前臂旋转对桡骨远端骨密度测量的影响。

IF 2.4 3区 医学 Q3 ENDOCRINOLOGY & METABOLISM Journal of Bone and Mineral Metabolism Pub Date : 2024-01-01 Epub Date: 2023-12-06 DOI:10.1007/s00774-023-01473-4
Keisuke Uemura, Satoshi Miyamura, Yoshito Otake, Hirokazu Mae, Kazuma Takashima, Hidetoshi Hamada, Kosuke Ebina, Tsuyoshi Murase, Yoshinobu Sato, Seiji Okada
{"title":"前臂旋转对桡骨远端骨密度测量的影响。","authors":"Keisuke Uemura, Satoshi Miyamura, Yoshito Otake, Hirokazu Mae, Kazuma Takashima, Hidetoshi Hamada, Kosuke Ebina, Tsuyoshi Murase, Yoshinobu Sato, Seiji Okada","doi":"10.1007/s00774-023-01473-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Forearm dual-energy X-ray absorptiometry (DXA) is often performed in clinics where central DXA is unavailable. Accurate bone mineral density (BMD) measurement is crucial for clinical assessment. Forearm rotation can affect BMD measurements, but this effect remains uncertain. Thus, we aimed to conduct a simulation study using CT images to clarify the effect of forearm rotation on BMD measurements.</p><p><strong>Materials and methods: </strong>Forearm CT images of 60 women were analyzed. BMD was measured at the total, ultra-distal (UD), mid-distal (MD), and distal 33% radius regions with the radius located at the neutral position using digitally reconstructed radiographs generated from CT images. Then, the rotation was altered from - 30° to 30° (supination set as positive) with a one-degree increment, and the percent BMD changes from the neutral position were quantified for all regions at each angle for each patient.</p><p><strong>Results: </strong>The maximum mean BMD changes were 5.8%, 7.0%, 6.2%, and 7.2% for the total, UD, MD, and distal 33% radius regions, respectively. The analysis of the absolute values of the percent BMD changes from the neutral position showed that BMD changes of all patients remained within 2% when the rotation was between - 5° and 7° for the total region, between - 3° and 2° for the UD region, between - 4° and 3° for the MD region, and between - 3° and 1° for the distal 33% radius region.</p><p><strong>Conclusion: </strong>Subtle rotational changes affected the BMD measurement of each region. The results showed the importance of forearm positioning when measuring the distal radius BMD.</p>","PeriodicalId":15116,"journal":{"name":"Journal of Bone and Mineral Metabolism","volume":" ","pages":"37-46"},"PeriodicalIF":2.4000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The effect of forearm rotation on the bone mineral density measurements of the distal radius.\",\"authors\":\"Keisuke Uemura, Satoshi Miyamura, Yoshito Otake, Hirokazu Mae, Kazuma Takashima, Hidetoshi Hamada, Kosuke Ebina, Tsuyoshi Murase, Yoshinobu Sato, Seiji Okada\",\"doi\":\"10.1007/s00774-023-01473-4\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Forearm dual-energy X-ray absorptiometry (DXA) is often performed in clinics where central DXA is unavailable. Accurate bone mineral density (BMD) measurement is crucial for clinical assessment. Forearm rotation can affect BMD measurements, but this effect remains uncertain. Thus, we aimed to conduct a simulation study using CT images to clarify the effect of forearm rotation on BMD measurements.</p><p><strong>Materials and methods: </strong>Forearm CT images of 60 women were analyzed. BMD was measured at the total, ultra-distal (UD), mid-distal (MD), and distal 33% radius regions with the radius located at the neutral position using digitally reconstructed radiographs generated from CT images. Then, the rotation was altered from - 30° to 30° (supination set as positive) with a one-degree increment, and the percent BMD changes from the neutral position were quantified for all regions at each angle for each patient.</p><p><strong>Results: </strong>The maximum mean BMD changes were 5.8%, 7.0%, 6.2%, and 7.2% for the total, UD, MD, and distal 33% radius regions, respectively. The analysis of the absolute values of the percent BMD changes from the neutral position showed that BMD changes of all patients remained within 2% when the rotation was between - 5° and 7° for the total region, between - 3° and 2° for the UD region, between - 4° and 3° for the MD region, and between - 3° and 1° for the distal 33% radius region.</p><p><strong>Conclusion: </strong>Subtle rotational changes affected the BMD measurement of each region. The results showed the importance of forearm positioning when measuring the distal radius BMD.</p>\",\"PeriodicalId\":15116,\"journal\":{\"name\":\"Journal of Bone and Mineral Metabolism\",\"volume\":\" \",\"pages\":\"37-46\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Bone and Mineral Metabolism\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s00774-023-01473-4\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/12/6 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Bone and Mineral Metabolism","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00774-023-01473-4","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/6 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

