{"title":"对快速 ChannelCheck™ 清洁测试进行模拟使用评估,以优化柔性内窥镜通道中有机残留物的检测。","authors":"K. Kulkarni , M. Gavette , M.J. Alfa","doi":"10.1016/j.jhin.2024.07.013","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>The need to monitor manual cleaning of high-risk endoscopes is recommended or more so required by the current endoscope reprocessing guidelines. The objective of this study was to establish the optimal extraction volume for colonoscopes and bronchoscopes and demonstrate the extraction efficacy for the ChannelCheck™ rapid test.</p></div><div><h3>Methods</h3><p>The test soil utilized as a positive control was ATS2015 containing 20% defibrinated bovine blood. The extraction from the instrument channel of a colonoscope and bronchoscope was evaluated to establish the optimal extraction volume and the extraction efficacy for protein, carbohydrate and haemoglobin.</p></div><div><h3>Results</h3><p>Of the extraction volumes tested, 10 mL was optimal for both colonoscopes and bronchoscopes. The extraction efficacy was 91% for carbohydrate, 83.7% for haemoglobin and 82.4% for protein.</p></div><div><h3>Conclusions</h3><p>The limit of detection for these analytes by the ChannelCheck rapid test meet or exceed the established levels that correlate with adequate manual cleaning of flexible endoscopes.</p></div>","PeriodicalId":54806,"journal":{"name":"Journal of Hospital Infection","volume":"152 ","pages":"Pages 66-72"},"PeriodicalIF":3.9000,"publicationDate":"2024-08-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Simulated-use evaluation of rapid ChannelCheck™ cleaning test for optimal detection of organic residues in flexible endoscope channels\",\"authors\":\"K. Kulkarni , M. Gavette , M.J. Alfa\",\"doi\":\"10.1016/j.jhin.2024.07.013\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>The need to monitor manual cleaning of high-risk endoscopes is recommended or more so required by the current endoscope reprocessing guidelines. The objective of this study was to establish the optimal extraction volume for colonoscopes and bronchoscopes and demonstrate the extraction efficacy for the ChannelCheck™ rapid test.</p></div><div><h3>Methods</h3><p>The test soil utilized as a positive control was ATS2015 containing 20% defibrinated bovine blood. The extraction from the instrument channel of a colonoscope and bronchoscope was evaluated to establish the optimal extraction volume and the extraction efficacy for protein, carbohydrate and haemoglobin.</p></div><div><h3>Results</h3><p>Of the extraction volumes tested, 10 mL was optimal for both colonoscopes and bronchoscopes. The extraction efficacy was 91% for carbohydrate, 83.7% for haemoglobin and 82.4% for protein.</p></div><div><h3>Conclusions</h3><p>The limit of detection for these analytes by the ChannelCheck rapid test meet or exceed the established levels that correlate with adequate manual cleaning of flexible endoscopes.</p></div>\",\"PeriodicalId\":54806,\"journal\":{\"name\":\"Journal of Hospital Infection\",\"volume\":\"152 \",\"pages\":\"Pages 66-72\"},\"PeriodicalIF\":3.9000,\"publicationDate\":\"2024-08-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Hospital Infection\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0195670124002676\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"INFECTIOUS DISEASES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Hospital Infection","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0195670124002676","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
Simulated-use evaluation of rapid ChannelCheck™ cleaning test for optimal detection of organic residues in flexible endoscope channels
Background
The need to monitor manual cleaning of high-risk endoscopes is recommended or more so required by the current endoscope reprocessing guidelines. The objective of this study was to establish the optimal extraction volume for colonoscopes and bronchoscopes and demonstrate the extraction efficacy for the ChannelCheck™ rapid test.
Methods
The test soil utilized as a positive control was ATS2015 containing 20% defibrinated bovine blood. The extraction from the instrument channel of a colonoscope and bronchoscope was evaluated to establish the optimal extraction volume and the extraction efficacy for protein, carbohydrate and haemoglobin.
Results
Of the extraction volumes tested, 10 mL was optimal for both colonoscopes and bronchoscopes. The extraction efficacy was 91% for carbohydrate, 83.7% for haemoglobin and 82.4% for protein.
Conclusions
The limit of detection for these analytes by the ChannelCheck rapid test meet or exceed the established levels that correlate with adequate manual cleaning of flexible endoscopes.
期刊介绍:
The Journal of Hospital Infection is the editorially independent scientific publication of the Healthcare Infection Society. The aim of the Journal is to publish high quality research and information relating to infection prevention and control that is relevant to an international audience.
The Journal welcomes submissions that relate to all aspects of infection prevention and control in healthcare settings. This includes submissions that:
provide new insight into the epidemiology, surveillance, or prevention and control of healthcare-associated infections and antimicrobial resistance in healthcare settings;
provide new insight into cleaning, disinfection and decontamination;
provide new insight into the design of healthcare premises;
describe novel aspects of outbreaks of infection;
throw light on techniques for effective antimicrobial stewardship;
describe novel techniques (laboratory-based or point of care) for the detection of infection or antimicrobial resistance in the healthcare setting, particularly if these can be used to facilitate infection prevention and control;
improve understanding of the motivations of safe healthcare behaviour, or describe techniques for achieving behavioural and cultural change;
improve understanding of the use of IT systems in infection surveillance and prevention and control.