Olaf Dietrich, Sergio Lentini, Osman Öcal, Pierre Bour, Thibaut L Faller, Valéry Ozenne, Jens Ricke, Max Seidensticker
{"title":"微波加热期间凝胶模型中三维实时磁共振成像温度绘图的准确性。","authors":"Olaf Dietrich, Sergio Lentini, Osman Öcal, Pierre Bour, Thibaut L Faller, Valéry Ozenne, Jens Ricke, Max Seidensticker","doi":"10.1186/s41747-024-00479-5","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Interventional magnetic resonance imaging (MRI) can provide a comprehensive setting for microwave ablation of tumors with real-time monitoring of the energy delivery using MRI-based temperature mapping. The purpose of this study was to quantify the accuracy of three-dimensional (3D) real-time MRI temperature mapping during microwave heating in vitro by comparing MRI thermometry data to reference data measured by fiber-optical thermometry.</p><p><strong>Methods: </strong>Nine phantom experiments were evaluated in agar-based gel phantoms using an in-room MR-conditional microwave system and MRI thermometry. MRI measurements were performed for 700 s (25 slices; temporal resolution 2 s). The temperature was monitored with two fiber-optical temperature sensors approximately 5 mm and 10 mm distant from the microwave antenna. Temperature curves of the sensors were compared to MRI temperature data of single-voxel regions of interest (ROIs) at the sensor tips; the accuracy of MRI thermometry was assessed as the root-mean-squared (RMS)-averaged temperature difference. Eighteen neighboring voxels around the original ROI were also evaluated and the voxel with the smallest temperature difference was additionally selected for further evaluation.</p><p><strong>Results: </strong>The maximum temperature changes measured by the fiber-optical sensors ranged from 7.3 K to 50.7 K. The median RMS-averaged temperature differences in the originally selected voxels ranged from 1.4 K to 3.4 K. When evaluating the minimum-difference voxel from the neighborhood, the temperature differences ranged from 0.5 K to 0.9 K. The microwave antenna and the MRI-conditional in-room microwave generator did not induce relevant radiofrequency artifacts.</p><p><strong>Conclusion: </strong>Accurate 3D real-time MRI temperature mapping during microwave heating with very low RMS-averaged temperature errors below 1 K is feasible in gel phantoms.</p><p><strong>Relevance statement: </strong>Accurate MRI-based volumetric real-time monitoring of temperature distribution and thermal dose is highly relevant in clinical MRI-based interventions and can be expected to improve local tumor control, as well as procedural safety by extending the limits of thermal (e.g., microwave) ablation of tumors in the liver and in other organs.</p><p><strong>Key points: </strong>Interventional MRI can provide a comprehensive setting for the microwave ablation of tumors. MRI can monitor the microwave ablation using real-time MRI-based temperature mapping. 3D real-time MRI temperature mapping during microwave heating is feasible. Measured temperature errors were below 1 °C in gel phantoms. The active in-room microwave generator did not induce any relevant radiofrequency artifacts.</p>","PeriodicalId":36926,"journal":{"name":"European Radiology Experimental","volume":"8 1","pages":"92"},"PeriodicalIF":3.7000,"publicationDate":"2024-08-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11324620/pdf/","citationCount":"0","resultStr":"{\"title\":\"Accuracy of 3D real-time MRI temperature mapping in gel phantoms during microwave heating.\",\"authors\":\"Olaf Dietrich, Sergio Lentini, Osman Öcal, Pierre Bour, Thibaut L Faller, Valéry Ozenne, Jens Ricke, Max Seidensticker\",\"doi\":\"10.1186/s41747-024-00479-5\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Interventional magnetic resonance imaging (MRI) can provide a comprehensive setting for microwave ablation of tumors with real-time monitoring of the energy delivery using MRI-based temperature mapping. The purpose of this study was to quantify the accuracy of three-dimensional (3D) real-time MRI temperature mapping during microwave heating in vitro by comparing MRI thermometry data to reference data measured by fiber-optical thermometry.</p><p><strong>Methods: </strong>Nine phantom experiments were evaluated in agar-based gel phantoms using an in-room MR-conditional microwave system and MRI thermometry. MRI measurements were performed for 700 s (25 slices; temporal resolution 2 s). The temperature was monitored with two fiber-optical temperature sensors approximately 5 mm and 10 mm distant from the microwave antenna. Temperature curves of the sensors were compared to MRI temperature data of single-voxel regions of interest (ROIs) at the sensor tips; the accuracy of MRI thermometry was assessed as the root-mean-squared (RMS)-averaged temperature difference. Eighteen neighboring voxels around the original ROI were also evaluated and the voxel with the smallest temperature difference was additionally selected for further evaluation.