M. Kennedy, D. Ristau, G. Dumitru, D. Sporea, Clementina A. Timus
{"title":"10.6 μm激光量热计校准程序","authors":"M. Kennedy, D. Ristau, G. Dumitru, D. Sporea, Clementina A. Timus","doi":"10.1117/12.312718","DOIUrl":null,"url":null,"abstract":"The paper presents the theoretical background of laser calorimetry as a method to investigate the absorptance of optical components, according to the recently adopted ISO Standard 11 551. An experimental set-up for optical absorption measurements, including a cw CO2 laser and a calorimetric chamber, is presented. The calibration of the laser calorimetric set-up is also described. The calibration involved several steps: laser power calibration, evaluation of chamber sensitivity, electrical calibration using electrically heated samples and optical calibration utilizing 100% absorptance samples. There were measured the conversion factors for the laser power and for the sample temperature; there were calculated over-all correction factors for the different substrates and they could be placed within +/- 4% around their mean value.","PeriodicalId":383583,"journal":{"name":"ROMOPTO International Conference on Micro- to Nano- Photonics III","volume":"114 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-07-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Calibration procedures for a 10.6-μm laser calorimeter\",\"authors\":\"M. Kennedy, D. Ristau, G. Dumitru, D. Sporea, Clementina A. Timus\",\"doi\":\"10.1117/12.312718\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The paper presents the theoretical background of laser calorimetry as a method to investigate the absorptance of optical components, according to the recently adopted ISO Standard 11 551. An experimental set-up for optical absorption measurements, including a cw CO2 laser and a calorimetric chamber, is presented. The calibration of the laser calorimetric set-up is also described. The calibration involved several steps: laser power calibration, evaluation of chamber sensitivity, electrical calibration using electrically heated samples and optical calibration utilizing 100% absorptance samples. There were measured the conversion factors for the laser power and for the sample temperature; there were calculated over-all correction factors for the different substrates and they could be placed within +/- 4% around their mean value.\",\"PeriodicalId\":383583,\"journal\":{\"name\":\"ROMOPTO International Conference on Micro- to Nano- Photonics III\",\"volume\":\"114 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-07-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ROMOPTO International Conference on Micro- to Nano- Photonics III\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1117/12.312718\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ROMOPTO International Conference on Micro- to Nano- Photonics III","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.312718","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Calibration procedures for a 10.6-μm laser calorimeter
The paper presents the theoretical background of laser calorimetry as a method to investigate the absorptance of optical components, according to the recently adopted ISO Standard 11 551. An experimental set-up for optical absorption measurements, including a cw CO2 laser and a calorimetric chamber, is presented. The calibration of the laser calorimetric set-up is also described. The calibration involved several steps: laser power calibration, evaluation of chamber sensitivity, electrical calibration using electrically heated samples and optical calibration utilizing 100% absorptance samples. There were measured the conversion factors for the laser power and for the sample temperature; there were calculated over-all correction factors for the different substrates and they could be placed within +/- 4% around their mean value.