{"title":"带可拆卸中心延长管的琼斯型电池中的非圆柱表面导致的方法误差","authors":"Oleksii Stennik, Oleksandr Mikhal, D. Meleshchuk","doi":"10.21014/actaimeko.v12i4.1604","DOIUrl":null,"url":null,"abstract":"The article shows an idealized model of a Jones-type cell with a removable central extension tube. Two main factors leading to the cylindricity distortion of the inner surface of the cell are considered. These are radial displacement and tube diameter inequality. Based on the finite element method (FEM), errors in measuring the resistance of a liquid column caused by the non-uniformity of the current density distribution inside the cell were determined. The methodological error with respect to the idealized model was estimated for each factor separately and in combination. The Authors show that at a radial displacement of 0.6 mm, the error can reach 0.1 %. The same error value occurs when the inequality of the diameters is only 20 μm.","PeriodicalId":37987,"journal":{"name":"Acta IMEKO","volume":"61 2","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-12-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Methodological errors due to a non-cylindrical surface in a Jones-type cell with a removable center extension tube\",\"authors\":\"Oleksii Stennik, Oleksandr Mikhal, D. Meleshchuk\",\"doi\":\"10.21014/actaimeko.v12i4.1604\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The article shows an idealized model of a Jones-type cell with a removable central extension tube. Two main factors leading to the cylindricity distortion of the inner surface of the cell are considered. These are radial displacement and tube diameter inequality. Based on the finite element method (FEM), errors in measuring the resistance of a liquid column caused by the non-uniformity of the current density distribution inside the cell were determined. The methodological error with respect to the idealized model was estimated for each factor separately and in combination. The Authors show that at a radial displacement of 0.6 mm, the error can reach 0.1 %. The same error value occurs when the inequality of the diameters is only 20 μm.\",\"PeriodicalId\":37987,\"journal\":{\"name\":\"Acta IMEKO\",\"volume\":\"61 2\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-12-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta IMEKO\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.21014/actaimeko.v12i4.1604\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Engineering\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta IMEKO","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.21014/actaimeko.v12i4.1604","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Engineering","Score":null,"Total":0}
Methodological errors due to a non-cylindrical surface in a Jones-type cell with a removable center extension tube
The article shows an idealized model of a Jones-type cell with a removable central extension tube. Two main factors leading to the cylindricity distortion of the inner surface of the cell are considered. These are radial displacement and tube diameter inequality. Based on the finite element method (FEM), errors in measuring the resistance of a liquid column caused by the non-uniformity of the current density distribution inside the cell were determined. The methodological error with respect to the idealized model was estimated for each factor separately and in combination. The Authors show that at a radial displacement of 0.6 mm, the error can reach 0.1 %. The same error value occurs when the inequality of the diameters is only 20 μm.
期刊介绍:
The main goal of this journal is the enhancement of academic activities of IMEKO and a wider dissemination of scientific output from IMEKO TC events. High-quality papers presented at IMEKO conferences, workshops or congresses are seleted by the event organizers and the authors are invited to publish an enhanced version of their paper in this journal. The journal also publishes scientific articles on measurement and instrumentation not related to an IMEKO event.