{"title":"热电偶阵列b模超声引导相控阵聚焦超声系统瞄准精度的实验评价","authors":"Bo-fan Zhou, Shou-bei Wang, Ke Li, Jing-feng Bai, Xiang Ji","doi":"10.1109/SPAWDA.2015.7364486","DOIUrl":null,"url":null,"abstract":"Image-guide high-intensity focused ultrasound (HIFU) has been widely used in the treatment of uterine fibroids. However, before HIFU system is ready for use, the targeting accuracy should be known for the safety of important abdominal organs surrounding the target. Researchers have used magnetic resonance (MR) images or computed tomography (CT) to evaluate the targeting accuracy of current MR-guided HIFU systems, whereas these methods cannot be directly used in the ultrasound-guided HIFU (USgHIFU) systems. In this study, a thermocouple array is used in combination with the in-house built acrylic plate to evaluate the targeting accuracy of the existing 112-channel phased-array USgHIFU system. The results of ex vivo experiment demonstrate the feasibility and the efficacy of the thermocouple array in the evaluation, and the targeting accuracy of the USgHIFU system is about 1mm in the focal plane. Thus, this accuracy is adequate for the treatment of uterine fibroids.","PeriodicalId":205914,"journal":{"name":"2015 Symposium on Piezoelectricity, Acoustic Waves, and Device Applications (SPAWDA)","volume":"30 3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Experimental evaluation of targeting accuracy of a B-mode ultrasound-guided phased-array focused ultrasound system using a thermocouple array\",\"authors\":\"Bo-fan Zhou, Shou-bei Wang, Ke Li, Jing-feng Bai, Xiang Ji\",\"doi\":\"10.1109/SPAWDA.2015.7364486\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Image-guide high-intensity focused ultrasound (HIFU) has been widely used in the treatment of uterine fibroids. However, before HIFU system is ready for use, the targeting accuracy should be known for the safety of important abdominal organs surrounding the target. Researchers have used magnetic resonance (MR) images or computed tomography (CT) to evaluate the targeting accuracy of current MR-guided HIFU systems, whereas these methods cannot be directly used in the ultrasound-guided HIFU (USgHIFU) systems. In this study, a thermocouple array is used in combination with the in-house built acrylic plate to evaluate the targeting accuracy of the existing 112-channel phased-array USgHIFU system. The results of ex vivo experiment demonstrate the feasibility and the efficacy of the thermocouple array in the evaluation, and the targeting accuracy of the USgHIFU system is about 1mm in the focal plane. Thus, this accuracy is adequate for the treatment of uterine fibroids.\",\"PeriodicalId\":205914,\"journal\":{\"name\":\"2015 Symposium on Piezoelectricity, Acoustic Waves, and Device Applications (SPAWDA)\",\"volume\":\"30 3 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2015-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2015 Symposium on Piezoelectricity, Acoustic Waves, and Device Applications (SPAWDA)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/SPAWDA.2015.7364486\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 Symposium on Piezoelectricity, Acoustic Waves, and Device Applications (SPAWDA)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SPAWDA.2015.7364486","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

图像引导高强度聚焦超声(HIFU)已广泛应用于子宫肌瘤的治疗。然而,在HIFU系统准备使用之前,应了解靶向准确性以及目标周围重要腹部器官的安全性。研究人员已经使用磁共振(MR)图像或计算机断层扫描(CT)来评估当前磁共振引导HIFU系统的靶向准确性,然而这些方法不能直接用于超声引导HIFU (USgHIFU)系统。在本研究中,热电偶阵列与内部自制的丙烯酸板结合使用,以评估现有112通道相控阵USgHIFU系统的瞄准精度。离体实验结果验证了热电偶阵列在评估中的可行性和有效性,USgHIFU系统在焦平面内的瞄准精度约为1mm。因此,这种准确性是足够的治疗子宫肌瘤。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Experimental evaluation of targeting accuracy of a B-mode ultrasound-guided phased-array focused ultrasound system using a thermocouple array
Image-guide high-intensity focused ultrasound (HIFU) has been widely used in the treatment of uterine fibroids. However, before HIFU system is ready for use, the targeting accuracy should be known for the safety of important abdominal organs surrounding the target. Researchers have used magnetic resonance (MR) images or computed tomography (CT) to evaluate the targeting accuracy of current MR-guided HIFU systems, whereas these methods cannot be directly used in the ultrasound-guided HIFU (USgHIFU) systems. In this study, a thermocouple array is used in combination with the in-house built acrylic plate to evaluate the targeting accuracy of the existing 112-channel phased-array USgHIFU system. The results of ex vivo experiment demonstrate the feasibility and the efficacy of the thermocouple array in the evaluation, and the targeting accuracy of the USgHIFU system is about 1mm in the focal plane. Thus, this accuracy is adequate for the treatment of uterine fibroids.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Potential of Al2O3/GaN/Sapphire layered structure for high temperature SAW sensors Application of SAW gas chromatography in the early screening of lung cancer Research on the loss-less compression algorithm of the ultrasonic testing of rail First-principles study on the electronic structure and optical properties of Mn-doped NaNbO3 Development of cryptophane A-coated SAW methane gas sensor
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1