{"title":"使用安捷伦3458A采样能力精确测量音频频段的交流电压","authors":"R. Lapuh, B. Voljc, M. Lindic","doi":"10.1109/CPEM.2014.6898598","DOIUrl":null,"url":null,"abstract":"The paper describes frequency dependence correction of the Agilent 3458A DMM in DCV sampling mode. The biggest error source identified is the input filter roll-off, which was measured and evaluated with an asynchronous sampling technique using phase sensitive sine fit algorithm. Corrected response is flat within 50 μV/V in the whole audio frequency band.","PeriodicalId":256575,"journal":{"name":"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)","volume":"10 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":"{\"title\":\"Accurate measurement of AC voltage in audio band using Agilent 3458A sampling capability\",\"authors\":\"R. Lapuh, B. Voljc, M. Lindic\",\"doi\":\"10.1109/CPEM.2014.6898598\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The paper describes frequency dependence correction of the Agilent 3458A DMM in DCV sampling mode. The biggest error source identified is the input filter roll-off, which was measured and evaluated with an asynchronous sampling technique using phase sensitive sine fit algorithm. Corrected response is flat within 50 μV/V in the whole audio frequency band.\",\"PeriodicalId\":256575,\"journal\":{\"name\":\"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)\",\"volume\":\"10 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-10-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"8\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/CPEM.2014.6898598\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CPEM.2014.6898598","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Accurate measurement of AC voltage in audio band using Agilent 3458A sampling capability
The paper describes frequency dependence correction of the Agilent 3458A DMM in DCV sampling mode. The biggest error source identified is the input filter roll-off, which was measured and evaluated with an asynchronous sampling technique using phase sensitive sine fit algorithm. Corrected response is flat within 50 μV/V in the whole audio frequency band.