使用 Sysmex XN-10 检测未成熟血小板部分的可重复性和稳定性。

IF 1.3 4区 医学 Q4 MEDICINE, RESEARCH & EXPERIMENTAL Scandinavian Journal of Clinical & Laboratory Investigation Pub Date : 2024-04-01 Epub Date: 2024-02-29 DOI:10.1080/00365513.2024.2321590
Usman Ali, Gavin Knight, Mridula Chopra, Dimitris A Tsitsikas
{"title":"使用 Sysmex XN-10 检测未成熟血小板部分的可重复性和稳定性。","authors":"Usman Ali, Gavin Knight, Mridula Chopra, Dimitris A Tsitsikas","doi":"10.1080/00365513.2024.2321590","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The Immature Platelet Fraction (IPF) is an indicator of thrombopoiesis which is a useful parameter in thrombocytopenia. It demonstrates compensatory mechanisms in production of platelets, but currently not implemented in routine clinical practice. The aim of this study was to establish the reproducibility and stability of IPF, for both percentage (%-IPF) and absolute (A-IPF) measurements.<b>Material/methods:</b> A total of 71 samples, of which 45 for reproducibility and 26 for stability analysis, were assayed for full blood count using the Sysmex XN-10 analyser at room temperature (RT:19-25 °C). For reproducibility analysis, IPF measurements were analysed 11 times by different appraisers using the same sample, while for stability analysis, IPF was measured over fourteen hourly-intervals up to 24 h (<i>n</i> = 21) and then separately extended beyond the point of stability to 72 h (<i>n</i> = 5).</p><p><strong>Results: </strong>Reproducibility analysis of %-IPF and A-IPF (<i>n</i> = 45) showed very reliable results, with the range of mean CV% values between 1.25-8.90% and 1.70-9.96%, respectively. On the other hand, overall, stability analysis of %-IPF and A-IPF (<i>n</i> = 21) at RT over 24 h showed reliable results, with pooled mean CV% values of 1.32% and 1.43%, respectively, with no significant difference between %-IPF and A-IPF (<i>p</i> = 0.767 and <i>p</i> = 0.821). All %-IPF and A-IPF values had exceeded the set acceptance criterion of stability (CV% ≥ 10.0%) before 72 h.</p><p><strong>Conclusions: </strong>Overall, %-IPF and A-IPF reproducibility and storage at RT for 24 h predominantly demonstrates the suitability of their usage for testing on the Sysmex XN-series analysers.</p>","PeriodicalId":21474,"journal":{"name":"Scandinavian Journal of Clinical & Laboratory Investigation","volume":" ","pages":"91-96"},"PeriodicalIF":1.3000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Reproducibility and stability of the immature platelet fraction using Sysmex XN-10.\",\"authors\":\"Usman Ali, Gavin Knight, Mridula Chopra, Dimitris A Tsitsikas\",\"doi\":\"10.1080/00365513.2024.2321590\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The Immature Platelet Fraction (IPF) is an indicator of thrombopoiesis which is a useful parameter in thrombocytopenia. It demonstrates compensatory mechanisms in production of platelets, but currently not implemented in routine clinical practice. The aim of this study was to establish the reproducibility and stability of IPF, for both percentage (%-IPF) and absolute (A-IPF) measurements.<b>Material/methods:</b> A total of 71 samples, of which 45 for reproducibility and 26 for stability analysis, were assayed for full blood count using the Sysmex XN-10 analyser at room temperature (RT:19-25 °C). For reproducibility analysis, IPF measurements were analysed 11 times by different appraisers using the same sample, while for stability analysis, IPF was measured over fourteen hourly-intervals up to 24 h (<i>n</i> = 21) and then separately extended beyond the point of stability to 72 h (<i>n</i> = 5).</p><p><strong>Results: </strong>Reproducibility analysis of %-IPF and A-IPF (<i>n</i> = 45) showed very reliable results, with the range of mean CV% values between 1.25-8.90% and 1.70-9.96%, respectively. On the other hand, overall, stability analysis of %-IPF and A-IPF (<i>n</i> = 21) at RT over 24 h showed reliable results, with pooled mean CV% values of 1.32% and 1.43%, respectively, with no significant difference between %-IPF and A-IPF (<i>p</i> = 0.767 and <i>p</i> = 0.821). All %-IPF and A-IPF values had exceeded the set acceptance criterion of stability (CV% ≥ 10.0%) before 72 h.</p><p><strong>Conclusions: </strong>Overall, %-IPF and A-IPF reproducibility and storage at RT for 24 h predominantly demonstrates the suitability of their usage for testing on the Sysmex XN-series analysers.</p>\",\"PeriodicalId\":21474,\"journal\":{\"name\":\"Scandinavian Journal of Clinical & Laboratory Investigation\",\"volume\":\" \",\"pages\":\"91-96\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Scandinavian Journal of Clinical & Laboratory Investigation\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/00365513.2024.2321590\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/29 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICINE, RESEARCH & EXPERIMENTAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian Journal of Clinical & Laboratory Investigation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00365513.2024.2321590","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/29 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0

