{"title":"计算机模拟 TOUCH 假体杯错位及解决方案。","authors":"Kevin Knappe , Mareike Schonhoff , Sebastian Jaeger , Berthold Bickert , Leila Harhaus , Benjamin Panzram","doi":"10.1016/j.hansur.2024.101712","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>Total joint replacement has become significantly more common as a treatment for advanced trapeziometacarpal joint osteoarthritis in recent years. The latest generation of prostheses with dual-mobility designs leads to very good functional results and low rates of loosening and dislocation in the short and medium term. Biomechanical studies showed that central placement and parallel alignment of the cup with respect to the proximal articular surface of the trapezium are crucial for both cup stability and prevention of dislocation. Despite correct positioning of the guidewire, incorrect placement or tilting of the inserted cup may occur, requiring immediate intraoperative revision.</p></div><div><h3>Methods</h3><p>The existing spherical and conical cup models in sizes 9 mm and 10 mm were transferred to a computer-aided design dataset. Depending on the intraoperative complication (tilting or incorrect placement), the revision options resulting from the various combinations of cup type and size were simulated and analyzed according to the resulting defect area and bony contact area.</p></div><div><h3>Results</h3><p>In well centered cups, a size 9 conical cup could be replaced by a size 9 spherical cup and still be fixed by press-fit. Conversely, a size 9 spherical cup could not be replaced by a size 9 conical cup, but only by a size 10 cup, of whatever shape. When a size 9 conical cup was tilted up to 20°, the best revision option was to resect the sclerotic margin and insert a size 10 conical cup deeper into the cancellous bone, to achieve the largest contact area with the surrounding bone. When a size 9 cup of whatever shape was poorly centered (misplaced with respect to the dorsopalmar or radioulnar line of the trapezium), placement should be corrected using a size 10 cup, combined with autologous bone grafting of the defect. Again, the size 10 conical cup showed the largest bony contact area.</p></div><div><h3>Conclusion</h3><p>Our computer-based measurements suggested options for intraoperative cup revision depending on cup shape and size and on type of misalignment with resulting bone defects. These suggestions, however, need to be confirmed in anatomic specimens before introducing them into clinical practice.</p></div>","PeriodicalId":54301,"journal":{"name":"Hand Surgery & Rehabilitation","volume":"43 3","pages":"Article 101712"},"PeriodicalIF":0.9000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2468122924001014/pdfft?md5=2b1095278d8d987d8b7dac4303c9c5d3&pid=1-s2.0-S2468122924001014-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Computer-simulated TOUCH prosthesis cup malposition and solutions\",\"authors\":\"Kevin Knappe , Mareike Schonhoff , Sebastian Jaeger , Berthold Bickert , Leila Harhaus , Benjamin Panzram\",\"doi\":\"10.1016/j.hansur.2024.101712\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>Total joint replacement has become significantly more common as a treatment for advanced trapeziometacarpal joint osteoarthritis in recent years. The latest generation of prostheses with dual-mobility designs leads to very good functional results and low rates of loosening and dislocation in the short and medium term. Biomechanical studies showed that central placement and parallel alignment of the cup with respect to the proximal articular surface of the trapezium are crucial for both cup stability and prevention of dislocation. Despite correct positioning of the guidewire, incorrect placement or tilting of the inserted cup may occur, requiring immediate intraoperative revision.</p></div><div><h3>Methods</h3><p>The existing spherical and conical cup models in sizes 9 mm and 10 mm were transferred to a computer-aided design dataset. Depending on the intraoperative complication (tilting or incorrect placement), the revision options resulting from the various combinations of cup type and size were simulated and analyzed according to the resulting defect area and bony contact area.</p></div><div><h3>Results</h3><p>In well centered cups, a size 9 conical cup could be replaced by a size 9 spherical cup and still be fixed by press-fit. Conversely, a size 9 spherical cup could not be replaced by a size 9 conical cup, but only by a size 10 cup, of whatever shape. When a size 9 conical cup was tilted up to 20°, the best revision option was to resect the sclerotic margin and insert a size 10 conical cup deeper into the cancellous bone, to achieve the largest contact area with the surrounding bone. When a size 9 cup of whatever shape was poorly centered (misplaced with respect to the dorsopalmar or radioulnar line of the trapezium), placement should be corrected using a size 10 cup, combined with autologous bone grafting of the defect. Again, the size 10 conical cup showed the largest bony contact area.