骨质疏松和骨质疏松症筛查检测:带校准因子和不带校准因子的钙骨定量超声与黄金标准双 X 射线吸收测定法的比较

IF 1.7 4区 医学 Q4 ENDOCRINOLOGY & METABOLISM Journal of Clinical Densitometry Pub Date : 2024-01-27 DOI:10.1016/j.jocd.2024.101470
Nutthapong Moonkum , Thanat Withayanuluck , Arun Somarungsan , Naphondej Sichai , Arisa Wongsiri , Witchayada Chawkhaodin , Podjana Ruengdach , Pasinee Boonsuk , Marut Pukdeeyorng , Gunjanaporn Tochaikul
{"title":"骨质疏松和骨质疏松症筛查检测:带校准因子和不带校准因子的钙骨定量超声与黄金标准双 X 射线吸收测定法的比较","authors":"Nutthapong Moonkum ,&nbsp;Thanat Withayanuluck ,&nbsp;Arun Somarungsan ,&nbsp;Naphondej Sichai ,&nbsp;Arisa Wongsiri ,&nbsp;Witchayada Chawkhaodin ,&nbsp;Podjana Ruengdach ,&nbsp;Pasinee Boonsuk ,&nbsp;Marut Pukdeeyorng ,&nbsp;Gunjanaporn Tochaikul","doi":"10.1016/j.jocd.2024.101470","DOIUrl":null,"url":null,"abstract":"<div><p><em>Background</em>: osteoporosis is a worldwide major health problem that normally diagnosed in advanced stages. So, an early detection at preclinical stage is now an interesting issue. A key factor to early diagnosis the disease is the used of noninvasive bone densitometry. Dual energy x-ray absorptiometry (DXA) is the gold standard techniques for the proposed. However, the high cost, non-widely available and exposed to ionizing radiation are still a drawback of the machine. Therefore, a cheaper, smaller and non-ionizing device such quantitative ultrasound (QUS) is now a favor alternative method, but the possibility of used QUS measurement instead of DXA is still limited due to their uncertainties. So, the aim of our study was to calibrated the QUS with the DXA to allowing the possible to establish a calibration factor (CF) to improve the measured value closer to the standard method.</p><p><em>Methodology</em>: 135 healthy men and women aged 30–88 years were recruited for lumbar spine/femoral neck DXA and calcaneal QUS scanning. The Pearson's correlation between T- and Z-score from the two systems were studied. Moreover, the sensitivity, specificity and percentage of diagnosed accuracy for both with and without CF were calculated.</p><p><em>Results</em>: The significant correlation between the two systems showed a positive trajectory in highly correlation (<em>r</em> = 0.784–0.899). Analyses showed a higher sensitivity, specificity and reduced the misdiagnosed rates when applied the CF in QUS values.</p><p><em>Conclusions</em>: QUS results showed a significantly correlated with DXA results for both lumbar spine and femoral neck sites with some percentage differences. These differences can be reduced by applied an individual specific machine CF to improve a QUS results. As identification of high risk of osteopenia and osteoporosis to reduce the demand of DXA propose, using a QUS alternative method can be a reliable that provide a cheaper and lack of ionizing radiation.</p></div>","PeriodicalId":50240,"journal":{"name":"Journal of Clinical Densitometry","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-01-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Osteopenia and Osteoporosis Screening Detection: Calcaneal Quantitative Ultrasound with and without Calibration Factor Comparison to Gold Standard Dual X-ray Absorptiometry\",\"authors\":\"Nutthapong Moonkum ,&nbsp;Thanat Withayanuluck ,&nbsp;Arun Somarungsan ,&nbsp;Naphondej Sichai ,&nbsp;Arisa Wongsiri ,&nbsp;Witchayada Chawkhaodin ,&nbsp;Podjana Ruengdach ,&nbsp;Pasinee Boonsuk ,&nbsp;Marut Pukdeeyorng ,&nbsp;Gunjanaporn Tochaikul\",\"doi\":\"10.1016/j.jocd.2024.101470\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p><em>Background</em>: osteoporosis is a worldwide major health problem that normally diagnosed in advanced stages. So, an early detection at preclinical stage is now an interesting issue. A key factor to early diagnosis the disease is the used of noninvasive bone densitometry. Dual energy x-ray absorptiometry (DXA) is the gold standard techniques for the proposed. However, the high cost, non-widely available and exposed to ionizing radiation are still a drawback of the machine. Therefore, a cheaper, smaller and non-ionizing device such quantitative ultrasound (QUS) is now a favor alternative method, but the possibility of used QUS measurement instead of DXA is still limited due to their uncertainties. So, the aim of our study was to calibrated the QUS with the DXA to allowing the possible to establish a calibration factor (CF) to improve the measured value closer to the standard method.</p><p><em>Methodology</em>: 135 healthy men and women aged 30–88 years were recruited for lumbar spine/femoral neck DXA and calcaneal QUS scanning. The Pearson's correlation between T- and Z-score from the two systems were studied. Moreover, the sensitivity, specificity and percentage of diagnosed accuracy for both with and without CF were calculated.</p><p><em>Results</em>: The significant correlation between the two systems showed a positive trajectory in highly correlation (<em>r</em> = 0.784–0.899). Analyses showed a higher sensitivity, specificity and reduced the misdiagnosed rates when applied the CF in QUS values.</p><p><em>Conclusions</em>: QUS results showed a significantly correlated with DXA results for both lumbar spine and femoral neck sites with some percentage differences. These differences can be reduced by applied an individual specific machine CF to improve a QUS results. As identification of high risk of osteopenia and osteoporosis to reduce the demand of DXA propose, using a QUS alternative method can be a reliable that provide a cheaper and lack of ionizing radiation.</p></div>\",\"PeriodicalId\":50240,\"journal\":{\"name\":\"Journal of Clinical Densitometry\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-01-27\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical Densitometry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1094695024000052\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Densitometry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1094695024000052","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