摘要

背景:未成熟血小板比例(IPF)是血小板生成的一个指标,是血小板减少症的一个有用参数。它显示了血小板生成的代偿机制,但目前尚未在常规临床实践中应用。本研究旨在确定 IPF 百分比(%-IPF)和绝对值(A-IPF)测量的可重复性和稳定性:使用 Sysmex XN-10 分析仪在室温(RT:19-25 °C)下对 71 份样本进行了全血细胞计数分析,其中 45 份用于重现性分析,26 份用于稳定性分析。在重现性分析中,由不同的鉴定人员使用同一样本进行了 11 次 IPF 测量分析;而在稳定性分析中,IPF 的测量时间间隔为每小时 14 次,直至 24 小时(n = 21),然后在稳定点之后分别延长至 72 小时(n = 5):结果:%-IPF 和 A-IPF 的重现性分析(n = 45)显示结果非常可靠,平均 CV% 值范围分别为 1.25-8.90% 和 1.70-9.96%。另一方面,总体而言,RT 24 小时内 %-IPF 和 A-IPF(n = 21)的稳定性分析结果可靠,汇总平均 CV% 值分别为 1.32% 和 1.43%,%-IPF 和 A-IPF 之间无显著差异(p = 0.767 和 p = 0.821)。所有 %-IPF 和 A-IPF 值在 72 小时前都超过了设定的稳定性接受标准(CV% ≥ 10.0%):总体而言,%-IPF 和 A-IPF 的重现性和在 RT 条件下保存 24 小时的稳定性表明,它们适合在 Sysmex XN 系列分析仪上进行测试。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Reproducibility and stability of the immature platelet fraction using Sysmex XN-10.

Background: The Immature Platelet Fraction (IPF) is an indicator of thrombopoiesis which is a useful parameter in thrombocytopenia. It demonstrates compensatory mechanisms in production of platelets, but currently not implemented in routine clinical practice. The aim of this study was to establish the reproducibility and stability of IPF, for both percentage (%-IPF) and absolute (A-IPF) measurements.Material/methods: A total of 71 samples, of which 45 for reproducibility and 26 for stability analysis, were assayed for full blood count using the Sysmex XN-10 analyser at room temperature (RT:19-25 °C). For reproducibility analysis, IPF measurements were analysed 11 times by different appraisers using the same sample, while for stability analysis, IPF was measured over fourteen hourly-intervals up to 24 h (n = 21) and then separately extended beyond the point of stability to 72 h (n = 5).

Results: Reproducibility analysis of %-IPF and A-IPF (n = 45) showed very reliable results, with the range of mean CV% values between 1.25-8.90% and 1.70-9.96%, respectively. On the other hand, overall, stability analysis of %-IPF and A-IPF (n = 21) at RT over 24 h showed reliable results, with pooled mean CV% values of 1.32% and 1.43%, respectively, with no significant difference between %-IPF and A-IPF (p = 0.767 and p = 0.821). All %-IPF and A-IPF values had exceeded the set acceptance criterion of stability (CV% ≥ 10.0%) before 72 h.

Conclusions: Overall, %-IPF and A-IPF reproducibility and storage at RT for 24 h predominantly demonstrates the suitability of their usage for testing on the Sysmex XN-series analysers.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.50
自引率
4.80%
发文量
85
审稿时长
4-8 weeks
期刊介绍: The Scandinavian Journal of Clinical and Laboratory Investigation is an international scientific journal covering clinically oriented biochemical and physiological research. Since the launch of the journal in 1949, it has been a forum for international laboratory medicine, closely related to, and edited by, The Scandinavian Society for Clinical Chemistry. The journal contains peer-reviewed articles, editorials, invited reviews, and short technical notes, as well as several supplements each year. Supplements consist of monographs, and symposium and congress reports covering subjects within clinical chemistry and clinical physiology.
期刊最新文献
Analytical interference on measurement of immunoglobulins in monoclonal gammopathy. FlowDiff: a simple, flow cytometry-based approach for performing a leukocyte differential count. Reference intervals for 23 common biochemical parameters during pregnancy and the first six postpartum months. Prevalence of monoclonal proteins in patients with isolated hypogammaglobulinemia on serum protein electrophoresis. Challenges of preanalytical variables in erythrocyte sedimentation rate: a CUBE 30 touch evaluation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1