{"title":"超高剂量率x射线辐照灭菌医疗设备","authors":"N. Barinov, G.I. Dolgachev, D. D. Maslennikov, M.S. Nitishinsky, A. Ushakov, L.P. Zakatov, O. Zinoviev","doi":"10.1109/BEAMS.1998.817016","DOIUrl":null,"url":null,"abstract":"The results of sterilization tests made at the RS-20 generator of Bremsstrahlung for sterilizing purpose are reported. The measurements were done under different conditions in the area of 10-30 kGy dose levels at 1 MGy/s peak dose rate range for various microorganisms. The immediately irradiated volume was 0.036 of cubic meter with average dose rate at 1 kGy/hour provided in the center of the volume. The Bremsstrahlung was created by 2.0 MeV 10 kA 100 ns pulsed e-beam directed toward the target at 2 Hz. The tests were compared with those made simultaneously at continuous radionucleic radiation source using Co-60 isotope providing same dose values. It was found that full sterilization effect at RS-20 using 1 MGy/s peak dose rate takes 2-5 times lesser dose against radionucleic source and increases total efficiency of the system. Total dose decrease down to 5-12 kGy allows to reduce radiation loads on the equipment and broadens the area of sterilizing application.","PeriodicalId":410823,"journal":{"name":"12th International Conference on High-Power Particle Beams. BEAMS'98. Proceedings (Cat. No.98EX103)","volume":"167 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-06-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Medical equipment sterilization using superhigh dose rate X-ray irradiation\",\"authors\":\"N. Barinov, G.I. Dolgachev, D. D. Maslennikov, M.S. Nitishinsky, A. Ushakov, L.P. Zakatov, O. Zinoviev\",\"doi\":\"10.1109/BEAMS.1998.817016\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The results of sterilization tests made at the RS-20 generator of Bremsstrahlung for sterilizing purpose are reported. The measurements were done under different conditions in the area of 10-30 kGy dose levels at 1 MGy/s peak dose rate range for various microorganisms. The immediately irradiated volume was 0.036 of cubic meter with average dose rate at 1 kGy/hour provided in the center of the volume. The Bremsstrahlung was created by 2.0 MeV 10 kA 100 ns pulsed e-beam directed toward the target at 2 Hz. The tests were compared with those made simultaneously at continuous radionucleic radiation source using Co-60 isotope providing same dose values. It was found that full sterilization effect at RS-20 using 1 MGy/s peak dose rate takes 2-5 times lesser dose against radionucleic source and increases total efficiency of the system. Total dose decrease down to 5-12 kGy allows to reduce radiation loads on the equipment and broadens the area of sterilizing application.\",\"PeriodicalId\":410823,\"journal\":{\"name\":\"12th International Conference on High-Power Particle Beams. BEAMS'98. Proceedings (Cat. No.98EX103)\",\"volume\":\"167 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-06-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"12th International Conference on High-Power Particle Beams. BEAMS'98. Proceedings (Cat. No.98EX103)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/BEAMS.1998.817016\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"12th International Conference on High-Power Particle Beams. BEAMS'98. Proceedings (Cat. No.98EX103)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BEAMS.1998.817016","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Medical equipment sterilization using superhigh dose rate X-ray irradiation
The results of sterilization tests made at the RS-20 generator of Bremsstrahlung for sterilizing purpose are reported. The measurements were done under different conditions in the area of 10-30 kGy dose levels at 1 MGy/s peak dose rate range for various microorganisms. The immediately irradiated volume was 0.036 of cubic meter with average dose rate at 1 kGy/hour provided in the center of the volume. The Bremsstrahlung was created by 2.0 MeV 10 kA 100 ns pulsed e-beam directed toward the target at 2 Hz. The tests were compared with those made simultaneously at continuous radionucleic radiation source using Co-60 isotope providing same dose values. It was found that full sterilization effect at RS-20 using 1 MGy/s peak dose rate takes 2-5 times lesser dose against radionucleic source and increases total efficiency of the system. Total dose decrease down to 5-12 kGy allows to reduce radiation loads on the equipment and broadens the area of sterilizing application.