摘要

简介:前臂双能x线吸收仪(DXA)通常在没有中央DXA的诊所进行。准确的骨密度(BMD)测量对临床评估至关重要。前臂旋转可以影响骨密度测量,但这种影响仍不确定。因此,我们旨在利用CT图像进行模拟研究,以阐明前臂旋转对骨密度测量的影响。材料与方法:对60例女性前臂CT图像进行分析。利用CT图像生成的数字重建x线片测量总、超远端(UD)、中远端(MD)和远33%半径区域的骨密度,半径位于中性位置。然后,将旋转从- 30°改变为30°(旋后为正),增加1度,并量化每位患者在每个角度的所有区域从中性位置的BMD变化百分比。结果:骨密度的最大平均变化分别为5.8%、7.0%、6.2%和7.2%,分别为总、UD、MD和远33%桡骨区。从中立位开始的骨密度变化百分比的绝对值分析显示,当旋转在- 5°至7°之间时,所有患者的骨密度变化在- 5°至7°之间,UD区域在- 3°至2°之间,MD区域在- 4°至3°之间,33%桡骨远端在- 3°至1°之间时,所有患者的骨密度变化都保持在2%以内。结论:轻微的旋转变化影响了各部位的骨密度测量。结果显示前臂定位在测量桡骨远端骨密度时的重要性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
The effect of forearm rotation on the bone mineral density measurements of the distal radius.

Introduction: Forearm dual-energy X-ray absorptiometry (DXA) is often performed in clinics where central DXA is unavailable. Accurate bone mineral density (BMD) measurement is crucial for clinical assessment. Forearm rotation can affect BMD measurements, but this effect remains uncertain. Thus, we aimed to conduct a simulation study using CT images to clarify the effect of forearm rotation on BMD measurements.

Materials and methods: Forearm CT images of 60 women were analyzed. BMD was measured at the total, ultra-distal (UD), mid-distal (MD), and distal 33% radius regions with the radius located at the neutral position using digitally reconstructed radiographs generated from CT images. Then, the rotation was altered from - 30° to 30° (supination set as positive) with a one-degree increment, and the percent BMD changes from the neutral position were quantified for all regions at each angle for each patient.

Results: The maximum mean BMD changes were 5.8%, 7.0%, 6.2%, and 7.2% for the total, UD, MD, and distal 33% radius regions, respectively. The analysis of the absolute values of the percent BMD changes from the neutral position showed that BMD changes of all patients remained within 2% when the rotation was between - 5° and 7° for the total region, between - 3° and 2° for the UD region, between - 4° and 3° for the MD region, and between - 3° and 1° for the distal 33% radius region.

Conclusion: Subtle rotational changes affected the BMD measurement of each region. The results showed the importance of forearm positioning when measuring the distal radius BMD.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Bone and Mineral Metabolism
Journal of Bone and Mineral Metabolism 医学-内分泌学与代谢
CiteScore
6.30
自引率
3.00%
发文量
89
审稿时长
6-12 weeks
期刊介绍: The Journal of Bone and Mineral Metabolism (JBMM) provides an international forum for researchers and clinicians to present and discuss topics relevant to bone, teeth, and mineral metabolism, as well as joint and musculoskeletal disorders. The journal welcomes the submission of manuscripts from any country. Membership in the society is not a prerequisite for submission. Acceptance is based on the originality, significance, and validity of the material presented. The journal is aimed at researchers and clinicians dedicated to improvements in research, development, and patient-care in the fields of bone and mineral metabolism.
期刊最新文献
Osteoporosis screening using X-ray assessment and osteoporosis self-assessment tool for Asians in hip surgery patients. Responders and non-responders to romosozumab treatment. Multiple thyroid disorders and risk of osteoporosis: a two-sample Mendelian randomization study. Histological assessments for anabolic effects in teriparatide/abaloparatide administered rodent models. List of reviewers 2024.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1