{"title":"用于分析脑脊液(CSF)分流阀的高分辨率放射学方法试点研究","authors":"Martin P. Pichotka, Moritz Weigt, Mukesch J. Shah, Maximilian F. Russe, Thomas Stein, T. Billoud, Jürgen Beck, Jakob Straehle, Christopher L. Schlett, Dominik v. Elverfeldt, Marco Reisert","doi":"10.1016/j.zemedi.2023.11.001","DOIUrl":null,"url":null,"abstract":"<h3>Objectives</h3><p>Despite their life-saving capabilities, cerebrospinal fluid (CSF) shunts exhibit high failure rates, with a large fraction of failures attributed to the regulating valve. Due to a lack of methods for the detailed analysis of valve malfunctions, failure mechanisms are not well understood, and valves often have to be surgically explanted on the mere suspicion of malfunction.</p><p>The presented pilot study aims to demonstrate radiological methods for comprehensive analysis of CSF shunt valves, considering both the potential for failure analysis in design optimization, and for future clinical in-vivo application to reduce the number of required shunt revision surgeries. The proposed method could also be utilized to develop and support in situ repair methods (e.g. by lysis or ultrasound) of malfunctioning CSF shunt valves.</p><h3>Materials and methods</h3><p>The primary methods described are contrast-enhanced radiographic time series of CSF shunt valves, taken in a favorable projection geometry at low radiation dose, and the machine-learning-based diagnosis of CSF shunt valve obstructions. Complimentarily, we investigate CT-based methods capable of providing accurate ground truth for the training of such diagnostic tools. Using simulated test and training data, the performance of the machine-learning diagnostics in identifying and localizing obstructions within a shunt valve is evaluated regarding per-pixel sensitivity and specificity, the Dice similarity coefficient, and the false positive rate in the case of obstruction free test samples.</p><h3>Results</h3><p>Contrast enhanced subtraction radiography allows high-resolution, time-resolved, low-dose analysis of fluid transport in CSF shunt valves. Complementarily, photon-counting micro-CT allows to investigate valve obstruction mechanisms in detail, and to generate valid ground truth for machine learning-based diagnostics.</p><p>Machine-learning-based detection of valve obstructions in simulated radiographies shows promising results, with a per-pixel sensitivity >70%, per-pixel specificity >90%, a median Dice coefficient >0.8 and <10% false positives at a detection threshold of 0.5.</p><h3>Conclusions</h3><p>This ex-vivo study demonstrates obstruction detection in cerebro-spinal fluid shunt valves, combining radiological methods with machine learning under conditions compatible to future in-vivo application.</p><p>Results indicate that high-resolution contrast-enhanced subtraction radiography, possibly including time-series data, combined with machine-learning image analysis, has the potential to strongly improve the diagnostics of CSF shunt valve failures. The presented method is in principle suitable for in-vivo application, considering both measurement geometry and radiological dose. Further research is needed to validate these results on real-world data and to refine the employed methods.</p><p>In combination, the presented methods enable comprehensive analysis of valve failure mechanisms, paving the way for improved product development and clinical diagnostics of CSF shunt valves.</p>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":"21 1","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2023-12-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Pilot study on high-resolution radiological methods for the analysis of cerebrospinal fluid (CSF) shunt valves\",\"authors\":\"Martin P. Pichotka, Moritz Weigt, Mukesch J. Shah, Maximilian F. Russe, Thomas Stein, T. Billoud, Jürgen Beck, Jakob Straehle, Christopher L. Schlett, Dominik v. Elverfeldt, Marco Reisert\",\"doi\":\"10.1016/j.zemedi.2023.11.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<h3>Objectives</h3><p>Despite their life-saving capabilities, cerebrospinal fluid (CSF) shunts exhibit high failure rates, with a large fraction of failures attributed to the regulating valve. Due to a lack of methods for the detailed analysis of valve malfunctions, failure mechanisms are not well understood, and valves often have to be surgically explanted on the mere suspicion of malfunction.</p><p>The presented pilot study aims to demonstrate radiological methods for comprehensive analysis of CSF shunt valves, considering both the potential for failure analysis in design optimization, and for future clinical in-vivo application to reduce the number of required shunt revision surgeries. The proposed method could also be utilized to develop and support in situ repair methods (e.g. by lysis or ultrasound) of malfunctioning CSF shunt valves.</p><h3>Materials and methods</h3><p>The primary methods described are contrast-enhanced radiographic time series of CSF shunt valves, taken in a favorable projection geometry at low radiation dose, and the machine-learning-based diagnosis of CSF shunt valve obstructions. Complimentarily, we investigate CT-based methods capable of providing accurate ground truth for the training of such diagnostic tools. Using simulated test and training data, the performance of the machine-learning diagnostics in identifying and localizing obstructions within a shunt valve is evaluated regarding per-pixel sensitivity and specificity, the Dice similarity coefficient, and the false positive rate in the case of obstruction free test samples.</p><h3>Results</h3><p>Contrast enhanced subtraction radiography allows high-resolution, time-resolved, low-dose analysis of fluid transport in CSF shunt valves. Complementarily, photon-counting micro-CT allows to investigate valve obstruction mechanisms in detail, and to generate valid ground truth for machine learning-based diagnostics.