{"title":"利用磁心动图评估冠状动脉微血管功能障碍:古老技术的新应用","authors":"Namrita Ashokprabhu , Khaled Ziada , Edouard Daher , Leslie Cho , Christian W. Schmidt , Yulith Roca , Cassady Palmer , Sukhleen Kaur , Timothy D. Henry , Carl J. Pepine , Odayme Quesada","doi":"10.1016/j.ahjo.2024.100424","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>In patients with angina and non-obstructive coronary artery disease (ANOCA), diagnosis of coronary microvascular dysfunction (CMD) remains an unmet need. Magnetocardiography (MCG), is a rest-based, non-invasive scan that can detect weak electrophysiological changes that occur at the early phase of ischemia.</p></div><div><h3>Objective</h3><p>This study assessed the ability of MCG to detect CMD in ANOCA patients as compared to reference standard, invasive coronary flow reserve (CFR).</p></div><div><h3>Methods</h3><p>Patients with ANOCA and invasive coronary physiologic assessment using intracoronary flow measurements with Doppler and thermodilution methods were enrolled. CMD was defined dichotomously as an invasive CFR < 2.0 by Doppler or thermodilution assessment. Noninvasive 36-channel 90-s MCG scan was performed and quantitative assessment of four distinct MCG features was completed. We evaluated the diagnostic performance of 2 or more abnormal MCG features to detect CMD in the overall cohort and performed a subgroup analysis in the subset of patients with Doppler CFR assessment.</p></div><div><h3>Results</h3><p>Among 79 ANOCA patients, 25 were CMD positive and 54 patients were CMD negative by CFR. Using invasive CFR as reference, MCG had an ROC AUC of 0.66 with a sensitivity of 68 % and specificity of 65 % for the detection of CMD. In the subgroup with Doppler CFR assessment, MCG had an ROC AUC of 0.76 with a sensitivity of 75 % and specificity of 77 %.</p></div><div><h3>Conclusions</h3><p>In ANOCA patients, MCG demonstrates the ability to detect CMD using a 90-second non-invasive scan without the need for an intravenous stressor or ionizing radiation. Further investigations are needed to validate an MCG-based diagnostic pathway for CMD.</p></div>","PeriodicalId":72158,"journal":{"name":"American heart journal plus : cardiology research and practice","volume":"44 ","pages":"Article 100424"},"PeriodicalIF":1.3000,"publicationDate":"2024-07-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666602224000673/pdfft?md5=68a2dc70cbc1a806b2918184b6b3d367&pid=1-s2.0-S2666602224000673-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Evaluation of coronary microvascular dysfunction using magnetocardiography: A new application to an old technology\",\"authors\":\"Namrita Ashokprabhu , Khaled Ziada , Edouard Daher , Leslie Cho , Christian W. Schmidt , Yulith Roca , Cassady Palmer , Sukhleen Kaur , Timothy D. Henry , Carl J. Pepine , Odayme Quesada\",\"doi\":\"10.1016/j.ahjo.2024.100424\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>In patients with angina and non-obstructive coronary artery disease (ANOCA), diagnosis of coronary microvascular dysfunction (CMD) remains an unmet need. Magnetocardiography (MCG), is a rest-based, non-invasive scan that can detect weak electrophysiological changes that occur at the early phase of ischemia.</p></div><div><h3>Objective</h3><p>This study assessed the ability of MCG to detect CMD in ANOCA patients as compared to reference standard, invasive coronary flow reserve (CFR).</p></div><div><h3>Methods</h3><p>Patients with ANOCA and invasive coronary physiologic assessment using intracoronary flow measurements with Doppler and thermodilution methods were enrolled. CMD was defined dichotomously as an invasive CFR < 2.0 by Doppler or thermodilution assessment. Noninvasive 36-channel 90-s MCG scan was performed and quantitative assessment of four distinct MCG features was completed. We evaluated the diagnostic performance of 2 or more abnormal MCG features to detect CMD in the overall cohort and performed a subgroup analysis in the subset of patients with Doppler CFR assessment.</p></div><div><h3>Results</h3><p>Among 79 ANOCA patients, 25 were CMD positive and 54 patients were CMD negative by CFR. Using invasive CFR as reference, MCG had an ROC AUC of 0.66 with a sensitivity of 68 % and specificity of 65 % for the detection of CMD. In the subgroup with Doppler CFR assessment, MCG had an ROC AUC of 0.76 with a sensitivity of 75 % and specificity of 77 %.</p></div><div><h3>Conclusions</h3><p>In ANOCA patients, MCG demonstrates the ability to detect CMD using a 90-second non-invasive scan without the need for an intravenous stressor or ionizing radiation. Further investigations are needed to validate an MCG-based diagnostic pathway for CMD.</p></div>\",\"PeriodicalId\":72158,\"journal\":{\"name\":\"American heart journal plus : cardiology research and practice\",\"volume\":\"44 \",\"pages\":\"Article 100424\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-07-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2666602224000673/pdfft?md5=68a2dc70cbc1a806b2918184b6b3d367&pid=1-s2.0-S2666602224000673-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"American heart journal plus : cardiology research and practice\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2666602224000673\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"American heart journal plus : cardiology research and practice","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666602224000673","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Evaluation of coronary microvascular dysfunction using magnetocardiography: A new application to an old technology
Background
In patients with angina and non-obstructive coronary artery disease (ANOCA), diagnosis of coronary microvascular dysfunction (CMD) remains an unmet need. Magnetocardiography (MCG), is a rest-based, non-invasive scan that can detect weak electrophysiological changes that occur at the early phase of ischemia.
Objective
This study assessed the ability of MCG to detect CMD in ANOCA patients as compared to reference standard, invasive coronary flow reserve (CFR).
Methods
Patients with ANOCA and invasive coronary physiologic assessment using intracoronary flow measurements with Doppler and thermodilution methods were enrolled. CMD was defined dichotomously as an invasive CFR < 2.0 by Doppler or thermodilution assessment. Noninvasive 36-channel 90-s MCG scan was performed and quantitative assessment of four distinct MCG features was completed. We evaluated the diagnostic performance of 2 or more abnormal MCG features to detect CMD in the overall cohort and performed a subgroup analysis in the subset of patients with Doppler CFR assessment.
Results
Among 79 ANOCA patients, 25 were CMD positive and 54 patients were CMD negative by CFR. Using invasive CFR as reference, MCG had an ROC AUC of 0.66 with a sensitivity of 68 % and specificity of 65 % for the detection of CMD. In the subgroup with Doppler CFR assessment, MCG had an ROC AUC of 0.76 with a sensitivity of 75 % and specificity of 77 %.
Conclusions
In ANOCA patients, MCG demonstrates the ability to detect CMD using a 90-second non-invasive scan without the need for an intravenous stressor or ionizing radiation. Further investigations are needed to validate an MCG-based diagnostic pathway for CMD.