{"title":"机器人辅助骨科手术的铣削精度","authors":"D. Meister, P. Pokrandt, A. Both","doi":"10.1109/IECON.1998.724120","DOIUrl":null,"url":null,"abstract":"Experiments with industrial robot systems showed that the accuracy of these systems is not sufficient for many surgical applications. Hence, we developed an algorithm which takes the trajectory generator of the robot control system into account. Furthermore, the inhomogeneous milling characteristics of a bone are predicted by its computed tomography (CT). Results from experiments demonstrated that the demanded accuracy can be achieved using this approach. Since December 1997 several patients have been treated with our surgical robot CASPAR, which encapsulates this strategy in the field of total hip replacement (THR).","PeriodicalId":377136,"journal":{"name":"IECON '98. Proceedings of the 24th Annual Conference of the IEEE Industrial Electronics Society (Cat. No.98CH36200)","volume":"45 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"18","resultStr":"{\"title\":\"Milling accuracy in robot assisted orthopaedic surgery\",\"authors\":\"D. Meister, P. Pokrandt, A. Both\",\"doi\":\"10.1109/IECON.1998.724120\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Experiments with industrial robot systems showed that the accuracy of these systems is not sufficient for many surgical applications. Hence, we developed an algorithm which takes the trajectory generator of the robot control system into account. Furthermore, the inhomogeneous milling characteristics of a bone are predicted by its computed tomography (CT). Results from experiments demonstrated that the demanded accuracy can be achieved using this approach. Since December 1997 several patients have been treated with our surgical robot CASPAR, which encapsulates this strategy in the field of total hip replacement (THR).\",\"PeriodicalId\":377136,\"journal\":{\"name\":\"IECON '98. Proceedings of the 24th Annual Conference of the IEEE Industrial Electronics Society (Cat. No.98CH36200)\",\"volume\":\"45 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-08-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"18\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IECON '98. Proceedings of the 24th Annual Conference of the IEEE Industrial Electronics Society (Cat. No.98CH36200)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IECON.1998.724120\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IECON '98. Proceedings of the 24th Annual Conference of the IEEE Industrial Electronics Society (Cat. No.98CH36200)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IECON.1998.724120","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Milling accuracy in robot assisted orthopaedic surgery
Experiments with industrial robot systems showed that the accuracy of these systems is not sufficient for many surgical applications. Hence, we developed an algorithm which takes the trajectory generator of the robot control system into account. Furthermore, the inhomogeneous milling characteristics of a bone are predicted by its computed tomography (CT). Results from experiments demonstrated that the demanded accuracy can be achieved using this approach. Since December 1997 several patients have been treated with our surgical robot CASPAR, which encapsulates this strategy in the field of total hip replacement (THR).