{"title":"通过硅基化提高火焰电离探测器的响应:羧酸的有效碳数","authors":"Judit Mátyási, Gyula Nyerges, József Balla","doi":"10.3311/ppch.22827","DOIUrl":null,"url":null,"abstract":"Detector response of carboxylic acids (C2–C12, straight and branched chain) were investigated using a flame ionization detector (FID) in a capillary gas chromatographic system. The response of the FID for hydrocarbons is almost directly proportional to the carbon quantity introduced into the flame. Heteroatoms in the molecule reduce signal magnitude, depending on their quality and on the bond they are involved in. We expressed this reduced response with the effective carbon number (ECN). We determined the ECN contribution (ΔECN) of the carboxyl group on the alkyl skeleton. We examined how the responses of carboxylic acids change if trimethylsilyl derivatives are evaluated and we compared the ECN of the neat and derivatized form.","PeriodicalId":19922,"journal":{"name":"Periodica Polytechnica Chemical Engineering","volume":"25 1","pages":"0"},"PeriodicalIF":1.4000,"publicationDate":"2023-09-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Increasing Flame Ionization Detector Response by Silylation: The Effective Carbon Number of Carboxylic Acids\",\"authors\":\"Judit Mátyási, Gyula Nyerges, József Balla\",\"doi\":\"10.3311/ppch.22827\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Detector response of carboxylic acids (C2–C12, straight and branched chain) were investigated using a flame ionization detector (FID) in a capillary gas chromatographic system. The response of the FID for hydrocarbons is almost directly proportional to the carbon quantity introduced into the flame. Heteroatoms in the molecule reduce signal magnitude, depending on their quality and on the bond they are involved in. We expressed this reduced response with the effective carbon number (ECN). We determined the ECN contribution (ΔECN) of the carboxyl group on the alkyl skeleton. We examined how the responses of carboxylic acids change if trimethylsilyl derivatives are evaluated and we compared the ECN of the neat and derivatized form.\",\"PeriodicalId\":19922,\"journal\":{\"name\":\"Periodica Polytechnica Chemical Engineering\",\"volume\":\"25 1\",\"pages\":\"0\"},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2023-09-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Periodica Polytechnica Chemical Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.3311/ppch.22827\",\"RegionNum\":4,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, CHEMICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Periodica Polytechnica Chemical Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3311/ppch.22827","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, CHEMICAL","Score":null,"Total":0}
Increasing Flame Ionization Detector Response by Silylation: The Effective Carbon Number of Carboxylic Acids
Detector response of carboxylic acids (C2–C12, straight and branched chain) were investigated using a flame ionization detector (FID) in a capillary gas chromatographic system. The response of the FID for hydrocarbons is almost directly proportional to the carbon quantity introduced into the flame. Heteroatoms in the molecule reduce signal magnitude, depending on their quality and on the bond they are involved in. We expressed this reduced response with the effective carbon number (ECN). We determined the ECN contribution (ΔECN) of the carboxyl group on the alkyl skeleton. We examined how the responses of carboxylic acids change if trimethylsilyl derivatives are evaluated and we compared the ECN of the neat and derivatized form.
期刊介绍:
The main scope of the journal is to publish original research articles in the wide field of chemical engineering including environmental and bioengineering.