Bauer, Khalid Hossain","doi":"10.1115/1.4063001","DOIUrl":null,"url":null,"abstract":"<p><p>This paper reports a nonbraided, bioresorbable polycaprolactone (PCL) flow diverter (FD) for the endovascular treatment of aneurysms. Bioresorbable FDs can reduce the risk associated with the permanent metallic FDs as they are resorbed by the body after curing of aneurysms. PCL FDs were designed and fabricated using an in-house hybrid electromelt spinning-fused deposition fabrication unit. Flow diverter's properties, surface qualities, and mechanical characteristics of PCL FDs of 50%, 60%, and 70% porosities were studied using scanning electron microscope (SEM), atomic force microscopy (AFM), and high precision universal testing machine (UTM). The deployability through a clinically relevant catheter was demonstrated in a PDMS aneurysm model. The angiographic visibility of the developed PCL FDs was evaluated using BaSO4 and Bi2O3 coatings of various concentration. The average strut thicknesses were 74.12 ± 6.63 μm, 63.07 ± 1.26 μm, and 56.82 ± 2.09 μm for PCL FDs with 50%, 60%, and 70% porosities, respectively. They average pore areas for the 50%, 60% and 70% porosities FDs were 0.055 ± 0.0056 mm2, 0. 0605 ± 0.0065 mm2, and 0.0712 ± 0.012 mm2, respectively. The surface quality was great with an RMS roughness value of 14.45 nm. The tensile, radial strength, and flexibility were found to be satisfactory and comparable to the nonbraided coronary stents. The developed PCL FDs were highly flexible and demonstrated to be deployable through conventional delivery system as low as 4 Fr catheters in a PDMS aneurysm model. The visibility under X-ray increases with the increasing concentration of coating materials BaSO4 and Bi2O3. The visibility intensity was slightly higher with Bi2O3 coating of PCL FDs. The overall results of the engineering analysis of the developed nonbraided PCL FDs are promising.</p>","PeriodicalId":54871,"journal":{"name":"Journal of Biomechanical Engineering-Transactions of the Asme","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10578075/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomechanical Engineering-Transactions of the Asme","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1115/1.4063001","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOPHYSICS","Score":null,"Total":0}
引用次数: 0

Abstract

This paper reports a nonbraided, bioresorbable polycaprolactone (PCL) flow diverter (FD) for the endovascular treatment of aneurysms. Bioresorbable FDs can reduce the risk associated with the permanent metallic FDs as they are resorbed by the body after curing of aneurysms. PCL FDs were designed and fabricated using an in-house hybrid electromelt spinning-fused deposition fabrication unit. Flow diverter's properties, surface qualities, and mechanical characteristics of PCL FDs of 50%, 60%, and 70% porosities were studied using scanning electron microscope (SEM), atomic force microscopy (AFM), and high precision universal testing machine (UTM). The deployability through a clinically relevant catheter was demonstrated in a PDMS aneurysm model. The angiographic visibility of the developed PCL FDs was evaluated using BaSO4 and Bi2O3 coatings of various concentration. The average strut thicknesses were 74.12 ± 6.63 μm, 63.07 ± 1.26 μm, and 56.82 ± 2.09 μm for PCL FDs with 50%, 60%, and 70% porosities, respectively. They average pore areas for the 50%, 60% and 70% porosities FDs were 0.055 ± 0.0056 mm2, 0. 0605 ± 0.0065 mm2, and 0.0712 ± 0.012 mm2, respectively. The surface quality was great with an RMS roughness value of 14.45 nm. The tensile, radial strength, and flexibility were found to be satisfactory and comparable to the nonbraided coronary stents. The developed PCL FDs were highly flexible and demonstrated to be deployable through conventional delivery system as low as 4 Fr catheters in a PDMS aneurysm model. The visibility under X-ray increases with the increasing concentration of coating materials BaSO4 and Bi2O3. The visibility intensity was slightly higher with Bi2O3 coating of PCL FDs. The overall results of the engineering analysis of the developed nonbraided PCL FDs are promising.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
动脉瘤用非编织聚己内酯生物可吸收分流器的工程分析。
本文报道了一种用于动脉瘤血管内治疗的非突袭、生物可吸收聚己内酯(PCL)分流器(FD)。生物可吸收的FDs可以降低与永久性金属FDs相关的风险,因为它们在动脉瘤治愈后被身体吸收。PCL-FD是使用内部混合静电纺丝熔融沉积制造单元设计和制造的。使用扫描电子显微镜(SEM)、原子力显微镜(AFM)和高精度通用试验机(UTM)研究了50%、60%和70%孔隙率的PCL-FD的分流器性能、表面质量和机械特性。在PDMS动脉瘤模型中证明了通过临床相关导管的可部署性。使用不同浓度的BaSO4和Bi2O3涂层评估所开发的PCL-FD的血管造影可见性。支柱的平均厚度为74.12 ± 6.63 μm,63.07 ± 1.26 μm和56.82 ± 2.09 μm的PCL-FD,孔隙率分别为50%、60%和70%。50%、60%和70%孔隙率FD的平均孔隙面积为0.055 ± 0.0056 mm2,0。0605 ± 0.0065 mm2和0.0712 ± 0.012 mm2。表面质量很好,RMS粗糙度值为14.45 nm。拉伸、径向强度和柔韧性均令人满意,可与非阻塞性冠状动脉支架相媲美。所开发的PCL-FD具有高度灵活性,并证明可通过低至4 PDMS动脉瘤模型中的Fr导管。X射线下的可见度随着涂层材料BaSO4和Bi2O3浓度的增加而增加。PCL-FD的Bi2O3涂层的可见度强度略高。对所开发的非突袭PCL-FD进行工程分析的总体结果是有希望的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.40
自引率
5.90%
发文量
169
审稿时长
4-8 weeks
期刊介绍: Artificial Organs and Prostheses; Bioinstrumentation and Measurements; Bioheat Transfer; Biomaterials; Biomechanics; Bioprocess Engineering; Cellular Mechanics; Design and Control of Biological Systems; Physiological Systems.
期刊最新文献
Image-Based Estimation of Left Ventricular Myocardial Stiffness. Phenomenological Muscle Constitutive Model With Actin-Titin Binding for Simulating Active Stretching. Regulatory Role of Collagen XI in the Establishment of Mechanical Properties of Tendons and Ligaments in Mice Is Tissue Dependent. Study of the Mechanism of Perceived Rotational Acceleration of a Bionic Semicircular Canal on the Basis of the "Circular Geometry Hypothesis". Walking Slope and Heavy Backpacks Affect Peak and Impulsive Lumbar Joint Contact Forces.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1