摘要

背景骨质疏松症是一个世界性的重大健康问题,通常要到晚期才能确诊。因此,在临床前阶段进行早期检测是一个令人关注的问题。使用无创骨密度测量法是早期诊断该疾病的关键因素。双能 X 射线吸收测量法(DXA)是目前提出的金标准技术。然而,该仪器的缺点是成本高、不能广泛使用,而且会受到电离辐射。因此,定量超声波(QUS)这种更便宜、更小且无电离辐射的设备现在成为了一种受欢迎的替代方法,但由于其不确定性,用 QUS 测量代替 DXA 的可能性仍然有限。因此,我们研究的目的是将 QUS 与 DXA 进行校准,以便建立校准因子(CF),使测量值更接近标准方法。研究了两种系统的 T 值和 Z 值之间的皮尔逊相关性。结果两个系统之间的显著相关性显示出高度相关的正向轨迹(r=0.784-0.899)。分析表明,当在 QUS 值中应用 CF 时,灵敏度和特异性更高,误诊率也更低。这些差异可以通过应用特定的机器 CF 来减少,从而改善 QUS 结果。在识别骨质疏松和骨质疏松症的高风险以减少对 DXA 的需求方面,使用 QUS 替代方法是一种可靠的方法,不仅成本更低,而且没有电离辐射。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Osteopenia and Osteoporosis Screening Detection: Calcaneal Quantitative Ultrasound with and without Calibration Factor Comparison to Gold Standard Dual X-ray Absorptiometry

Background: osteoporosis is a worldwide major health problem that normally diagnosed in advanced stages. So, an early detection at preclinical stage is now an interesting issue. A key factor to early diagnosis the disease is the used of noninvasive bone densitometry. Dual energy x-ray absorptiometry (DXA) is the gold standard techniques for the proposed. However, the high cost, non-widely available and exposed to ionizing radiation are still a drawback of the machine. Therefore, a cheaper, smaller and non-ionizing device such quantitative ultrasound (QUS) is now a favor alternative method, but the possibility of used QUS measurement instead of DXA is still limited due to their uncertainties. So, the aim of our study was to calibrated the QUS with the DXA to allowing the possible to establish a calibration factor (CF) to improve the measured value closer to the standard method.

Methodology: 135 healthy men and women aged 30–88 years were recruited for lumbar spine/femoral neck DXA and calcaneal QUS scanning. The Pearson's correlation between T- and Z-score from the two systems were studied. Moreover, the sensitivity, specificity and percentage of diagnosed accuracy for both with and without CF were calculated.

Results: The significant correlation between the two systems showed a positive trajectory in highly correlation (r = 0.784–0.899). Analyses showed a higher sensitivity, specificity and reduced the misdiagnosed rates when applied the CF in QUS values.

Conclusions: QUS results showed a significantly correlated with DXA results for both lumbar spine and femoral neck sites with some percentage differences. These differences can be reduced by applied an individual specific machine CF to improve a QUS results. As identification of high risk of osteopenia and osteoporosis to reduce the demand of DXA propose, using a QUS alternative method can be a reliable that provide a cheaper and lack of ionizing radiation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Clinical Densitometry
Journal of Clinical Densitometry 医学-内分泌学与代谢
CiteScore
4.90
自引率
8.00%
发文量
92
审稿时长
90 days
期刊介绍: The Journal is committed to serving ISCD''s mission - the education of heterogenous physician specialties and technologists who are involved in the clinical assessment of skeletal health. The focus of JCD is bone mass measurement, including epidemiology of bone mass, how drugs and diseases alter bone mass, new techniques and quality assurance in bone mass imaging technologies, and bone mass health/economics. Combining high quality research and review articles with sound, practice-oriented advice, JCD meets the diverse diagnostic and management needs of radiologists, endocrinologists, nephrologists, rheumatologists, gynecologists, family physicians, internists, and technologists whose patients require diagnostic clinical densitometry for therapeutic management.
期刊最新文献
Technical and biological reliability of pQCT measured bone and muscle tissue quality across the age-span Correlation between body composition and bone mineral density differs by sex and skeletal site in overweight and obese Chinese subjects Lumbar spine densitometry in people with spinal cord injury: Investigation of potential sources of errors Comparison of Mineral Density of Dental Structures of Ancient Human Skeletons from the Roman Period with Modern-Day Human Dental Structures Lumbar Osteoarthritis in Patients with Osteoporosis Evaluated with the Lane and Wilke Scale and its Effect on Dexa Scores
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1