使用立体放大镜获得的图像测量树脂复合材料颜色的可靠性。

IF 1.5 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE Brazilian oral research Pub Date : 2024-05-13 eCollection Date: 2024-01-01 DOI:10.1590/1807-3107bor-2024.vol38.0032
Alexsandra Santos Albuquerque, Rafaella Bragança, Oscar Emilio Pecho, André Luis Faria-E-Silva
{"title":"使用立体放大镜获得的图像测量树脂复合材料颜色的可靠性。","authors":"Alexsandra Santos Albuquerque, Rafaella Bragança, Oscar Emilio Pecho, André Luis Faria-E-Silva","doi":"10.1590/1807-3107bor-2024.vol38.0032","DOIUrl":null,"url":null,"abstract":"<p><p>This study assessed the reliability of a color measurement method using images obtained from a charge-coupled device (CCD) camera and a stereoscopic loupe. Disc-shaped specimens were created using the composite Filtek Z350 XT (shades DA1, DA2, DA3, and DA4) (n = 3). CIELAB color coordinates of the specimens were measured using the spectrophotometer SP60 over white and black backgrounds. Images of the same specimens were taken using a CCD camera attached to a stereoscopic loupe. The color of the image was measured (red-green-blue [RGB]) using an image processing software and converted to CIELAB coordinates. For each color coordinate, data from images were adjusted using linear regressions predicting those values from SP60. The whiteness index for dentistry (WID) and translucency parameter (TP00) of the specimens as well as the color differences (ΔE00) among pairwise shades were calculated. Data were analyzed via repeated-measures analysis of variance and Tukey's post hoc test (α = 0.05). Images obtained using the loupe tended to be darker and redder than the actual color. Data adjustment resulted in similar WID, ΔE00, and TP00 values to those observed for the spectrophotometer. Differences were observed only for the WID of shade DA3 and ΔE00 for comparing DA1 and DA3 over the black background. However, these differences were not clinically relevant. The use of adjusted data from images taken using a stereoscopic loupe is considered a feasible method for color measurement.</p>","PeriodicalId":9240,"journal":{"name":"Brazilian oral research","volume":null,"pages":null},"PeriodicalIF":1.5000,"publicationDate":"2024-05-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Reliability of the color measurement of resin composites using images obtained using a stereoscopic loupe.\",\"authors\":\"Alexsandra Santos Albuquerque, Rafaella Bragança, Oscar Emilio Pecho, André Luis Faria-E-Silva\",\"doi\":\"10.1590/1807-3107bor-2024.vol38.0032\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This study assessed the reliability of a color measurement method using images obtained from a charge-coupled device (CCD) camera and a stereoscopic loupe. Disc-shaped specimens were created using the composite Filtek Z350 XT (shades DA1, DA2, DA3, and DA4) (n = 3). CIELAB color coordinates of the specimens were measured using the spectrophotometer SP60 over white and black backgrounds. Images of the same specimens were taken using a CCD camera attached to a stereoscopic loupe. The color of the image was measured (red-green-blue [RGB]) using an image processing software and converted to CIELAB coordinates. For each color coordinate, data from images were adjusted using linear regressions predicting those values from SP60. The whiteness index for dentistry (WID) and translucency parameter (TP00) of the specimens as well as the color differences (ΔE00) among pairwise shades were calculated. Data were analyzed via repeated-measures analysis of variance and Tukey's post hoc test (α = 0.05). Images obtained using the loupe tended to be darker and redder than the actual color. Data adjustment resulted in similar WID, ΔE00, and TP00 values to those observed for the spectrophotometer. Differences were observed only for the WID of shade DA3 and ΔE00 for comparing DA1 and DA3 over the black background. However, these differences were not clinically relevant. The use of adjusted data from images taken using a stereoscopic loupe is considered a feasible method for color measurement.</p>\",\"PeriodicalId\":9240,\"journal\":{\"name\":\"Brazilian oral research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2024-05-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Brazilian oral research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1590/1807-3107bor-2024.vol38.0032\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brazilian oral research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1590/1807-3107bor-2024.vol38.0032","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

本研究使用电荷耦合器件 (CCD) 相机和立体放大镜获得的图像,评估了颜色测量方法的可靠性。使用复合 Filtek Z350 XT(色调 DA1、DA2、DA3 和 DA4)制作了圆盘状试样(n = 3)。使用分光光度计 SP60 在白色和黑色背景上测量试样的 CIELAB 色坐标。使用安装在立体放大镜上的 CCD 相机拍摄相同标本的图像。使用图像处理软件测量图像的颜色(红-绿-蓝 [RGB])并转换为 CIELAB 坐标。对于每个颜色坐标,使用线性回归预测 SP60 的值来调整图像数据。计算样本的牙科白度指数(WID)和半透明参数(TP00),以及成对色调之间的色差(ΔE00)。数据分析采用重复测量方差分析和 Tukey 后检验(α = 0.05)。使用放大镜获得的图像往往比实际颜色更深更红。数据调整后的 WID、ΔE00 和 TP00 值与分光光度计观察到的值相似。只有在黑色背景上比较 DA1 和 DA3 的色调 DA3 的 WID 和 ΔE00 时才会观察到差异。不过,这些差异与临床无关。使用立体放大镜拍摄的图像调整数据被认为是一种可行的颜色测量方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Reliability of the color measurement of resin composites using images obtained using a stereoscopic loupe.

This study assessed the reliability of a color measurement method using images obtained from a charge-coupled device (CCD) camera and a stereoscopic loupe. Disc-shaped specimens were created using the composite Filtek Z350 XT (shades DA1, DA2, DA3, and DA4) (n = 3). CIELAB color coordinates of the specimens were measured using the spectrophotometer SP60 over white and black backgrounds. Images of the same specimens were taken using a CCD camera attached to a stereoscopic loupe. The color of the image was measured (red-green-blue [RGB]) using an image processing software and converted to CIELAB coordinates. For each color coordinate, data from images were adjusted using linear regressions predicting those values from SP60. The whiteness index for dentistry (WID) and translucency parameter (TP00) of the specimens as well as the color differences (ΔE00) among pairwise shades were calculated. Data were analyzed via repeated-measures analysis of variance and Tukey's post hoc test (α = 0.05). Images obtained using the loupe tended to be darker and redder than the actual color. Data adjustment resulted in similar WID, ΔE00, and TP00 values to those observed for the spectrophotometer. Differences were observed only for the WID of shade DA3 and ΔE00 for comparing DA1 and DA3 over the black background. However, these differences were not clinically relevant. The use of adjusted data from images taken using a stereoscopic loupe is considered a feasible method for color measurement.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.70
自引率
4.00%
发文量
107
审稿时长
12 weeks
期刊最新文献
Do SOD2 and SOD3 gene polymorphisms impact the oral health-related quality of life in Para athletes? Evaluation of polymerization shrinkage stress and cuspal strain in natural and typodont teeth. Exploring the factors driving higher interactions in false amber necklace posts on Facebook. Impact of COVID-19 on pediatric dental care in two epicenters: Italy and Brazil. Impact of streptozotocin-induced diabetes on experimental masseter pain in rats.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1