两种血液分析仪的结果比较:Dirui BF-7200 和 Sysmex XN-1000

Muhammed Seyithanoğlu, Hilmi İsmet Tuncer, F. Tolun, Metin Kilinç
{"title":"两种血液分析仪的结果比较:Dirui BF-7200 和 Sysmex XN-1000","authors":"Muhammed Seyithanoğlu, Hilmi İsmet Tuncer, F. Tolun, Metin Kilinç","doi":"10.1515/tjb-2023-0184","DOIUrl":null,"url":null,"abstract":"Abstract Objectives Complete blood count (CBC) is performed using automated hematology analyzers. It is important that CBC results are comparable, reproducible, and reliable. In this study, our aim is to compare the results of Sysmex XN-1000 and Dirui BF-7200 hematology analyzers. Methods Patient samples randomly selected from the routine workflow for each instrument were measured 20 consecutive times to assess reproducibility. The mean, standard deviation, and coefficient of variation (CV%) were calculated for each hematological parameter. A comparison of results from the evaluated Dirui BF-7200 system with those from the current hematology analyzer Sysmex XN-1000 system was made for all of the samples included in the study. The compatibility between the parameters was evaluated using Passing–Bablok and Bland–Altman analyses. Results The within-run CV% values were outside the desirable biological variation database (European Federation of Clinical Chemistry and Laboratory Medicine) specification for CV% for eosinophil, basophil, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, mean platelet volume and platelet distribution width expressed as standard deviation in the Sysmex XN-1000 instrument and eosinophil, basophil, hematocrit and indexes of red blood cell and platelet in the Dirui BF-7200 instrument. When the Bland–Altman and Passing–Bablok analysis results were evaluated together, most parameters showed poor agreement; only white blood cells and lymphocytes showed good agreement between the two instruments. Conclusions As there is variability between results from different hematology analyzers, we recommend analyzing patient samples in the same laboratory using the same analyzer to avoid different results that could be misinterpreted.","PeriodicalId":23344,"journal":{"name":"Turkish Journal of Biochemistry","volume":"226 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-11-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison of results of two hematological analyzer systems: Dirui BF-7200 and Sysmex XN-1000\",\"authors\":\"Muhammed Seyithanoğlu, Hilmi İsmet Tuncer, F. Tolun, Metin Kilinç\",\"doi\":\"10.1515/tjb-2023-0184\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Objectives Complete blood count (CBC) is performed using automated hematology analyzers. It is important that CBC results are comparable, reproducible, and reliable. In this study, our aim is to compare the results of Sysmex XN-1000 and Dirui BF-7200 hematology analyzers. Methods Patient samples randomly selected from the routine workflow for each instrument were measured 20 consecutive times to assess reproducibility. The mean, standard deviation, and coefficient of variation (CV%) were calculated for each hematological parameter. A comparison of results from the evaluated Dirui BF-7200 system with those from the current hematology analyzer Sysmex XN-1000 system was made for all of the samples included in the study. The compatibility between the parameters was evaluated using Passing–Bablok and Bland–Altman analyses. Results The within-run CV% values were outside the desirable biological variation database (European Federation of Clinical Chemistry and Laboratory Medicine) specification for CV% for eosinophil, basophil, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, mean platelet volume and platelet distribution width expressed as standard deviation in the Sysmex XN-1000 instrument and eosinophil, basophil, hematocrit and indexes of red blood cell and platelet in the Dirui BF-7200 instrument. When the Bland–Altman and Passing–Bablok analysis results were evaluated together, most parameters showed poor agreement; only white blood cells and lymphocytes showed good agreement between the two instruments. Conclusions As there is variability between results from different hematology analyzers, we recommend analyzing patient samples in the same laboratory using the same analyzer to avoid different results that could be misinterpreted.\",\"PeriodicalId\":23344,\"journal\":{\"name\":\"Turkish Journal of Biochemistry\",\"volume\":\"226 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-11-27\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Turkish Journal of Biochemistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1515/tjb-2023-0184\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Turkish Journal of Biochemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/tjb-2023-0184","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

摘要 目的 使用全自动血液分析仪进行全血细胞计数(CBC)。全血细胞计数结果的可比性、可重复性和可靠性非常重要。本研究旨在比较 Sysmex XN-1000 和 Dirui BF-7200 血液分析仪的结果。方法 从每台仪器的常规工作流程中随机抽取病人样本,连续测量 20 次,以评估重现性。计算每个血液参数的平均值、标准偏差和变异系数(CV%)。针对研究中的所有样本,将所评估的迪瑞 BF-7200 系统的结果与当前血液分析仪 Sysmex XN-1000 系统的结果进行了比较。使用 Passing-Bablok 和 Bland-Altman 分析法评估了参数之间的兼容性。结果 Sysmex XN-1000 仪器中以标准偏差表示的嗜酸性粒细胞、嗜碱性粒细胞、平均血红蛋白、平均血红蛋白浓度、平均血小板体积和血小板分布宽度,以及迪瑞 BF-7200 仪器中以标准偏差表示的嗜酸性粒细胞、嗜碱性粒细胞、血细胞比容、红细胞和血小板指数的运行内 CV% 值均超出了理想生物变异数据库(欧洲临床化学和实验室医学联合会)的规范。当同时评估 Bland-Altman 和 Passing-Bablok 分析结果时,大多数参数的一致性较差;只有白细胞和淋巴细胞在两台仪器之间的一致性较好。结论 由于不同血液分析仪的结果之间存在差异,我们建议在同一实验室使用同一分析仪对患者样本进行分析,以避免出现可能被误读的不同结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Comparison of results of two hematological analyzer systems: Dirui BF-7200 and Sysmex XN-1000
Abstract Objectives Complete blood count (CBC) is performed using automated hematology analyzers. It is important that CBC results are comparable, reproducible, and reliable. In this study, our aim is to compare the results of Sysmex XN-1000 and Dirui BF-7200 hematology analyzers. Methods Patient samples randomly selected from the routine workflow for each instrument were measured 20 consecutive times to assess reproducibility. The mean, standard deviation, and coefficient of variation (CV%) were calculated for each hematological parameter. A comparison of results from the evaluated Dirui BF-7200 system with those from the current hematology analyzer Sysmex XN-1000 system was made for all of the samples included in the study. The compatibility between the parameters was evaluated using Passing–Bablok and Bland–Altman analyses. Results The within-run CV% values were outside the desirable biological variation database (European Federation of Clinical Chemistry and Laboratory Medicine) specification for CV% for eosinophil, basophil, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, mean platelet volume and platelet distribution width expressed as standard deviation in the Sysmex XN-1000 instrument and eosinophil, basophil, hematocrit and indexes of red blood cell and platelet in the Dirui BF-7200 instrument. When the Bland–Altman and Passing–Bablok analysis results were evaluated together, most parameters showed poor agreement; only white blood cells and lymphocytes showed good agreement between the two instruments. Conclusions As there is variability between results from different hematology analyzers, we recommend analyzing patient samples in the same laboratory using the same analyzer to avoid different results that could be misinterpreted.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
CXCL12/CXCR4 as a potential axis in diagnosis and predicting disease severity in COVID-19 patients: a new perspective Human estrogen receptor alpha (ERα) targeted cyclic peptides inhibit cell growth and induce apoptosis in MCF-7 cells In vitro and in silico investigations of the pro-apoptotic activity of Opuntia ficus-indica cladode extracts against K562 cells Evaluation of the implementation of the rational use of laboratory tests in the clinical chemistry laboratory Right vagotomy alters heart rate variability temporarily and increases total choline levels in rats
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1