使用美国医学物理学家协会计算机断层扫描性能模型测量切片间距的实用方法。

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of Medical Physics Pub Date : 2024-01-01 Epub Date: 2024-03-30 DOI:10.4103/jmp.jmp_155_23
Choirul Anam, Ariij Naufal, Yanurita Dwihapsari, Toshioh Fujibuchi, Geoff Dougherty
{"title":"使用美国医学物理学家协会计算机断层扫描性能模型测量切片间距的实用方法。","authors":"Choirul Anam, Ariij Naufal, Yanurita Dwihapsari, Toshioh Fujibuchi, Geoff Dougherty","doi":"10.4103/jmp.jmp_155_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The slice spacing has a crucial role in the accuracy of computed tomography (CT) images in sagittal and coronal planes. However, there is no practical method for measuring the accuracy of the slice spacing.</p><p><strong>Purpose: </strong>This study proposes a novel method to automatically measure the slice spacing using the American Association of Physicists in Medicine (AAPM) CT performance phantom.</p><p><strong>Methods: </strong>The AAPM CT performance phantom module 610-04 was used to measure slice spacing. The process of slice spacing measurement involves a pair of axial images of the module containing ramp aluminum objects located at adjacent slice positions. The middle aluminum plate of each image was automatically segmented. Next, the two segmented images were combined to produce one image with two stair objects. The centroid coordinates of two stair objects were automatically determined. Subsequently, the distance between these two centroids was measured to directly indicate the slice spacing. For comparison, the slice spacing was calculated by accessing the slice position attributes from the DICOM header of both images. The proposed method was tested on phantom images with variations in slice spacing and field of view (FOV).</p><p><strong>Results: </strong>The results showed that the automatic measurement of slice spacing was quite accurate for all variations of slice spacing and FOV, with average differences of 9.0% and 9.3%, respectively.</p><p><strong>Conclusion: </strong>A new automated method for measuring the slice spacing using the AAPM CT phantom was successfully demonstrated and tested for variations of slice spacing and FOV. Slice spacing measurement may be considered an additional parameter to be checked in addition to other established parameters.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 1","pages":"103-109"},"PeriodicalIF":0.7000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11141755/pdf/","citationCount":"0","resultStr":"{\"title\":\"A Practical Method for Slice Spacing Measurement Using the American Association of Physicists in Medicine Computed Tomography Performance Phantom.\",\"authors\":\"Choirul Anam, Ariij Naufal, Yanurita Dwihapsari, Toshioh Fujibuchi, Geoff Dougherty\",\"doi\":\"10.4103/jmp.jmp_155_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The slice spacing has a crucial role in the accuracy of computed tomography (CT) images in sagittal and coronal planes. However, there is no practical method for measuring the accuracy of the slice spacing.</p><p><strong>Purpose: </strong>This study proposes a novel method to automatically measure the slice spacing using the American Association of Physicists in Medicine (AAPM) CT performance phantom.</p><p><strong>Methods: </strong>The AAPM CT performance phantom module 610-04 was used to measure slice spacing. The process of slice spacing measurement involves a pair of axial images of the module containing ramp aluminum objects located at adjacent slice positions. The middle aluminum plate of each image was automatically segmented. Next, the two segmented images were combined to produce one image with two stair objects. The centroid coordinates of two stair objects were automatically determined. Subsequently, the distance between these two centroids was measured to directly indicate the slice spacing. For comparison, the slice spacing was calculated by accessing the slice position attributes from the DICOM header of both images. The proposed method was tested on phantom images with variations in slice spacing and field of view (FOV).</p><p><strong>Results: </strong>The results showed that the automatic measurement of slice spacing was quite accurate for all variations of slice spacing and FOV, with average differences of 9.0% and 9.3%, respectively.</p><p><strong>Conclusion: </strong>A new automated method for measuring the slice spacing using the AAPM CT phantom was successfully demonstrated and tested for variations of slice spacing and FOV. Slice spacing measurement may be considered an additional parameter to be checked in addition to other established parameters.</p>\",\"PeriodicalId\":51719,\"journal\":{\"name\":\"Journal of Medical Physics\",\"volume\":\"49 1\",\"pages\":\"103-109\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11141755/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jmp.jmp_155_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/3/30 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_155_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/30 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

背景:切片间距对计算机断层扫描(CT)矢状面和冠状面图像的准确性起着至关重要的作用。目的:本研究提出了一种使用美国医学物理学家协会(AAPM)CT 性能模型自动测量切片间距的新方法:方法:使用 AAPM CT 性能模型模块 610-04 测量切片间距。切片间距的测量过程包括在相邻切片位置拍摄一对包含斜面铝制物体的模块轴向图像。自动分割每张图像的中间铝板。然后,将两张分割后的图像合并,生成一张包含两个阶梯对象的图像。两个楼梯对象的中心点坐标被自动确定。随后,测量这两个中心点之间的距离,以直接显示切片间距。为了进行比较,切片间距是通过访问两幅图像的 DICOM 标头中的切片位置属性计算得出的。在切片间距和视野(FOV)变化的幻影图像上测试了所提出的方法:结果表明,自动测量切片间距对所有切片间距和视场变化都相当准确,平均差异分别为 9.0% 和 9.3%:利用 AAPM CT 模型测量切片间距的新自动方法已成功演示,并针对切片间距和 FOV 的变化进行了测试。除其他既定参数外,切片间距测量可被视为需要检查的额外参数。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A Practical Method for Slice Spacing Measurement Using the American Association of Physicists in Medicine Computed Tomography Performance Phantom.

Background: The slice spacing has a crucial role in the accuracy of computed tomography (CT) images in sagittal and coronal planes. However, there is no practical method for measuring the accuracy of the slice spacing.

Purpose: This study proposes a novel method to automatically measure the slice spacing using the American Association of Physicists in Medicine (AAPM) CT performance phantom.

Methods: The AAPM CT performance phantom module 610-04 was used to measure slice spacing. The process of slice spacing measurement involves a pair of axial images of the module containing ramp aluminum objects located at adjacent slice positions. The middle aluminum plate of each image was automatically segmented. Next, the two segmented images were combined to produce one image with two stair objects. The centroid coordinates of two stair objects were automatically determined. Subsequently, the distance between these two centroids was measured to directly indicate the slice spacing. For comparison, the slice spacing was calculated by accessing the slice position attributes from the DICOM header of both images. The proposed method was tested on phantom images with variations in slice spacing and field of view (FOV).

Results: The results showed that the automatic measurement of slice spacing was quite accurate for all variations of slice spacing and FOV, with average differences of 9.0% and 9.3%, respectively.

Conclusion: A new automated method for measuring the slice spacing using the AAPM CT phantom was successfully demonstrated and tested for variations of slice spacing and FOV. Slice spacing measurement may be considered an additional parameter to be checked in addition to other established parameters.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Physics
Journal of Medical Physics RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.10
自引率
11.10%
发文量
55
审稿时长
30 weeks
期刊介绍: JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.
期刊最新文献
A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans. A Study on Radiation Level at the Treatment Plane Due to Induced Activity in Linear Accelerator Head. Advancements and Applications of Three-dimensional Printing Technology in Surgery. Agar-based Phantom for Evaluating Targeting of High-intensity Focused Ultrasound Systems for Breast Ablation. An Analysis of Radiotherapy Machine Requirements in India: Impact of the Pandemic and Regional Disparities.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1