CA 19-9 免疫测定的分析干扰导致一名患有低级别阑尾肿瘤的年轻女性出现假阳性结果:一份教育性病例报告和简要文献综述。

Jérôme A Denis, Mathieu Boissan, Eric Rouzaud, Jean-Marc Lacorte, Marc Pocard, Laurent Genser
{"title":"CA 19-9 免疫测定的分析干扰导致一名患有低级别阑尾肿瘤的年轻女性出现假阳性结果:一份教育性病例报告和简要文献综述。","authors":"Jérôme A Denis, Mathieu Boissan, Eric Rouzaud, Jean-Marc Lacorte, Marc Pocard, Laurent Genser","doi":"10.1684/abc.2024.1892","DOIUrl":null,"url":null,"abstract":"<p><p>CA 19-9 (carbohydrate antigen 19-9) is a tumor marker widely used for the follow-up of patients with pancreatic cancer and other digestive neoplasia. This case report describes a discrepancy between the results of serum CA 19-9 analyses using the Alinity analytical platform (Abbott™) and two other techniques, Kryptor Gold (ThermoFisher Scientific™) and Cobas E411 (Roche™), in the context of a young woman with appendiceal mucocele. In this context, when the serum concentration of CA 19-9 is high, it may raise concerns about potential malignancy or rupture of the mucocele that may lead to tumoral dissemination in the abdominal cavity. In the present case, we observed with Alinity a false elevation in CA 19-9 concentration at 190 kU/L (normal range < 37 kU/L) before appendix resection that continued to increase until reaching 619 kU/L six months after surgery. This situation led to unnecessary additional tests, increased hospitalization time and stress for the patient who also had to interrupt her medically assisted reproduction project. We solved this case using new measurements in CA 19-9 concentration with two other techniques, Kryptor Gold and Cobas E411, and we identified an analytical interference caused by the presence of heterophile antibodies. In all cases, abnormal result initially obtained with Alinity was found below normal range not only with the two other techniques but also with Alinity after a neutralisation step by using Heterophile Blocking Tubes (Scantibodies Laboratory™). Analytical interferences in medical tests can lead to inappropriate medical care. It is an important issue requiring a continuing training of biologists who must be aware of these problems, which are recurring concerns and are not always easy to identify in laboratories of medical biology, in particular when immunoassays are used. This case report also provides an opportunity to do a brief review of the literature and to remind some recommendations and actions to take into consideration in the presence of discrepancies between the clinic and the biology, in particular, one of them is to measure the biological analyte with a different technique. Moreover, the use of Heterophile Blocking Tubes neutralizing specifically the heterophile antibodies may be useful. In all cases, dialogue between clinicians and biologists remains essential.</p>","PeriodicalId":93870,"journal":{"name":"Annales de biologie clinique","volume":"82 3","pages":"321-328"},"PeriodicalIF":0.0000,"publicationDate":"2024-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Analytical interference in CA 19-9 immunoassay generates false-positive results in a young woman with a low-grade appendiceal tumor: An educational case report and brief literature review.\",\"authors\":\"Jérôme A Denis, Mathieu Boissan, Eric Rouzaud, Jean-Marc Lacorte, Marc Pocard, Laurent Genser\",\"doi\":\"10.1684/abc.2024.1892\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>CA 19-9 (carbohydrate antigen 19-9) is a tumor marker widely used for the follow-up of patients with pancreatic cancer and other digestive neoplasia. This case report describes a discrepancy between the results of serum CA 19-9 analyses using the Alinity analytical platform (Abbott™) and two other techniques, Kryptor Gold (ThermoFisher Scientific™) and Cobas E411 (Roche™), in the context of a young woman with appendiceal mucocele. In this context, when the serum concentration of CA 19-9 is high, it may raise concerns about potential malignancy or rupture of the mucocele that may lead to tumoral dissemination in the abdominal cavity. In the present case, we observed with Alinity a false elevation in CA 19-9 concentration at 190 kU/L (normal range < 37 kU/L) before appendix resection that continued to increase until reaching 619 kU/L six months after surgery. This situation led to unnecessary additional tests, increased hospitalization time and stress for the patient who also had to interrupt her medically assisted reproduction project. We solved this case using new measurements in CA 19-9 concentration with two other techniques, Kryptor Gold and Cobas E411, and we identified an analytical interference caused by the presence of heterophile antibodies. In all cases, abnormal result initially obtained with Alinity was found below normal range not only with the two other techniques but also with Alinity after a neutralisation step by using Heterophile Blocking Tubes (Scantibodies Laboratory™). Analytical interferences in medical tests can lead to inappropriate medical care. It is an important issue requiring a continuing training of biologists who must be aware of these problems, which are recurring concerns and are not always easy to identify in laboratories of medical biology, in particular when immunoassays are used. This case report also provides an opportunity to do a brief review of the literature and to remind some recommendations and actions to take into consideration in the presence of discrepancies between the clinic and the biology, in particular, one of them is to measure the biological analyte with a different technique. Moreover, the use of Heterophile Blocking Tubes neutralizing specifically the heterophile antibodies may be useful. In all cases, dialogue between clinicians and biologists remains essential.</p>\",\"PeriodicalId\":93870,\"journal\":{\"name\":\"Annales de biologie clinique\",\"volume\":\"82 3\",\"pages\":\"321-328\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-08-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annales de biologie clinique\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1684/abc.2024.1892\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales de biologie clinique","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1684/abc.2024.1892","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