</p><p><strong>Results: </strong>The maximum temperature changes measured by the fiber-optical sensors ranged from 7.3 K to 50.7 K. The median RMS-averaged temperature differences in the originally selected voxels ranged from 1.4 K to 3.4 K. When evaluating the minimum-difference voxel from the neighborhood, the temperature differences ranged from 0.5 K to 0.9 K. The microwave antenna and the MRI-conditional in-room microwave generator did not induce relevant radiofrequency artifacts.</p><p><strong>Conclusion: </strong>Accurate 3D real-time MRI temperature mapping during microwave heating with very low RMS-averaged temperature errors below 1 K is feasible in gel phantoms.</p><p><strong>Relevance statement: </strong>Accurate MRI-based volumetric real-time monitoring of temperature distribution and thermal dose is highly relevant in clinical MRI-based interventions and can be expected to improve local tumor control, as well as procedural safety by extending the limits of thermal (e.g., microwave) ablation of tumors in the liver and in other organs.</p><p><strong>Key points: </strong>Interventional MRI can provide a comprehensive setting for the microwave ablation of tumors. MRI can monitor the microwave ablation using real-time MRI-based temperature mapping. 3D real-time MRI temperature mapping during microwave heating is feasible. Measured temperature errors were below 1 °C in gel phantoms. The active in-room microwave generator did not induce any relevant radiofrequency artifacts.</p>\",\"PeriodicalId\":36926,\"journal\":{\"name\":\"European Radiology Experimental\",\"volume\":\"8 1\",\"pages\":\"92\"},\"PeriodicalIF\":3.7000,\"publicationDate\":\"2024-08-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11324620/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Radiology Experimental\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1186/s41747-024-00479-5\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Radiology Experimental","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41747-024-00479-5","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Accuracy of 3D real-time MRI temperature mapping in gel phantoms during microwave heating.
Background: Interventional magnetic resonance imaging (MRI) can provide a comprehensive setting for microwave ablation of tumors with real-time monitoring of the energy delivery using MRI-based temperature mapping. The purpose of this study was to quantify the accuracy of three-dimensional (3D) real-time MRI temperature mapping during microwave heating in vitro by comparing MRI thermometry data to reference data measured by fiber-optical thermometry.
Methods: Nine phantom experiments were evaluated in agar-based gel phantoms using an in-room MR-conditional microwave system and MRI thermometry. MRI measurements were performed for 700 s (25 slices; temporal resolution 2 s). The temperature was monitored with two fiber-optical temperature sensors approximately 5 mm and 10 mm distant from the microwave antenna. Temperature curves of the sensors were compared to MRI temperature data of single-voxel regions of interest (ROIs) at the sensor tips; the accuracy of MRI thermometry was assessed as the root-mean-squared (RMS)-averaged temperature difference. Eighteen neighboring voxels around the original ROI were also evaluated and the voxel with the smallest temperature difference was additionally selected for further evaluation.
Results: The maximum temperature changes measured by the fiber-optical sensors ranged from 7.3 K to 50.7 K. The median RMS-averaged temperature differences in the originally selected voxels ranged from 1.4 K to 3.4 K. When evaluating the minimum-difference voxel from the neighborhood, the temperature differences ranged from 0.5 K to 0.9 K. The microwave antenna and the MRI-conditional in-room microwave generator did not induce relevant radiofrequency artifacts.
Conclusion: Accurate 3D real-time MRI temperature mapping during microwave heating with very low RMS-averaged temperature errors below 1 K is feasible in gel phantoms.
Relevance statement: Accurate MRI-based volumetric real-time monitoring of temperature distribution and thermal dose is highly relevant in clinical MRI-based interventions and can be expected to improve local tumor control, as well as procedural safety by extending the limits of thermal (e.g., microwave) ablation of tumors in the liver and in other organs.
Key points: Interventional MRI can provide a comprehensive setting for the microwave ablation of tumors. MRI can monitor the microwave ablation using real-time MRI-based temperature mapping. 3D real-time MRI temperature mapping during microwave heating is feasible. Measured temperature errors were below 1 °C in gel phantoms. The active in-room microwave generator did not induce any relevant radiofrequency artifacts.