用于头颈癌放疗计划的基于人工智能的自动轮廓解决方案

Virginia Marin Anaya
{"title":"用于头颈癌放疗计划的基于人工智能的自动轮廓解决方案","authors":"Virginia Marin Anaya","doi":"10.1016/j.ipemt.2023.100018","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Manual contouring is time-consuming and subjective. Thus, auto-segmentation methods, which can be deployed in the existing workflow, are needed. The objective of this study was to assess the feasibility of Limbus AI and AI Rad Companion auto-contours for head and neck treatment planning.</p></div><div><h3>Methods</h3><p>Head and neck patients treated with RapidArc were selected retrospectively. The manual contours on the planning CT were used as reference. Geometric analysis of the auto-contours was performed using several evaluation metrics such as the Dice Similarity Coefficient (DSC) and the Mean Distance to Conformity (MDC). Dosimetric analysis was performed by recalculating the original plan on the auto-contours and comparing Dose Volume Histogram (DVH) metrics to the original plan.</p></div><div><h3>Results and discussion</h3><p>Both AI tools tend to underestimate the volumes of brainstem and cord. For brainstem and parotids, median DSC values were ≥ 0.8. For all auto-contours, median MDC values were ∼ 3–6 mm. Median differences were found of up to ±7 % in DVH points on the auto-contours relative to the planning CT contours, but these were not statistically-significant.</p></div><div><h3>Conclusion</h3><p>The auto-contours could be used as a starting point to assist the clinician with the manual contouring of structures on the planning and re-scanning planning CT.</p></div>","PeriodicalId":73507,"journal":{"name":"IPEM-translation","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-11-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2667258823000031/pdfft?md5=9b7269c600230bd5b8b66a2cab079e41&pid=1-s2.0-S2667258823000031-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Artificial intelligence based auto-contouring solutions for use in radiotherapy treatment planning of head and neck cancer\",\"authors\":\"Virginia Marin Anaya\",\"doi\":\"10.1016/j.ipemt.2023.100018\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>Manual contouring is time-consuming and subjective. Thus, auto-segmentation methods, which can be deployed in the existing workflow, are needed. The objective of this study was to assess the feasibility of Limbus AI and AI Rad Companion auto-contours for head and neck treatment planning.</p></div><div><h3>Methods</h3><p>Head and neck patients treated with RapidArc were selected retrospectively. The manual contours on the planning CT were used as reference. Geometric analysis of the auto-contours was performed using several evaluation metrics such as the Dice Similarity Coefficient (DSC) and the Mean Distance to Conformity (MDC). Dosimetric analysis was performed by recalculating the original plan on the auto-contours and comparing Dose Volume Histogram (DVH) metrics to the original plan.</p></div><div><h3>Results and discussion</h3><p>Both AI tools tend to underestimate the volumes of brainstem and cord. For brainstem and parotids, median DSC values were ≥ 0.8. For all auto-contours, median MDC values were ∼ 3–6 mm. Median differences were found of up to ±7 % in DVH points on the auto-contours relative to the planning CT contours, but these were not statistically-significant.</p></div><div><h3>Conclusion</h3><p>The auto-contours could be used as a starting point to assist the clinician with the manual contouring of structures on the planning and re-scanning planning CT.</p></div>\",\"PeriodicalId\":73507,\"journal\":{\"name\":\"IPEM-translation\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-11-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2667258823000031/pdfft?md5=9b7269c600230bd5b8b66a2cab079e41&pid=1-s2.0-S2667258823000031-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IPEM-translation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2667258823000031\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IPEM-translation","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2667258823000031","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

手工轮廓是费时和主观的。因此,需要在现有工作流中部署的自动分割方法。本研究的目的是评估Limbus AI和AI Rad Companion自动轮廓在头颈部治疗计划中的可行性。方法回顾性分析使用RapidArc治疗的头颈部患者。参考规划CT上的手工等高线。使用骰子相似系数(DSC)和平均一致性距离(MDC)等几个评估指标对自动轮廓进行几何分析。通过在自动轮廓上重新计算原计划并将剂量体积直方图(DVH)指标与原计划进行比较,进行剂量学分析。结果和讨论两种人工智能工具都倾向于低估脑干和脊髓的体积。脑干和腮腺的DSC中位数≥0.8。对于所有自动轮廓,中位MDC值为~ 3-6 mm。相对于规划CT轮廓,自动轮廓上的DVH点的中位数差异高达±7%,但这些差异没有统计学意义。结论自动轮廓可以作为辅助临床医生在规划和重扫描规划CT上进行人工结构轮廓的起点。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Artificial intelligence based auto-contouring solutions for use in radiotherapy treatment planning of head and neck cancer

Background

Manual contouring is time-consuming and subjective. Thus, auto-segmentation methods, which can be deployed in the existing workflow, are needed. The objective of this study was to assess the feasibility of Limbus AI and AI Rad Companion auto-contours for head and neck treatment planning.

Methods

Head and neck patients treated with RapidArc were selected retrospectively. The manual contours on the planning CT were used as reference. Geometric analysis of the auto-contours was performed using several evaluation metrics such as the Dice Similarity Coefficient (DSC) and the Mean Distance to Conformity (MDC). Dosimetric analysis was performed by recalculating the original plan on the auto-contours and comparing Dose Volume Histogram (DVH) metrics to the original plan.

Results and discussion

Both AI tools tend to underestimate the volumes of brainstem and cord. For brainstem and parotids, median DSC values were ≥ 0.8. For all auto-contours, median MDC values were ∼ 3–6 mm. Median differences were found of up to ±7 % in DVH points on the auto-contours relative to the planning CT contours, but these were not statistically-significant.

Conclusion

The auto-contours could be used as a starting point to assist the clinician with the manual contouring of structures on the planning and re-scanning planning CT.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
IPEM-translation
IPEM-translation Medicine and Dentistry (General)
自引率
0.00%
发文量
0
审稿时长
63 days
期刊最新文献
Experimental measurement of dosimetric parameters relevant to radioactive needlestick injury The use of solar film elements on a neonate manikin surface to estimate the received output power of neonatal phototherapy lamp systems Role of Coriolis flow measurement technology in validation of model of syringe driver performance Photoplethysmogram-based heart rate and blood pressure estimation with hypertension classification Towards clinical translation of 7 Tesla MRI in the human brain
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1