Irma Cerić Andelius, David Minarik, Eva Persson, Henrik Mosén, Kristian Valind, Elin Trägårdh, Jenny Oddstig
{"title":"环形镉锌碲化照相机的首次临床应用:与传统伽马照相机系统的比较研究。","authors":"Irma Cerić Andelius, David Minarik, Eva Persson, Henrik Mosén, Kristian Valind, Elin Trägårdh, Jenny Oddstig","doi":"10.1111/cpf.12853","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>A novel semiconductor cadmium zinc telluride (CZT) gamma camera system using a block sequential regularized expectation maximization (BSREM) reconstruction algorithm is now clinically available. Here we investigate how a multi-purpose ring-configurated CZT system can be safely applied in clinics and describe the initial optimization process.</p>\n </section>\n \n <section>\n \n <h3> Method</h3>\n \n <p>Seventy-six patients (bone-, cardiac- and lung scan) were scanned on a conventional gamma camera (planar and/or single-photon emission computed tomography [SPECT]/SPECT-CT) used in clinical routine and on the ring-configurated CZT camera Starguide (GE Healthcare). These data were used to validate and optimize the Starguide system for routine clinical use.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Comparable image quality for the Starguide system, to that of the conventional gamma camera, was achieved for bone scan (4 min/bed position [BP] using a relative difference prior [RDP] with gamma 2 and beta 0.4, along with 10 iterations and 10 subsets), cardiac scan (8 min [stress] and 3 min 20 s [rest] using median root prior [MRP] with beta 0.07 non attenuation corrected and 0.008 attenuation corrected and 50 interations and 10 subsets for both stress and rest) and lung scan (10 min [vent] and 5 min [perf] using RDP with gamma 0.5 and beta 0.03 [vent] and 0.02 [perf] and 20 interations and 10 subsets for both vent and perf).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>It was possible to transition from a conventional gamma camera to the Starguide system as part of the clinical routine, with acceptable image quality. Images from the Starguide system were deemed to be at least as good as those from a conventional gamma camera.</p>\n </section>\n </div>","PeriodicalId":10504,"journal":{"name":"Clinical Physiology and Functional Imaging","volume":"44 1","pages":"79-88"},"PeriodicalIF":1.3000,"publicationDate":"2023-08-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/cpf.12853","citationCount":"0","resultStr":"{\"title\":\"First clinical experience of a ring-configured cadmium zinc telluride camera: A comparative study versus conventional gamma camera systems\",\"authors\":\"Irma Cerić Andelius, David Minarik, Eva Persson, Henrik Mosén, Kristian Valind, Elin Trägårdh, Jenny Oddstig\",\"doi\":\"10.1111/cpf.12853\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>A novel semiconductor cadmium zinc telluride (CZT) gamma camera system using a block sequential regularized expectation maximization (BSREM) reconstruction algorithm is now clinically available. Here we investigate how a multi-purpose ring-configurated CZT system can be safely applied in clinics and describe the initial optimization process.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Method</h3>\\n \\n <p>Seventy-six patients (bone-, cardiac- and lung scan) were scanned on a conventional gamma camera (planar and/or single-photon emission computed tomography [SPECT]/SPECT-CT) used in clinical routine and on the ring-configurated CZT camera Starguide (GE Healthcare). These data were used to validate and optimize the Starguide system for routine clinical use.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Comparable image quality for the Starguide system, to that of the conventional gamma camera, was achieved for bone scan (4 min/bed position [BP] using a relative difference prior [RDP] with gamma 2 and beta 0.4, along with 10 iterations and 10 subsets), cardiac scan (8 min [stress] and 3 min 20 s [rest] using median root prior [MRP] with beta 0.07 non attenuation corrected and 0.008 attenuation corrected and 50 interations and 10 subsets for both stress and rest) and lung scan (10 min [vent] and 5 min [perf] using RDP with gamma 0.5 and beta 0.03 [vent] and 0.02 [perf] and 20 interations and 10 subsets for both vent and perf).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>It was possible to transition from a conventional gamma camera to the Starguide system as part of the clinical routine, with acceptable image quality. Images from the Starguide system were deemed to be at least as good as those from a conventional gamma camera.</p>\\n </section>\\n </div>\",\"PeriodicalId\":10504,\"journal\":{\"name\":\"Clinical Physiology and Functional Imaging\",\"volume\":\"44 1\",\"pages\":\"79-88\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2023-08-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1111/cpf.12853\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical Physiology and Functional Imaging\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/cpf.12853\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"PHYSIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Physiology and Functional Imaging","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/cpf.12853","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHYSIOLOGY","Score":null,"Total":0}
First clinical experience of a ring-configured cadmium zinc telluride camera: A comparative study versus conventional gamma camera systems
Background
A novel semiconductor cadmium zinc telluride (CZT) gamma camera system using a block sequential regularized expectation maximization (BSREM) reconstruction algorithm is now clinically available. Here we investigate how a multi-purpose ring-configurated CZT system can be safely applied in clinics and describe the initial optimization process.
Method
Seventy-six patients (bone-, cardiac- and lung scan) were scanned on a conventional gamma camera (planar and/or single-photon emission computed tomography [SPECT]/SPECT-CT) used in clinical routine and on the ring-configurated CZT camera Starguide (GE Healthcare). These data were used to validate and optimize the Starguide system for routine clinical use.
Results
Comparable image quality for the Starguide system, to that of the conventional gamma camera, was achieved for bone scan (4 min/bed position [BP] using a relative difference prior [RDP] with gamma 2 and beta 0.4, along with 10 iterations and 10 subsets), cardiac scan (8 min [stress] and 3 min 20 s [rest] using median root prior [MRP] with beta 0.07 non attenuation corrected and 0.008 attenuation corrected and 50 interations and 10 subsets for both stress and rest) and lung scan (10 min [vent] and 5 min [perf] using RDP with gamma 0.5 and beta 0.03 [vent] and 0.02 [perf] and 20 interations and 10 subsets for both vent and perf).
Conclusions
It was possible to transition from a conventional gamma camera to the Starguide system as part of the clinical routine, with acceptable image quality. Images from the Starguide system were deemed to be at least as good as those from a conventional gamma camera.
期刊介绍:
Clinical Physiology and Functional Imaging publishes reports on clinical and experimental research pertinent to human physiology in health and disease. The scope of the Journal is very broad, covering all aspects of the regulatory system in the cardiovascular, renal and pulmonary systems with special emphasis on methodological aspects. The focus for the journal is, however, work that has potential clinical relevance. The Journal also features review articles on recent front-line research within these fields of interest.
Covered by the major abstracting services including Current Contents and Science Citation Index, Clinical Physiology and Functional Imaging plays an important role in providing effective and productive communication among clinical physiologists world-wide.