Weiqiang Zhao, Hui Liu, Jinyun Yan, Ying Su, Lin Jiang, Zhifeng Wu
{"title":"光度计光谱响应测量使用opo可调谐激光","authors":"Weiqiang Zhao, Hui Liu, Jinyun Yan, Ying Su, Lin Jiang, Zhifeng Wu","doi":"10.25039/X46.2019.OP79","DOIUrl":null,"url":null,"abstract":"This paper shows an photometer spectral response measurement method using an pulsed tunable laser.The laser is the comparator. the standard detector is a three-silicon S1337 trap detector and the detector under test is a silicon photodiode with a V(λ)-correction filter. Due to the low noise and strong signal of the laser, we believe it will give a lower measurement uncertainty compared to traditional method for measuring S(λ) of the photometer","PeriodicalId":121032,"journal":{"name":"PROCEEDINGS OF the 29th Quadrennial Session of the CIE","volume":"47 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-06-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"PHOTOMETER SPECTRAL RESPONSE MEASUREMENT USING OPO TUNABLE LASER\",\"authors\":\"Weiqiang Zhao, Hui Liu, Jinyun Yan, Ying Su, Lin Jiang, Zhifeng Wu\",\"doi\":\"10.25039/X46.2019.OP79\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper shows an photometer spectral response measurement method using an pulsed tunable laser.The laser is the comparator. the standard detector is a three-silicon S1337 trap detector and the detector under test is a silicon photodiode with a V(λ)-correction filter. Due to the low noise and strong signal of the laser, we believe it will give a lower measurement uncertainty compared to traditional method for measuring S(λ) of the photometer\",\"PeriodicalId\":121032,\"journal\":{\"name\":\"PROCEEDINGS OF the 29th Quadrennial Session of the CIE\",\"volume\":\"47 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-06-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"PROCEEDINGS OF the 29th Quadrennial Session of the CIE\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.25039/X46.2019.OP79\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"PROCEEDINGS OF the 29th Quadrennial Session of the CIE","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.25039/X46.2019.OP79","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
PHOTOMETER SPECTRAL RESPONSE MEASUREMENT USING OPO TUNABLE LASER
This paper shows an photometer spectral response measurement method using an pulsed tunable laser.The laser is the comparator. the standard detector is a three-silicon S1337 trap detector and the detector under test is a silicon photodiode with a V(λ)-correction filter. Due to the low noise and strong signal of the laser, we believe it will give a lower measurement uncertainty compared to traditional method for measuring S(λ) of the photometer