</p><p>Machine-learning-based detection of valve obstructions in simulated radiographies shows promising results, with a per-pixel sensitivity >70%, per-pixel specificity >90%, a median Dice coefficient >0.8 and <10% false positives at a detection threshold of 0.5.</p><h3>Conclusions</h3><p>This ex-vivo study demonstrates obstruction detection in cerebro-spinal fluid shunt valves, combining radiological methods with machine learning under conditions compatible to future in-vivo application.</p><p>Results indicate that high-resolution contrast-enhanced subtraction radiography, possibly including time-series data, combined with machine-learning image analysis, has the potential to strongly improve the diagnostics of CSF shunt valve failures. The presented method is in principle suitable for in-vivo application, considering both measurement geometry and radiological dose. Further research is needed to validate these results on real-world data and to refine the employed methods.</p><p>In combination, the presented methods enable comprehensive analysis of valve failure mechanisms, paving the way for improved product development and clinical diagnostics of CSF shunt valves.</p>\",\"PeriodicalId\":54397,\"journal\":{\"name\":\"Zeitschrift fur Medizinische Physik\",\"volume\":\"21 1\",\"pages\":\"\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2023-12-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Zeitschrift fur Medizinische Physik\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1016/j.zemedi.2023.11.001\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeitschrift fur Medizinische Physik","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.zemedi.2023.11.001","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Pilot study on high-resolution radiological methods for the analysis of cerebrospinal fluid (CSF) shunt valves
Objectives
Despite their life-saving capabilities, cerebrospinal fluid (CSF) shunts exhibit high failure rates, with a large fraction of failures attributed to the regulating valve. Due to a lack of methods for the detailed analysis of valve malfunctions, failure mechanisms are not well understood, and valves often have to be surgically explanted on the mere suspicion of malfunction.
The presented pilot study aims to demonstrate radiological methods for comprehensive analysis of CSF shunt valves, considering both the potential for failure analysis in design optimization, and for future clinical in-vivo application to reduce the number of required shunt revision surgeries. The proposed method could also be utilized to develop and support in situ repair methods (e.g. by lysis or ultrasound) of malfunctioning CSF shunt valves.
Materials and methods
The primary methods described are contrast-enhanced radiographic time series of CSF shunt valves, taken in a favorable projection geometry at low radiation dose, and the machine-learning-based diagnosis of CSF shunt valve obstructions. Complimentarily, we investigate CT-based methods capable of providing accurate ground truth for the training of such diagnostic tools. Using simulated test and training data, the performance of the machine-learning diagnostics in identifying and localizing obstructions within a shunt valve is evaluated regarding per-pixel sensitivity and specificity, the Dice similarity coefficient, and the false positive rate in the case of obstruction free test samples.
Results
Contrast enhanced subtraction radiography allows high-resolution, time-resolved, low-dose analysis of fluid transport in CSF shunt valves. Complementarily, photon-counting micro-CT allows to investigate valve obstruction mechanisms in detail, and to generate valid ground truth for machine learning-based diagnostics.
Machine-learning-based detection of valve obstructions in simulated radiographies shows promising results, with a per-pixel sensitivity >70%, per-pixel specificity >90%, a median Dice coefficient >0.8 and <10% false positives at a detection threshold of 0.5.
Conclusions
This ex-vivo study demonstrates obstruction detection in cerebro-spinal fluid shunt valves, combining radiological methods with machine learning under conditions compatible to future in-vivo application.
Results indicate that high-resolution contrast-enhanced subtraction radiography, possibly including time-series data, combined with machine-learning image analysis, has the potential to strongly improve the diagnostics of CSF shunt valve failures. The presented method is in principle suitable for in-vivo application, considering both measurement geometry and radiological dose. Further research is needed to validate these results on real-world data and to refine the employed methods.
In combination, the presented methods enable comprehensive analysis of valve failure mechanisms, paving the way for improved product development and clinical diagnostics of CSF shunt valves.
期刊介绍:
Zeitschrift fur Medizinische Physik (Journal of Medical Physics) is an official organ of the German and Austrian Society of Medical Physic and the Swiss Society of Radiobiology and Medical Physics.The Journal is a platform for basic research and practical applications of physical procedures in medical diagnostics and therapy. The articles are reviewed following international standards of peer reviewing.
Focuses of the articles are:
-Biophysical methods in radiation therapy and nuclear medicine
-Dosimetry and radiation protection
-Radiological diagnostics and quality assurance
-Modern imaging techniques, such as computed tomography, magnetic resonance imaging, positron emission tomography
-Ultrasonography diagnostics, application of laser and UV rays
-Electronic processing of biosignals
-Artificial intelligence and machine learning in medical physics
In the Journal, the latest scientific insights find their expression in the form of original articles, reviews, technical communications, and information for the clinical practice.