Erica C. Hord, Melanie P. Hager, Christina M. Bolch, Katherine Bonugli, Lee-Jae Guo, Egemen Tuzun, John C. Criscione
{"title":"用于小儿心脏支持的微创直接心脏压迫装置的临床前概念验证","authors":"Erica C. Hord, Melanie P. Hager, Christina M. Bolch, Katherine Bonugli, Lee-Jae Guo, Egemen Tuzun, John C. Criscione","doi":"10.1007/s13239-023-00703-0","DOIUrl":null,"url":null,"abstract":"<h3 data-test=\"abstract-sub-heading\">Purpose</h3><p>For pediatric patients, extracorporeal membrane oxygenation (ECMO) remains the predominant mechanical circulatory support (MCS) modality for heart failure (HF) although survival to discharge rates remain between 50 and 60% for these patients. The device-blood interface and disruption of physiologic hemodynamics are significant contributors to poor outcomes.</p><h3 data-test=\"abstract-sub-heading\">Methods</h3><p>In this study, we evaluate the preclinical feasibility of a minimally invasive, non-blood-contacting pediatric DCC prototype for temporary MCS. Proof-of-concept is demonstrated in vivo in an animal model of HF. Hemodynamic pressures and flows were examined.</p><h3 data-test=\"abstract-sub-heading\">Results</h3><p>Minimally invasive deployment on the beating heart was successful without cardiopulmonary bypass or anticoagulation. During HF, device operation resulted in an immediate 43% increase in cardiac output while maintaining pulsatile hemodynamics. Compared to the pre-HF baseline, the device recovered up to 95% of ventricular stroke volume. At the conclusion of the study, the device was easily removed from the beating heart.</p><h3 data-test=\"abstract-sub-heading\">Conclusions</h3><p>This preclinical proof-of-concept study demonstrated the feasibility of a DCC device on a pediatric scale that is minimally invasive and non-blood contacting, with promising hemodynamic support and durability for the initial intended duration of use. The ability of DCC to maintain pulsatile MCS without blood contact represents an opportunity to mitigate the mortality and morbidity observed in non-pulsatile, blood-contacting MCS.</p>","PeriodicalId":54322,"journal":{"name":"Cardiovascular Engineering and Technology","volume":"236 1","pages":""},"PeriodicalIF":1.6000,"publicationDate":"2023-12-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Preclinical Proof-of-Concept of a Minimally Invasive Direct Cardiac Compression Device for Pediatric Heart Support\",\"authors\":\"Erica C. Hord, Melanie P. Hager, Christina M. Bolch, Katherine Bonugli, Lee-Jae Guo, Egemen Tuzun, John C. Criscione\",\"doi\":\"10.1007/s13239-023-00703-0\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<h3 data-test=\\\"abstract-sub-heading\\\">Purpose</h3><p>For pediatric patients, extracorporeal membrane oxygenation (ECMO) remains the predominant mechanical circulatory support (MCS) modality for heart failure (HF) although survival to discharge rates remain between 50 and 60% for these patients. The device-blood interface and disruption of physiologic hemodynamics are significant contributors to poor outcomes.</p><h3 data-test=\\\"abstract-sub-heading\\\">Methods</h3><p>In this study, we evaluate the preclinical feasibility of a minimally invasive, non-blood-contacting pediatric DCC prototype for temporary MCS. Proof-of-concept is demonstrated in vivo in an animal model of HF. Hemodynamic pressures and flows were examined.</p><h3 data-test=\\\"abstract-sub-heading\\\">Results</h3><p>Minimally invasive deployment on the beating heart was successful without cardiopulmonary bypass or anticoagulation. During HF, device operation resulted in an immediate 43% increase in cardiac output while maintaining pulsatile hemodynamics. Compared to the pre-HF baseline, the device recovered up to 95% of ventricular stroke volume. At the conclusion of the study, the device was easily removed from the beating heart.</p><h3 data-test=\\\"abstract-sub-heading\\\">Conclusions</h3><p>This preclinical proof-of-concept study demonstrated the feasibility of a DCC device on a pediatric scale that is minimally invasive and non-blood contacting, with promising hemodynamic support and durability for the initial intended duration of use. The ability of DCC to maintain pulsatile MCS without blood contact represents an opportunity to mitigate the mortality and morbidity observed in non-pulsatile, blood-contacting MCS.</p>\",\"PeriodicalId\":54322,\"journal\":{\"name\":\"Cardiovascular Engineering and Technology\",\"volume\":\"236 1\",\"pages\":\"\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2023-12-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cardiovascular Engineering and Technology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1007/s13239-023-00703-0\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Engineering and Technology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s13239-023-00703-0","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Preclinical Proof-of-Concept of a Minimally Invasive Direct Cardiac Compression Device for Pediatric Heart Support
Purpose
For pediatric patients, extracorporeal membrane oxygenation (ECMO) remains the predominant mechanical circulatory support (MCS) modality for heart failure (HF) although survival to discharge rates remain between 50 and 60% for these patients. The device-blood interface and disruption of physiologic hemodynamics are significant contributors to poor outcomes.
Methods
In this study, we evaluate the preclinical feasibility of a minimally invasive, non-blood-contacting pediatric DCC prototype for temporary MCS. Proof-of-concept is demonstrated in vivo in an animal model of HF. Hemodynamic pressures and flows were examined.
Results
Minimally invasive deployment on the beating heart was successful without cardiopulmonary bypass or anticoagulation. During HF, device operation resulted in an immediate 43% increase in cardiac output while maintaining pulsatile hemodynamics. Compared to the pre-HF baseline, the device recovered up to 95% of ventricular stroke volume. At the conclusion of the study, the device was easily removed from the beating heart.
Conclusions
This preclinical proof-of-concept study demonstrated the feasibility of a DCC device on a pediatric scale that is minimally invasive and non-blood contacting, with promising hemodynamic support and durability for the initial intended duration of use. The ability of DCC to maintain pulsatile MCS without blood contact represents an opportunity to mitigate the mortality and morbidity observed in non-pulsatile, blood-contacting MCS.
期刊介绍:
Cardiovascular Engineering and Technology is a journal publishing the spectrum of basic to translational research in all aspects of cardiovascular physiology and medical treatment. It is the forum for academic and industrial investigators to disseminate research that utilizes engineering principles and methods to advance fundamental knowledge and technological solutions related to the cardiovascular system. Manuscripts spanning from subcellular to systems level topics are invited, including but not limited to implantable medical devices, hemodynamics and tissue biomechanics, functional imaging, surgical devices, electrophysiology, tissue engineering and regenerative medicine, diagnostic instruments, transport and delivery of biologics, and sensors. In addition to manuscripts describing the original publication of research, manuscripts reviewing developments in these topics or their state-of-art are also invited.