胸部和腹部计算机断层扫描时手臂位置对图像质量和辐射剂量的影响

IF 2.5 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Radiography Pub Date : 2024-05-08 DOI:10.1016/j.radi.2024.04.015
E.Ø. Buchhave , C.D. Rasmussen , R. Lentz , E.H. Frederiksen , J.O. Kipp , M.Ø. Leftes
{"title":"胸部和腹部计算机断层扫描时手臂位置对图像质量和辐射剂量的影响","authors":"E.Ø. Buchhave ,&nbsp;C.D. Rasmussen ,&nbsp;R. Lentz ,&nbsp;E.H. Frederiksen ,&nbsp;J.O. Kipp ,&nbsp;M.Ø. Leftes","doi":"10.1016/j.radi.2024.04.015","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose <em>(E</em>), when positioning the arms below shoulder level in CT-TAP.</p></div><div><h3>Methods</h3><p>An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate <em>E</em>. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.</p></div><div><h3>Results</h3><p>Significant differences in <em>E</em> between all arm positions were seen (p = 0.005). The lowest <em>E</em> was obtained in position C, reducing <em>E</em> by 8.42%. Position A provided the best image quality but the highest <em>E.</em></p></div><div><h3>Conclusion</h3><p>This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest <em>E</em>. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.</p></div><div><h3>Implication for practice</h3><p>This study illustrates tendencies for objective image quality and <em>E</em> when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.</p></div>","PeriodicalId":47416,"journal":{"name":"Radiography","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-05-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans\",\"authors\":\"E.Ø. Buchhave ,&nbsp;C.D. Rasmussen ,&nbsp;R. Lentz ,&nbsp;E.H. Frederiksen ,&nbsp;J.O. Kipp ,&nbsp;M.Ø. Leftes\",\"doi\":\"10.1016/j.radi.2024.04.015\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose <em>(E</em>), when positioning the arms below shoulder level in CT-TAP.</p></div><div><h3>Methods</h3><p>An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate <em>E</em>. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.</p></div><div><h3>Results</h3><p>Significant differences in <em>E</em> between all arm positions were seen (p = 0.005). The lowest <em>E</em> was obtained in position C, reducing <em>E</em> by 8.42%. Position A provided the best image quality but the highest <em>E.</em></p></div><div><h3>Conclusion</h3><p>This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest <em>E</em>. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.</p></div><div><h3>Implication for practice</h3><p>This study illustrates tendencies for objective image quality and <em>E</em> when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.</p></div>\",\"PeriodicalId\":47416,\"journal\":{\"name\":\"Radiography\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.5000,\"publicationDate\":\"2024-05-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiography\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1078817424000920\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiography","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078817424000920","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

导言在对胸部-腹部-盆腔(TAP)进行计算机断层扫描(CT)时,应将患者的手臂置于头部上方,以获得最佳图像质量,并使患者受到尽可能小的辐射剂量。肩部有问题的患者可能无法做到这一点,因此需要以其他方式摆放手臂。本研究旨在调查在 CT-TAP 中将手臂置于肩部水平以下时,客观图像质量和估计有效剂量(E)的差异。测试了四种手臂位置:测试了四种手臂位置:沿躯干(A)、骨盆上(B)、骨盆上的枕头上(C),以及一只手臂放在骨盆上的枕头上,另一只手臂放在骨盆上(D)。使用的是配有 CareDose 4D 的西门子 SOMATOM Definition Flash CT 扫描仪。通过测量肝脏和膀胱相关区域内的噪声,对图像质量进行客观评估。结果所有手臂位置的 E 值均有显著差异(P = 0.005)。C位的E值最低,降低了8.42%。结论这项研究表明,CT-TAP 没有明确的最佳手臂位置。这些结果可能会影响 CT 诊断的规划,因为双臂的位置可能会影响最佳图像质量和辐射剂量。对实践的意义这项研究说明了当双臂位置低于肩部水平时,客观图像质量和 E 值的变化趋势。需要进一步研究以评估其临床相关性和诊断潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans

Introduction

During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose (E), when positioning the arms below shoulder level in CT-TAP.

Methods

An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate E. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.

Results

Significant differences in E between all arm positions were seen (p = 0.005). The lowest E was obtained in position C, reducing E by 8.42%. Position A provided the best image quality but the highest E.

Conclusion

This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest E. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.

Implication for practice

This study illustrates tendencies for objective image quality and E when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Radiography
Radiography RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
4.70
自引率
34.60%
发文量
169
审稿时长
63 days
期刊介绍: Radiography is an International, English language, peer-reviewed journal of diagnostic imaging and radiation therapy. Radiography is the official professional journal of the College of Radiographers and is published quarterly. Radiography aims to publish the highest quality material, both clinical and scientific, on all aspects of diagnostic imaging and radiation therapy and oncology.
期刊最新文献
An enquiry into the usefulness of an information portal for Deaf and hard of hearing people prior to x-ray examinations and improvement ideas Morphometry-based radiomics for predicting prognosis in soft tissue sarcomas of extremities following radiotherapy Virtual reality simulation for mastery learning of wrist radiograph technique An initial exploration of factors that may impact radiographer performance in reporting mammograms Diagnostic radiography workforce expectations of learners against the 2023 HCPC standards of proficiency: Results of a UK Delphi study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1