</p></div><div><h3>Conclusion</h3><p>Our computer-based measurements suggested options for intraoperative cup revision depending on cup shape and size and on type of misalignment with resulting bone defects. These suggestions, however, need to be confirmed in anatomic specimens before introducing them into clinical practice.</p></div>\",\"PeriodicalId\":54301,\"journal\":{\"name\":\"Hand Surgery & Rehabilitation\",\"volume\":\"43 3\",\"pages\":\"Article 101712\"},\"PeriodicalIF\":0.9000,\"publicationDate\":\"2024-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2468122924001014/pdfft?md5=2b1095278d8d987d8b7dac4303c9c5d3&pid=1-s2.0-S2468122924001014-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Hand Surgery & Rehabilitation\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2468122924001014\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hand Surgery & Rehabilitation","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2468122924001014","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
Computer-simulated TOUCH prosthesis cup malposition and solutions
Introduction
Total joint replacement has become significantly more common as a treatment for advanced trapeziometacarpal joint osteoarthritis in recent years. The latest generation of prostheses with dual-mobility designs leads to very good functional results and low rates of loosening and dislocation in the short and medium term. Biomechanical studies showed that central placement and parallel alignment of the cup with respect to the proximal articular surface of the trapezium are crucial for both cup stability and prevention of dislocation. Despite correct positioning of the guidewire, incorrect placement or tilting of the inserted cup may occur, requiring immediate intraoperative revision.
Methods
The existing spherical and conical cup models in sizes 9 mm and 10 mm were transferred to a computer-aided design dataset. Depending on the intraoperative complication (tilting or incorrect placement), the revision options resulting from the various combinations of cup type and size were simulated and analyzed according to the resulting defect area and bony contact area.
Results
In well centered cups, a size 9 conical cup could be replaced by a size 9 spherical cup and still be fixed by press-fit. Conversely, a size 9 spherical cup could not be replaced by a size 9 conical cup, but only by a size 10 cup, of whatever shape. When a size 9 conical cup was tilted up to 20°, the best revision option was to resect the sclerotic margin and insert a size 10 conical cup deeper into the cancellous bone, to achieve the largest contact area with the surrounding bone. When a size 9 cup of whatever shape was poorly centered (misplaced with respect to the dorsopalmar or radioulnar line of the trapezium), placement should be corrected using a size 10 cup, combined with autologous bone grafting of the defect. Again, the size 10 conical cup showed the largest bony contact area.
Conclusion
Our computer-based measurements suggested options for intraoperative cup revision depending on cup shape and size and on type of misalignment with resulting bone defects. These suggestions, however, need to be confirmed in anatomic specimens before introducing them into clinical practice.
期刊介绍:
As the official publication of the French, Belgian and Swiss Societies for Surgery of the Hand, as well as of the French Society of Rehabilitation of the Hand & Upper Limb, ''Hand Surgery and Rehabilitation'' - formerly named "Chirurgie de la Main" - publishes original articles, literature reviews, technical notes, and clinical cases. It is indexed in the main international databases (including Medline). Initially a platform for French-speaking hand surgeons, the journal will now publish its articles in English to disseminate its author''s scientific findings more widely. The journal also includes a biannual supplement in French, the monograph of the French Society for Surgery of the Hand, where comprehensive reviews in the fields of hand, peripheral nerve and upper limb surgery are presented.
Organe officiel de la Société française de chirurgie de la main, de la Société française de Rééducation de la main (SFRM-GEMMSOR), de la Société suisse de chirurgie de la main et du Belgian Hand Group, indexée dans les grandes bases de données internationales (Medline, Embase, Pascal, Scopus), Hand Surgery and Rehabilitation - anciennement titrée Chirurgie de la main - publie des articles originaux, des revues de la littérature, des notes techniques, des cas clinique. Initialement plateforme d''expression francophone de la spécialité, la revue s''oriente désormais vers l''anglais pour devenir une référence scientifique et de formation de la spécialité en France et en Europe. Avec 6 publications en anglais par an, la revue comprend également un supplément biannuel, la monographie du GEM, où sont présentées en français, des mises au point complètes dans les domaines de la chirurgie de la main, des nerfs périphériques et du membre supérieur.