CA 19-9(碳水化合物抗原 19-9)是一种肿瘤标志物,广泛用于胰腺癌和其他消化道肿瘤患者的随访。本病例报告描述了在对一名患有阑尾粘液瘤的年轻女性进行血清 CA 19-9 分析时,使用 Alinity 分析平台(Abbott™)和其他两种技术(Kryptor Gold(ThermoFisher Scientific™)和 Cobas E411(Roche™))得出的结果之间的差异。在这种情况下,当血清中的 CA 19-9 浓度较高时,可能会引发对潜在恶性肿瘤或粘液瘤破裂的担忧,从而导致肿瘤在腹腔内扩散。在本病例中,我们用 Alinity 观察到 CA 19-9 浓度假性升高,达到 190 kU/L(正常范围
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Analytical interference in CA 19-9 immunoassay generates false-positive results in a young woman with a low-grade appendiceal tumor: An educational case report and brief literature review.

CA 19-9 (carbohydrate antigen 19-9) is a tumor marker widely used for the follow-up of patients with pancreatic cancer and other digestive neoplasia. This case report describes a discrepancy between the results of serum CA 19-9 analyses using the Alinity analytical platform (Abbott™) and two other techniques, Kryptor Gold (ThermoFisher Scientific™) and Cobas E411 (Roche™), in the context of a young woman with appendiceal mucocele. In this context, when the serum concentration of CA 19-9 is high, it may raise concerns about potential malignancy or rupture of the mucocele that may lead to tumoral dissemination in the abdominal cavity. In the present case, we observed with Alinity a false elevation in CA 19-9 concentration at 190 kU/L (normal range < 37 kU/L) before appendix resection that continued to increase until reaching 619 kU/L six months after surgery. This situation led to unnecessary additional tests, increased hospitalization time and stress for the patient who also had to interrupt her medically assisted reproduction project. We solved this case using new measurements in CA 19-9 concentration with two other techniques, Kryptor Gold and Cobas E411, and we identified an analytical interference caused by the presence of heterophile antibodies. In all cases, abnormal result initially obtained with Alinity was found below normal range not only with the two other techniques but also with Alinity after a neutralisation step by using Heterophile Blocking Tubes (Scantibodies Laboratory™). Analytical interferences in medical tests can lead to inappropriate medical care. It is an important issue requiring a continuing training of biologists who must be aware of these problems, which are recurring concerns and are not always easy to identify in laboratories of medical biology, in particular when immunoassays are used. This case report also provides an opportunity to do a brief review of the literature and to remind some recommendations and actions to take into consideration in the presence of discrepancies between the clinic and the biology, in particular, one of them is to measure the biological analyte with a different technique. Moreover, the use of Heterophile Blocking Tubes neutralizing specifically the heterophile antibodies may be useful. In all cases, dialogue between clinicians and biologists remains essential.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Complementarity between Bayesian Internal Quality Control results management and External Quality Assessment bivariate z-score analysis: application to a concrete case study. [Can TDABC (Time-Driven Activity-Based Costing) help the NOVO hospital's medical biology laboratory to value the appropriate prescription? - Role of the biologist in reducing healthcare costs]. [Clinical associations of anti-Jo1 antibodies in a Moroccan population]. [Clinical-biological approaches to the spectrophotometric detection of oxyhemoglobin and bilirubin in CSF in the management of aneurysmal subarachnoid hemorrhage]. [Evaluation of the CellaVision® DM-1200 system for detecting and quantifying schistocytes].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1