{"title":"巴西南部城市初级卫生保健中公共配发血糖仪的分析性能","authors":"Isabelle L. Silva, Flávia Martinello","doi":"10.1016/j.plabm.2024.e00421","DOIUrl":null,"url":null,"abstract":"<div><h3>Aims</h3><p>This study aimed to assess the use of glucometers by patients and the analytical performance of glucometers provided by the primary care services.</p></div><div><h3>Methods</h3><p>The analytical performance of 48 glucometers Accu-Chek® Active, was assessed through quintuplicate analyses of one Roche and one PNCQ (National Quality Control Program) control sample at different concentrations; 31 were also evaluated by a single proficiency testing sample. The evaluation metrics included imprecision, bias, and total error and were measured according to quality specifications based on biological variation (QSBV). Glucometer users answered a questionnaire regarding their experience.</p></div><div><h3>Results</h3><p>Among the 48 glucometers evaluated with internal control samples, 17 met precision criteria at both control levels according to QSBV, while 24 met the criteria at only one control level. Of the 31 glucometers further evaluated through proficiency test, 11 met accuracy criteria according to QSBV, and only one device showed an unacceptable result. Out of these 31, only 15 demonstrated a total error within the acceptable maximum limits based on QSBV.</p></div><div><h3>Conclusions</h3><p>Overall, our findings showed that patients had a good understanding of glucometer usage and suggested that some glucometers should be replaced, as they sometimes failed to meet even the manufacturer’s acceptable variation limits, and/or did not meet QSBV.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551724000672/pdfft?md5=4da41e980a7e2125105f74c43088cc49&pid=1-s2.0-S2352551724000672-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Analytical performance of publicly dispensed glucometers in primary health care in a southern Brazilian city\",\"authors\":\"Isabelle L. Silva, Flávia Martinello\",\"doi\":\"10.1016/j.plabm.2024.e00421\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Aims</h3><p>This study aimed to assess the use of glucometers by patients and the analytical performance of glucometers provided by the primary care services.</p></div><div><h3>Methods</h3><p>The analytical performance of 48 glucometers Accu-Chek® Active, was assessed through quintuplicate analyses of one Roche and one PNCQ (National Quality Control Program) control sample at different concentrations; 31 were also evaluated by a single proficiency testing sample. The evaluation metrics included imprecision, bias, and total error and were measured according to quality specifications based on biological variation (QSBV). Glucometer users answered a questionnaire regarding their experience.</p></div><div><h3>Results</h3><p>Among the 48 glucometers evaluated with internal control samples, 17 met precision criteria at both control levels according to QSBV, while 24 met the criteria at only one control level. Of the 31 glucometers further evaluated through proficiency test, 11 met accuracy criteria according to QSBV, and only one device showed an unacceptable result. Out of these 31, only 15 demonstrated a total error within the acceptable maximum limits based on QSBV.</p></div><div><h3>Conclusions</h3><p>Overall, our findings showed that patients had a good understanding of glucometer usage and suggested that some glucometers should be replaced, as they sometimes failed to meet even the manufacturer’s acceptable variation limits, and/or did not meet QSBV.</p></div>\",\"PeriodicalId\":20421,\"journal\":{\"name\":\"Practical Laboratory Medicine\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2352551724000672/pdfft?md5=4da41e980a7e2125105f74c43088cc49&pid=1-s2.0-S2352551724000672-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2352551724000672\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551724000672","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
Analytical performance of publicly dispensed glucometers in primary health care in a southern Brazilian city
Aims
This study aimed to assess the use of glucometers by patients and the analytical performance of glucometers provided by the primary care services.
Methods
The analytical performance of 48 glucometers Accu-Chek® Active, was assessed through quintuplicate analyses of one Roche and one PNCQ (National Quality Control Program) control sample at different concentrations; 31 were also evaluated by a single proficiency testing sample. The evaluation metrics included imprecision, bias, and total error and were measured according to quality specifications based on biological variation (QSBV). Glucometer users answered a questionnaire regarding their experience.
Results
Among the 48 glucometers evaluated with internal control samples, 17 met precision criteria at both control levels according to QSBV, while 24 met the criteria at only one control level. Of the 31 glucometers further evaluated through proficiency test, 11 met accuracy criteria according to QSBV, and only one device showed an unacceptable result. Out of these 31, only 15 demonstrated a total error within the acceptable maximum limits based on QSBV.
Conclusions
Overall, our findings showed that patients had a good understanding of glucometer usage and suggested that some glucometers should be replaced, as they sometimes failed to meet even the manufacturer’s acceptable variation limits, and/or did not meet QSBV.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.