{"title":"心血管图像:一只狗的起搏器导线断裂和过度盘绕","authors":"A.K. McManamey DVM , M.M. Lake DVM , T.C. DeFrancesco DVM","doi":"10.1016/j.jvc.2024.02.004","DOIUrl":null,"url":null,"abstract":"<div><p>A seven-year-old male castrated Labrador Retriever presented emergently due to concern for pacemaker malfunction five years after successful transvenous pacemaker implantation to treat partial atrial standstill. On presentation, the dog's pulse rate was 30–50 beats per minute. An electrocardiogram showed no spontaneous atrial activity or paced ventricular activity. Pacemaker interrogation revealed an increased impedance of 7557 ohms, indicating a lead malfunction. Thoracic radiographs confirmed the lead was fractured and had excessive coiling. The transvenous pacing system was turned off, left in place, and an epicardial pacing system was implanted the following day. The dog was discharged with no perioperative complications. The dog eventually required escalated medical therapy for progressive cardiac disease and was euthanized two years after implantation of the replacement pacemaker. This manuscript illustrates a complete lead fracture and excessive lead coiling, which has not previously been detailed in veterinary medicine.</p></div>","PeriodicalId":48788,"journal":{"name":"Journal of Veterinary Cardiology","volume":"52 ","pages":"Pages 68-71"},"PeriodicalIF":1.5000,"publicationDate":"2024-02-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Cardiovascular images: pacemaker-lead fracture and excessive coiling in a dog\",\"authors\":\"A.K. McManamey DVM , M.M. Lake DVM , T.C. DeFrancesco DVM\",\"doi\":\"10.1016/j.jvc.2024.02.004\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>A seven-year-old male castrated Labrador Retriever presented emergently due to concern for pacemaker malfunction five years after successful transvenous pacemaker implantation to treat partial atrial standstill. On presentation, the dog's pulse rate was 30–50 beats per minute. An electrocardiogram showed no spontaneous atrial activity or paced ventricular activity. Pacemaker interrogation revealed an increased impedance of 7557 ohms, indicating a lead malfunction. Thoracic radiographs confirmed the lead was fractured and had excessive coiling. The transvenous pacing system was turned off, left in place, and an epicardial pacing system was implanted the following day. The dog was discharged with no perioperative complications. The dog eventually required escalated medical therapy for progressive cardiac disease and was euthanized two years after implantation of the replacement pacemaker. This manuscript illustrates a complete lead fracture and excessive lead coiling, which has not previously been detailed in veterinary medicine.</p></div>\",\"PeriodicalId\":48788,\"journal\":{\"name\":\"Journal of Veterinary Cardiology\",\"volume\":\"52 \",\"pages\":\"Pages 68-71\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2024-02-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Veterinary Cardiology\",\"FirstCategoryId\":\"97\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1760273424000079\",\"RegionNum\":2,\"RegionCategory\":\"农林科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"VETERINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Veterinary Cardiology","FirstCategoryId":"97","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1760273424000079","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
Cardiovascular images: pacemaker-lead fracture and excessive coiling in a dog
A seven-year-old male castrated Labrador Retriever presented emergently due to concern for pacemaker malfunction five years after successful transvenous pacemaker implantation to treat partial atrial standstill. On presentation, the dog's pulse rate was 30–50 beats per minute. An electrocardiogram showed no spontaneous atrial activity or paced ventricular activity. Pacemaker interrogation revealed an increased impedance of 7557 ohms, indicating a lead malfunction. Thoracic radiographs confirmed the lead was fractured and had excessive coiling. The transvenous pacing system was turned off, left in place, and an epicardial pacing system was implanted the following day. The dog was discharged with no perioperative complications. The dog eventually required escalated medical therapy for progressive cardiac disease and was euthanized two years after implantation of the replacement pacemaker. This manuscript illustrates a complete lead fracture and excessive lead coiling, which has not previously been detailed in veterinary medicine.
期刊介绍:
The mission of the Journal of Veterinary Cardiology is to publish peer-reviewed reports of the highest quality that promote greater understanding of cardiovascular disease, and enhance the health and well being of animals and humans. The Journal of Veterinary Cardiology publishes original contributions involving research and clinical practice that include prospective and retrospective studies, clinical trials, epidemiology, observational studies, and advances in applied and basic research.
The Journal invites submission of original manuscripts. Specific content areas of interest include heart failure, arrhythmias, congenital heart disease, cardiovascular medicine, surgery, hypertension, health outcomes research, diagnostic imaging, interventional techniques, genetics, molecular cardiology, and cardiovascular pathology, pharmacology, and toxicology.