{"title":"平板探测器中的串扰对 CT 图像质量的影响","authors":"Seemeen Karimi;James Hall;Jerel A. Smith;Joseph Tringe","doi":"10.1109/OJIM.2023.3332342","DOIUrl":null,"url":null,"abstract":"Spatial resolution and image noise are two aspects of image quality of an X-ray computed tomography (CT) system and are determined by the X-ray source, the detector, and mathematical operations for image reconstruction. In CT scanners with flat panel detectors (FPDs), there is cross-talk (signal leakage) between detector pixels. The contribution of the cross-talk to spatial resolution and noise in reconstructed images has not been adequately modeled. Previously, we estimated cross-talk from autocovariance measurements in air, and modeled the impact of cross-talk on spatial resolution. We have extended that work to calculate the impact of cross-talk on signal-to-noise ratio in radiographs and to reconstructed image noise. We modeled the spatial resolution and noise of a CT scanner that uses a flat-panel detector with 0.2-mm pixels and a gadolinium oxysulfide scintillator, and a 450 kVp, dual-focus X-ray tube. Our noise model agrees with measurements from experimental data and simulations to within 10%. We show that cross-talk in FPDs can reduce resolution by over 30%, reduce noise by approximately a factor of two, and introduce correlation in the noise, and therefore, cannot be disregarded when assessing CT image quality.","PeriodicalId":100630,"journal":{"name":"IEEE Open Journal of Instrumentation and Measurement","volume":"2 ","pages":"1-9"},"PeriodicalIF":0.0000,"publicationDate":"2023-11-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10317879","citationCount":"0","resultStr":"{\"title\":\"The Impact of Cross-Talk in a Flat Panel Detector on CT Image Quality\",\"authors\":\"Seemeen Karimi;James Hall;Jerel A. Smith;Joseph Tringe\",\"doi\":\"10.1109/OJIM.2023.3332342\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Spatial resolution and image noise are two aspects of image quality of an X-ray computed tomography (CT) system and are determined by the X-ray source, the detector, and mathematical operations for image reconstruction. In CT scanners with flat panel detectors (FPDs), there is cross-talk (signal leakage) between detector pixels. The contribution of the cross-talk to spatial resolution and noise in reconstructed images has not been adequately modeled. Previously, we estimated cross-talk from autocovariance measurements in air, and modeled the impact of cross-talk on spatial resolution. We have extended that work to calculate the impact of cross-talk on signal-to-noise ratio in radiographs and to reconstructed image noise. We modeled the spatial resolution and noise of a CT scanner that uses a flat-panel detector with 0.2-mm pixels and a gadolinium oxysulfide scintillator, and a 450 kVp, dual-focus X-ray tube. Our noise model agrees with measurements from experimental data and simulations to within 10%. We show that cross-talk in FPDs can reduce resolution by over 30%, reduce noise by approximately a factor of two, and introduce correlation in the noise, and therefore, cannot be disregarded when assessing CT image quality.\",\"PeriodicalId\":100630,\"journal\":{\"name\":\"IEEE Open Journal of Instrumentation and Measurement\",\"volume\":\"2 \",\"pages\":\"1-9\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-11-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10317879\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE Open Journal of Instrumentation and Measurement\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://ieeexplore.ieee.org/document/10317879/\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Open Journal of Instrumentation and Measurement","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10317879/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
The Impact of Cross-Talk in a Flat Panel Detector on CT Image Quality
Spatial resolution and image noise are two aspects of image quality of an X-ray computed tomography (CT) system and are determined by the X-ray source, the detector, and mathematical operations for image reconstruction. In CT scanners with flat panel detectors (FPDs), there is cross-talk (signal leakage) between detector pixels. The contribution of the cross-talk to spatial resolution and noise in reconstructed images has not been adequately modeled. Previously, we estimated cross-talk from autocovariance measurements in air, and modeled the impact of cross-talk on spatial resolution. We have extended that work to calculate the impact of cross-talk on signal-to-noise ratio in radiographs and to reconstructed image noise. We modeled the spatial resolution and noise of a CT scanner that uses a flat-panel detector with 0.2-mm pixels and a gadolinium oxysulfide scintillator, and a 450 kVp, dual-focus X-ray tube. Our noise model agrees with measurements from experimental data and simulations to within 10%. We show that cross-talk in FPDs can reduce resolution by over 30%, reduce noise by approximately a factor of two, and introduce correlation in the noise, and therefore, cannot be disregarded when assessing CT image quality.