{"title":"[根据 GOST R ISO 11095-2007 标准对乙醇质量浓度测量校准线性函数的充分评估]。","authors":"M S Zatyamina, L A Smirnova, O V Ostrovskiy","doi":"10.17116/sudmed20246703134","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To assess the adequacy of linear function of calibration according to GOST R ISO 11095-2007 for ethanol mass concentration measurement using internal reference materials (RMs).</p><p><strong>Material and methods: </strong>An experiment on calibration in accordance with the GOST R ISO 11095-2007 National standard of the RF was carried out using internal RMs, namely aqueous solutions of ethanol at different concentrations. Measurements were performed for two subbands of ethanol concentrations at RMs: 0.15-1.05 and 1.0-7.0 mg/ml - according to the certified methodology.</p><p><strong>Results: </strong>The graphs of the calibration's functions based on experimental data are consistent with the assumption of the calibration function's linearity, as well as the assumption of the standard deviation's constance of residues is equitable for two subbands of RMs.</p><p><strong>Conclusion: </strong>Proven linear models in the calibration experiment may be recommended for use in the ethanol mass concentration measurement.</p>","PeriodicalId":35704,"journal":{"name":"Sudebno-Meditsinskaya Ekspertisa","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Adequate assessment of linear function of calibration according to GOST R ISO 11095-2007 for ethanol mass concentration measurement].\",\"authors\":\"M S Zatyamina, L A Smirnova, O V Ostrovskiy\",\"doi\":\"10.17116/sudmed20246703134\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To assess the adequacy of linear function of calibration according to GOST R ISO 11095-2007 for ethanol mass concentration measurement using internal reference materials (RMs).</p><p><strong>Material and methods: </strong>An experiment on calibration in accordance with the GOST R ISO 11095-2007 National standard of the RF was carried out using internal RMs, namely aqueous solutions of ethanol at different concentrations. Measurements were performed for two subbands of ethanol concentrations at RMs: 0.15-1.05 and 1.0-7.0 mg/ml - according to the certified methodology.</p><p><strong>Results: </strong>The graphs of the calibration's functions based on experimental data are consistent with the assumption of the calibration function's linearity, as well as the assumption of the standard deviation's constance of residues is equitable for two subbands of RMs.</p><p><strong>Conclusion: </strong>Proven linear models in the calibration experiment may be recommended for use in the ethanol mass concentration measurement.</p>\",\"PeriodicalId\":35704,\"journal\":{\"name\":\"Sudebno-Meditsinskaya Ekspertisa\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Sudebno-Meditsinskaya Ekspertisa\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.17116/sudmed20246703134\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sudebno-Meditsinskaya Ekspertisa","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17116/sudmed20246703134","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的评估根据 GOST R ISO 11095-2007 标准使用内部参考物质(RMs)测量乙醇质量浓度时校准线性函数的适当性:根据 GOST R ISO 11095-2007 RF 国家标准,使用内部参考物质(即不同浓度的乙醇水溶液)进行了校准实验。对两个子波段的乙醇浓度(RMs)进行了测量:结果:结果:根据实验数据绘制的校准函数图与校准函数线性假设一致,两个子波段 RM 的残留物标准偏差常数假设也一致:结论:校准实验中经过验证的线性模型可推荐用于乙醇质量浓度测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[Adequate assessment of linear function of calibration according to GOST R ISO 11095-2007 for ethanol mass concentration measurement].

Objective: To assess the adequacy of linear function of calibration according to GOST R ISO 11095-2007 for ethanol mass concentration measurement using internal reference materials (RMs).

Material and methods: An experiment on calibration in accordance with the GOST R ISO 11095-2007 National standard of the RF was carried out using internal RMs, namely aqueous solutions of ethanol at different concentrations. Measurements were performed for two subbands of ethanol concentrations at RMs: 0.15-1.05 and 1.0-7.0 mg/ml - according to the certified methodology.

Results: The graphs of the calibration's functions based on experimental data are consistent with the assumption of the calibration function's linearity, as well as the assumption of the standard deviation's constance of residues is equitable for two subbands of RMs.

Conclusion: Proven linear models in the calibration experiment may be recommended for use in the ethanol mass concentration measurement.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Sudebno-Meditsinskaya Ekspertisa
Sudebno-Meditsinskaya Ekspertisa Medicine-Pathology and Forensic Medicine
CiteScore
0.60
自引率
0.00%
发文量
70
期刊介绍: The journal is concerned with the theory and practice of forensic medicine - the problems of thanatology, traumatology, toxicology, serology, forensic obstetrics, forensic dentistry, forensic psychiatry, forensic chemistry, physicotechnical methods of investigation, history of forensic medicine and some problems of criminology and legal laws related to forensic medicine. It publishes original studies by Russian authors, casuistry surveys, abstracts and reviews of Russian and foreign literature, scientific information, reports on scientific conferences.
期刊最新文献
[Comparative analysis of factors influencing suicidal actions performance before and during the COVID-19 pandemic]. [Correlation between defects in medical care and data of causal analysis]. [Determination of 2,4-dimethylhydroxybenzene by chromatographic methods in forensic toxicological research of biological material]. [Development of a carbamazepine determination method based on high-performance liquid chromatography with diode array]. [Development of selective method for mebeverine detection in blood].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1